Simultaneous quantification of acylcarnitine isomers containing dicarboxylic acylcarnitines in human serum and urine by high-performance liquid chromatography/electrospray ionization tandem mass spectrometry

被引:49
作者
Maeda, Yasuhiro
Ito, Tetsuya
Suzuki, Atsuko
Kurono, Yukihisa
Ueta, Akihito
Yokoi, Kyoko
Sumi, Satoshi
Togari, Hajime
Sugiyama, Naruji
机构
[1] Nagoya City Univ, Grad Sch Pharmaceut Sci, Mizuho Ku, Lab Hosp Pharm, Nagoya, Aichi 4678603, Japan
[2] Nagoya City Univ, Grad Sch Med Sci, Dept Pediat, Nagoya, Aichi, Japan
[3] Aichi Gakuin Univ, Sch Dent, Dept Pediat, Nagoya, Aichi 464, Japan
关键词
D O I
10.1002/rcm.2905
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Tandem mass spectrometry (MS/MS) has become a prominent method for screening newborns for diseases such as organic acidemia and fatty acid oxidation defects, although current methods cannot separate acylcarnitine isomers. Accurate determination of dicarboxylic acylcarnitines such as methylmalonylcarnitine and glutarylcarnitine has not been carried out, because obtaining standards of these acylcarnitines is difficult. We attempted the individual determinations of acylcarnitines with isomers and dicarboxylic acylcarnitines by applying high-performance liquid chromatography (HPLC). Chromatographic separation was performed by gradient elution using a mixture of 0.08% aqueous ion-pairing agent and acetonitrile as the mobile phase. Mass transitions of m/z 161.8 -> 84.8 for carnitine and m/z 164.8 -> 84.8 for deuterated carnitine were monitored in positive ion electrospray ionization mode. One carnitine and 16 acylcarnitines were quantified. The limit of quantitation (LOQ) was 0.1 mu mol/L for methylmalonylcarnitine and 0.05 mu mol/L for the other acylcarnitines. Intra-day and inter-day coefficients of variance (CVs) were < 8.3% and < 8.8%, respectively, for all acylcarnitines in serum, and both were < 9.2% in urine. Mean recoveries were > 90% for all acylcarnitines. Human samples were quantified by this method. After addition of deuterated acylcarnitines as internal standards, acylcarnitines in serum or urine were extracted using a solid-phase extraction cartridge. In healthy adult individuals, isobutyryl-, 2-methylbutyryl- and isovalerylcarnitine were detected in serum and urine. Dicarboxylic acylcarnitines were detected in urine. High concentrations of methylmalonylcarnitine and propionylcarnitine were found in both the serum and the urine of a patient with methylmalonic acidemia. The described HPLC/MS/MS method could separate most acylcarnitine isomers and quantify them, potentially allowing detailed diagnoses and follow-up treatment for those diseases. Copyright (c) 2007 John Wiley & Sons, Ltd.
引用
收藏
页码:799 / 806
页数:8
相关论文
共 29 条
[1]  
Al-Dirbashi OY, 2005, CLIN CHEM, V51, P235
[2]   NEW SYNTHESIS OF (RS)-CARNITINE CHLORIDE [J].
BOOTS, SG ;
BOOTS, MR .
JOURNAL OF PHARMACEUTICAL SCIENCES, 1975, 64 (07) :1262-1264
[3]   CARNITINE - METABOLISM AND FUNCTIONS [J].
BREMER, J .
PHYSIOLOGICAL REVIEWS, 1983, 63 (04) :1420-1480
[4]   Age-related variations in acylcarnitine and free carnitine concentrations measured by tandem mass spectrometry [J].
Cavedon, CT ;
Bourdoux, P ;
Mertens, K ;
Van Thi, HV ;
Herremans, N ;
de Laet, C ;
Goyens, P .
CLINICAL CHEMISTRY, 2005, 51 (04) :745-752
[5]   Use of tandem mass spectrometry for multianalyte screening of dried blood specimens from newborns [J].
Chace, DH ;
Kalas, TA ;
Naylor, EW .
CLINICAL CHEMISTRY, 2003, 49 (11) :1797-1817
[6]  
Chace DH, 2001, CLIN CHEM, V47, P2040
[7]  
Chace DH, 2001, CLIN CHEM, V47, P1166
[8]   Modified and improved method for the measurement of plasma acylcarnitines by online extraction coupled to liquid chromatography-tandem mass spectrometry without derivatization [J].
Ghoshal, AK ;
Balay, J ;
Soldin, SJ .
CLINICA CHIMICA ACTA, 2006, 365 (1-2) :352-353
[9]   Rapid measurement of plasma acylcarnitines by liquid chromatography - tandem mass spectrometry without derivatization [J].
Ghoshal, AK ;
Guo, TD ;
Soukhova, N ;
Soldin, SJ .
CLINICA CHIMICA ACTA, 2005, 358 (1-2) :104-112
[10]   Prenatal diagnosis for organic acid disorders using two mass spectrometric methods, gas chromatography mass spectrometry and tandem mass spectrometry [J].
Hasegawa, Y ;
Iga, M ;
Kimura, M ;
Shigematsu, Y ;
Yamaguchi, S .
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2005, 823 (01) :13-17