Flow-injection chemiluminescence sensor for the determination of free chlorine in tap water

被引:0
作者
Qin, W
Zhang, ZJ
Liu, SN
机构
[1] SHAANXI NORMAL UNIV,DEPT CHEM,XIAN 710062,PEOPLES R CHINA
[2] NANJING UNIV,DEPT CHEM,NANJING 210008,PEOPLES R CHINA
关键词
chemical sensor; chemiluminescence; flow-injection; free chlorine;
D O I
暂无
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A novel chemiluminescence(CL) sensor combined with flow-injection analysis has been developed for deternlining free chlorine in tap water. The analytical reagent luminol was immobilized on an anion exchange resin column. While a volumn of sodium hydroxide passed through the column. luminol was eluted from the resin in alkaline aqueous solution and then mixed with a sample stream to react and produce CL. The CL emission intensity was correlated with the standard ClO- concentration in the range 1x10(-8) to 4x10(-5) g ml(-1). and the detection limit was 8x10(-9) g ml(-1) ClO-. Interfering metal ions present in water were effectively separated by a pre-column cation exchanger. A complete analysis, including sampling and washing, could be performed in 1 min with a relative standard deviation of less than 5%. The sensor was stable for over 200 times and has been applied successfully to the determination of ClO- in tap water.
引用
收藏
页码:11 / 19
页数:9
相关论文
共 12 条
[1]  
*APHA AWWA WPCF, 1985, STAND METH EX WAT WA, P294
[2]   DETERMINATION OF AIRBORNE FREE CHLORINE IN THE PRESENCE OF AMMONIA BY CAPILLARY COLUMN GAS-CHROMATOGRAPHY [J].
CHEPLEN, JM ;
BARROW, C ;
WHITE, EL .
ANALYTICAL CHEMISTRY, 1984, 56 (07) :1194-1196
[3]   PERFORMANCE OF THE ORION 97-70 TOTAL RESIDUAL CHLORINE ELECTRODE AT LOW CONCENTRATIONS AND ITS APPLICATION TO THE ANALYSIS OF COOLING WATERS [J].
DIMMOCK, NA ;
MIDGLEY, D .
TALANTA, 1982, 29 (07) :557-567
[4]  
DU TG, 1988, FENXI HUAXUE, V16, P467
[5]   SELECTIVE CHLORINE DETERMINATION BY GAS-DIFFUSION FLOW-INJECTION ANALYSIS WITH CHEMI-LUMINESCENT DETECTION [J].
GORD, JR ;
GORDON, G ;
PACEY, GE .
ANALYTICAL CHEMISTRY, 1988, 60 (01) :2-4
[6]   DETERMINATION OF CHLORINE IN WATER BY LUMINOL CHEMI-LUMINESCENCE [J].
MARINO, DF ;
INGLE, JD .
ANALYTICAL CHEMISTRY, 1981, 53 (03) :455-458
[7]   RHODAMINE-6G CHEMI-LUMINESCENCE FOR THE DETERMINATION OF FREE CHLORINE IN TAP WATER BY A FLOW-INJECTION METHOD [J].
NAKAGAMA, T ;
YAMADA, M ;
HOBO, T .
ANALYST, 1989, 114 (10) :1275-1278
[8]   CHEMILUMINESCENCE SENSOR WITH URANINE IMMOBILIZED ON AN ANION-EXCHANGE RESIN FOR MONITORING FREE CHLORINE IN TAP WATER [J].
NAKAGAMA, T ;
YAMADA, M ;
HOBO, T .
ANALYTICA CHIMICA ACTA, 1990, 231 (01) :7-12
[9]  
VANDENBERG A, 1993, SENS ACTUATORS B, V13, P369
[10]   DETERMINATION OF FREE AND COMBINED RESIDUAL CHLORINE BY FLOW-INJECTION SPECTROPHOTOMETRY [J].
VERMA, KK ;
JAIN, A ;
TOWNSHEND, A .
ANALYTICA CHIMICA ACTA, 1992, 261 (1-2) :233-240