Comparison and Evaluation of Ocular Biometry Using a New Noncontact Optical Low-Coherence Reflectometer

被引:138
作者
Rohrer, Kaspar [2 ]
Frueh, Beatrice E. [2 ]
Waelti, Rudolf [3 ]
Clemetson, Isabelle A. [2 ]
Tappeiner, Christoph [2 ]
Goldblum, David [1 ]
机构
[1] Univ Basel, Univ Basel Hosp, Dept Ophthalmol, CH-4031 Basel, Switzerland
[2] Univ Bern, Inselspital, Dept Ophthalmol, CH-3010 Bern, Switzerland
[3] Haag Streit AG, Koniz, Switzerland
关键词
ANTERIOR-CHAMBER DEPTH; INTRAOCULAR LENSES; CORNEAL THICKNESS; INTERFEROMETRY; POWER; PACHYMETRY; PRECISION; EYE;
D O I
10.1016/j.ophtha.2009.04.019
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To evaluate a new high-resolution noncontact biometer (Lenstar; Haag-Streit AG, Koeniz, Switzerland) using optical low-coherence reflectometry and to compare the clinical measurements with those obtained from the IOLMaster (Carl Zeiss, Jena, Germany) and the Pach mu meter (Haag-Streit AG). Design: Exploratory evaluation of diagnostic technology and nonrandomized, prospective clinical trial. Participants: Eighty subjects (144 eyes) aged 20 to 90 years with cataractous, pseudophakic, aphakic, silicon oil-filled, or normal eyes. Methods: Measurements of axial length (AL), anterior chamber depth (ACD), central corneal thickness (CCT), corneal radius (R1 [flattest radius of corneal curvature] and R2 [steep radius, 90 degrees apart from R1]), and axis of the flattest radius (Ax1) obtained with the Lenstar were compared with those obtained with the IOLMaster or Pach mu meter. The results were evaluated using Bland-Altman analyses. The differences between both methods were assessed using the paired t test, and its correlation was evaluated by Pearson coefficient. Main Outcome Measures: Axial length, CCT, ACD, R1, R2, and Ax1. Results: The overall mean AL measured with the Lenstar and the IOLMaster was 24.1 mm (r = 0.999). Anterior chamber depth was 3.19 mm (Lenstar) and 3.17 mm (IOLMaster; r = 0.875). Excellent correlations also were found for the corneal radius and the axis of flattest radius (R1, r = 0.927; R2, r = 0.929; and Ax1, r = 0.938). Mean CCT was 0.557 mm (r = 0.978) for both Lenstar and Pach mu meter. Conclusions: Measurements with the new Lenstar correlated well with those with the IOLMaster and Pach mu meter in cataractous, pseudophakic, aphakic, silicon oil-filled, and normal eyes. It is an accurate, fast instrument that provides additional information of interest to any cataract or refractive surgeon. Financial Disclosure(s): Proprietary or commercial disclosure may be found after the references. Ophthalmology 2009;116:2087-2092 (C) 2009 by the American Academy of Ophthalmology.
引用
收藏
页码:2087 / 2092
页数:6
相关论文
共 20 条
[1]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[2]   Real-time pachymetry during photorefractive keratectomy using optical low-coherence reflectometry [J].
Böhnke, M ;
Widmer, S ;
Wälti, R .
JOURNAL OF BIOMEDICAL OPTICS, 2001, 6 (04) :412-417
[3]   Precision and reproducibility of measurements of human corneal thickness with rapid optical low-coherence reflectometry (OLCR) [J].
Böhnke, M ;
Masters, BR ;
Wälti, R ;
Ballif, JJ ;
Chavanne, P ;
Gianotti, R ;
Salathé, RP .
JOURNAL OF BIOMEDICAL OPTICS, 1999, 4 (01) :152-156
[4]  
Bohnke M, 1998, J REFRACT SURG, V14, P140
[5]   Comparison of three methods of measuring corneal thickness and anterior chamber depth [J].
Buehl, W ;
Stojanac, D ;
Sacu, S ;
Drexler, W ;
Findl, O .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 2006, 141 (01) :7-12
[6]   Partial coherence interferometry: A novel approach to biometry in cataract surgery [J].
Drexler, W ;
Findl, O ;
Menapace, R ;
Rainer, G ;
Vass, C ;
Hitzenberger, CK ;
Fercher, AF .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 1998, 126 (04) :524-534
[7]   High precision biometry of pseudophakic eyes using partial coherence interferometry [J].
Findl, O ;
Drexler, W ;
Menapace, R ;
Hitzenberger, CK ;
Fercher, AF .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 1998, 24 (08) :1087-1093
[8]   Optical low coherence reflectometry for noncontact measurements of flap thickness during laser in situ keratomileusis [J].
Genth, U ;
Mrochen, M ;
Wälti, R ;
Salaheldine, MM ;
Seiler, T .
OPHTHALMOLOGY, 2002, 109 (05) :973-978
[9]   Comparison of immersion ultrasound biometry and partial coherence interferometry for intraocular lens calculation according to Haigis [J].
Haigis, W ;
Lege, B ;
Miller, N ;
Schneider, B .
GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2000, 238 (09) :765-773
[10]   REFRACTIVE POWER CALCULATIONS FOR INTRAOCULAR LENSES IN THE PHAKIC EYE [J].
HOLLADAY, JT .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 1993, 116 (01) :63-66