Ceramic Component Fracture: Trends and Recommendations with Modern Components based on improved Reporting Methods

被引:5
作者
Garino, J. P. [1 ]
机构
[1] Dept Orthoped Surg, Philadelphia, PA 19104 USA
来源
BIOCERAMICS AND ALTERNATIVE BEARINGS IN JOINT ARTHROPLASTY | 2005年
关键词
D O I
10.1007/3-7985-1540-9_25
中图分类号
Q81 [生物工程学(生物技术)]; Q93 [微生物学];
学科分类号
071005 ; 0836 ; 090102 ; 100705 ;
摘要
Fractures of ceramic components have been reported in the literature with highly variable rates, many of which include obsolete designs and previous generations of ceramic materials. With improved worldwide reporting since the year 2000, the database of the largest manufacturer of medical grade ceramics was analyzed with respect to the demographics of modern ceramic component fractures. Although the overall fracture rate based on an excess of 1 million components being produced with the latest technology was a safe and low 0.015% and compared favorably with other complications such as infection, stem breakage, dislocation, etc, the analysis revealed many interesting findings. Three of the most significant were: 28 mm ball heads fracture at a substantially higher rate than the 32mm ball heads; 90% of fractures occur within the 36 months of implantation; due to continued quality improvements the rate continues to decrease (actual > expected rates).
引用
收藏
页码:157 / 168
页数:12
相关论文
共 21 条
[1]   Ceramic-on-ceramic bearings in total hip arthroplasty [J].
Bierbaum, BE ;
Nairus, J ;
Kuesis, D ;
Morrison, JC ;
Ward, D .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2002, (405) :158-163
[2]  
Bizot P, 2000, CLIN ORTHOP RELAT R, P134
[3]  
Boehler M, 2000, CLIN ORTHOP RELAT R, P85
[4]  
BONICOLI F, 1997, PERFORMANCE WEAR COU, P38
[5]  
Callaway GH, 1995, J ARTHROPLASTY, V10, P855
[6]   An analysis of Food and Drug Administration medical device reports relating to total joint components [J].
Castro, FP ;
Chimento, G ;
Munn, BG ;
Levy, RS ;
Timon, S ;
Barrack, RL .
JOURNAL OF ARTHROPLASTY, 1997, 12 (07) :765-771
[7]  
D'Antonio J, 2003, ORTHOPEDICS, V26, P39
[8]  
DANTONIO J, 2003, INT SOC TECH ARTHR S
[9]  
Degreif J, 2001, ORTHOPEDICS, V24, P129
[10]  
*FDA, 2003, TRANSC LIN SYST WEIG