Quantification of human high-energy phosphate metabolite concentrations at 3 T with partial volume and sensitivity corrections

被引:23
作者
El-Sharkawy, AbdEl-Monem M. [1 ]
Gabr, Refaat E. [1 ]
Schaer, Michael [1 ,2 ]
Weiss, Robert G. [1 ,3 ]
Bottomley, Paul A. [1 ]
机构
[1] Johns Hopkins Univ, Sch Med, Russell H Morgan Dept Radiol & Radiol Sci, Baltimore, MD 21287 USA
[2] Philips Healthcare, Cleveland, OH USA
[3] Johns Hopkins Univ, Sch Med, Dept Med, Div Cardiol, Baltimore, MD 21287 USA
基金
美国国家卫生研究院;
关键词
quantification; metabolism; phosphorus; high field; human heart; 3T; MAGNETIC-RESONANCE-SPECTROSCOPY; CARDIAC PHOSPHORUS METABOLITES; IN-VIVO; CREATINE-KINASE; HUMAN-HEART; P-31; MRS; MYOCARDIAL-INFARCTION; RELAXATION-TIMES; MUSCLE; TRIGLYCERIDE;
D O I
10.1002/nbm.2961
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Practical noninvasive methods for the measurement of absolute metabolite concentrations are key to the assessment of the depletion of myocardial metabolite pools which occurs with several cardiac diseases, including infarction and heart failure. Localized MRS offers unique noninvasive access to many metabolites, but is often confounded by nonuniform sensitivity and partial volume effects in the large, poorly defined voxels commonly used for the detection of low-concentration metabolites with surface coils. These problems are exacerbated at higher magnetic field strengths by greater radiofrequency (RF) field inhomogeneity and differences in RF penetration with heteronuclear concentration referencing. An example is the P-31 measurement of cardiac adenosine triphosphate (ATP) and phosphocreatine (PCr) concentrations, which, although central to cardiac energetics, have not been measured at field strengths above 1.5T. Here, practical acquisition and analysis protocols are presented for the quantification of [PCr] and [ATP] with one-dimensionally resolved surface coil spectra and concentration referencing at 3T. The effects of nonuniform sensitivity and partial tissue volumes are addressed at 3T by the application of MRI-based three-dimensional sensitivity weighting and tissue segmentation. The method is validated in phantoms of different sizes and concentrations, and used to measure [PCr] and [ATP] in healthy subjects. In calf muscle (n=8), [PCr]=24.7 +/- 3.4 and [ATP]=5.7 +/- 1.3 mu mol/g wet weight, whereas, in heart (n=18), [PCr]=10.4 +/- 1.5 and [ATP]=6.0 +/- 1.1 mu mol/g wet weight (all mean +/- SD), consistent with previous reports at lower fields. The method enables, for the first time, the efficient, semi-automated quantification of high-energy phosphate metabolites in humans at 3T with nonuniform excitation and detection. Copyright (c) 2013 John Wiley & Sons, Ltd.
引用
收藏
页码:1363 / 1371
页数:9
相关论文
共 41 条
[1]   Absolute concentrations of high-energy phosphate metabolites in normal, hypertrophied, and failing human myocardium measured noninvasively with 31P-SLOOP magnetic resonance spectroscopy [J].
Beer, M ;
Seyfarth, T ;
Sandstede, J ;
Landschütz, W ;
Lipke, C ;
Köstler, H ;
von Kienlin, M ;
Harre, K ;
Hahn, D ;
Neubauer, S .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2002, 40 (07) :1267-1274
[3]   Non-invasive magnetic-resonance detection of creatine depletion in non-viable infarcted myocardium [J].
Bottomley, PA ;
Weiss, RG .
LANCET, 1998, 351 (9104) :714-718
[4]   Total creatine in muscle: Imaging and quantification with proton MR spectroscopy [J].
Bottomley, PA ;
Lee, YH ;
Weiss, RG .
RADIOLOGY, 1997, 204 (02) :403-410
[5]   PHOSPHATE METABOLITE IMAGING AND CONCENTRATION MEASUREMENTS IN HUMAN HEART BY NUCLEAR-MAGNETIC-RESONANCE [J].
BOTTOMLEY, PA ;
HARDY, CJ ;
ROEMER, PB .
MAGNETIC RESONANCE IN MEDICINE, 1990, 14 (03) :425-434
[6]   Human cardiac high-energy phosphate metabolite concentrations by 1D-resolved NMR spectroscopy [J].
Bottomley, PA ;
Atalar, E ;
Weiss, RG .
MAGNETIC RESONANCE IN MEDICINE, 1996, 35 (05) :664-670
[7]   ALTERED PHOSPHATE-METABOLISM IN MYOCARDIAL-INFARCTION - P-31 MR SPECTROSCOPY [J].
BOTTOMLEY, PA ;
HERFKENS, RJ ;
SMITH, LS ;
BASHORE, TM .
RADIOLOGY, 1987, 165 (03) :703-707
[8]   AIDS DEMENTIA COMPLEX - BRAIN HIGH-ENERGY PHOSPHATE METABOLITE DEFICITS [J].
BOTTOMLEY, PA ;
HARDY, CJ ;
COUSINS, JP ;
ARMSTRONG, M ;
WAGLE, WA .
RADIOLOGY, 1990, 176 (02) :407-411
[9]   Reduced Myocardial Creatine Kinase Flux in Human Myocardial Infarction An In Vivo Phosphorus Magnetic Resonance Spectroscopy Study [J].
Bottomley, Paul A. ;
Wu, Katherine C. ;
Gerstenblith, Gary ;
Schulman, Steven P. ;
Steinberg, Angela ;
Weiss, Robert G. .
CIRCULATION, 2009, 119 (14) :1918-1924
[10]   ASSESSMENT OF ABSOLUTE METABOLITE CONCENTRATIONS IN HUMAN TISSUE BY P-31 MRS IN-VIVO .2. MUSCLE, LIVER, KIDNEY [J].
BUCHLI, R ;
MEIER, D ;
MARTIN, E ;
BOESIGER, P .
MAGNETIC RESONANCE IN MEDICINE, 1994, 32 (04) :453-458