Monte Carlo simulations of a kilovoltage external beam radiotherapy system on phantoms and breast patients

被引:13
作者
Breitkreutz, Dylan Y. [1 ]
Weil, Michael D. [2 ]
Zavgorodni, Sergei [3 ]
Bazalova-Carter, Magdalena [1 ,3 ]
机构
[1] Univ Victoria, Dept Phys & Astron, POB 1700 ST CSC, Victoria, BC V8W 2Y2, Canada
[2] Sirius Med LLC, POB 414, Half Moon Bay, CA 94019 USA
[3] BC Canc Agcy, Vancouver Isl Ctr, 2410 Lee Ave, Victoria, BC V8R 6V5, Canada
基金
加拿大自然科学与工程研究理事会;
关键词
kilovoltage x-rays; Monte Carlo; partial breast irradiations; FEASIBILITY; VMC++;
D O I
10.1002/mp.12619
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To determine the most suitable lesion size and depth for radiotherapy treatments with a prototype kilovoltage x-ray arc therapy (KVAT) system through Monte Carlo simulations of the dose delivered to lesion, dose homogeneity, and lesion-to-skin ratio. Methods: Monte Carlo simulations were used to calculate dose distributions generated by a novel low-energy kilovoltage x-ray system to a variety of clinically relevant lesion sizes and depths in phantoms and for hypothetical partial breast irradiations of patients in supine and prone positions. The treatments by 200 kV KVAT system were modeled for four sizes of tumor (1-4 cm diameter) at three depths (superficial, middle, and deep) in two sizes of cylindrical water phantoms (16.2-cm and 32.2-cm diameter). In addition, treatments of 3-cm and 4-cm diameter lesions were modeled for two breast patients in prone and supine positions. Dose distributions were calculated using the EGSnrc/DOS-XYZnrc code package. Phantom study metrics included lesion-to-skin ratio, dose delivered to isocenter (cGy/min), dose homogeneity, dose profiles, and cumulative dose volume histograms. Lesion-toskin ratio, lesion-to-rib ratio, dose profiles, and cumulative dose volume histograms were used to evaluate simulated breast patient treatments. Supine breast irradiations were compared to 6-MV VMAT plans. The criterion applied to evaluate the dose distributions was derived from NSABP-B39/ RTOG 0413 for accelerated partial breast irradiation. Skin dose was limited to a maximum of 250 cGy for a prescribed lesion dose of 385 cGy per fraction (with the whole treatment being delivered in 10 fractions). This produced the minimum lesion-to-skin dose ratio of 1.5 that served as the main guideline, along with other metrics, for evaluation of future clinical viability of treatments. Results: Phantom dose distributions in the centrally located lesions treated with 360-degree KVAT were found to be superior to dose distributions in off-center lesions with the exception of isocenter dose, which was highest for lesions located closer to the phantom surface. Dose metrics were more favorable for smaller lesions, suggesting that KVAT might be most suitable for treatment of lesions of 1-2 cm in diameter down to depths of 8.1 cm along with 3 cm lesions at depths from 3 cm to 8.1 cm. In addition, treatments of 4-cm lesions were found to be acceptable down to the depths of 4.1 cm (in the 16.2-cm phantom) and 8.1 cm (in the 32.2-cm phantom). At depths from 8.1-cm to 16.1-cm, treatments of 1-cm to 4-cm lesions are possible at the cost of decreased dose rate. KVAT breast treatments in the supine patient position demonstrated that increasing the arc angle and decreasing lesion size improved lesion-to-skin ratio and lesion-to-rib ratio. Supine breast data indicate that 3-cm lesions are treatable at a minimum depth of 3 cm. The 6-MV VMAT plan resulted in lower doses to the ipsilateral lung and the body, but a higher heart dose compared to the KVAT plans. Dose distributions for the prone breast phantoms were superior to the supine cases due to the increased treatment angle of 360-degrees. Conclusions: Although nonoptimized KVAT dose distributions presented here were of inferior quality to VMAT plans, this work has demonstrated the feasibility of delivering low-energy kilovoltage xrays to lesions up to 4 cm in diameter to depths of 8.1 cm while sparing surrounding tissue. (C) 2017 American Association of Physicists in Medicine
引用
收藏
页码:6548 / 6559
页数:12
相关论文
共 23 条
  • [1] Efficiency improvements of x-ray simulations in EGSnrc user-codes using bremsstrahlung cross-section enhancement (BCSE)
    Ali, E. S. M.
    Rogers, D. W. O.
    [J]. MEDICAL PHYSICS, 2007, 34 (06) : 2143 - 2154
  • [2] [Anonymous], 0319 RTOG
  • [3] Radiation-induced rib fracture after stereotactic body radiotherapy with a total dose of 54-56 Gy given in 9-7 fractions for patients with peripheral lung tumor: impact of maximum dose and fraction size
    Aoki, Masahiko
    Sato, Mariko
    Hirose, Katsumi
    Akimoto, Hiroyoshi
    Kawaguchi, Hideo
    Hatayama, Yoshiomi
    Ono, Shuichi
    Takai, Yoshihiro
    [J]. RADIATION ONCOLOGY, 2015, 10
  • [4] Monte Carlo model of the scanning beam digital x-ray (SBDX) source
    Bazalova, M.
    Weil, M. D.
    Wilfley, B.
    Graves, E. E.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2012, 57 (22) : 7381 - 7394
  • [5] Feasibility of external beam radiation therapy to deep-seated targets with kilovoltage x-rays
    Bazalova-Carter, Magdalena
    Weil, Michael D.
    Breitkreutz, Dylan Yamabe
    Wilfley, Brian P.
    Graves, Edward E.
    [J]. MEDICAL PHYSICS, 2017, 44 (02) : 597 - 607
  • [6] Towards evidence-based guidelines for radiotherapy infrastructure and staffing needs in Europe: the ESTRO QUARTS project
    Bentzen, SM
    Heeren, G
    Cottier, B
    Slotman, B
    Glimelius, B
    Lievens, Y
    van den Bogaert, W
    [J]. RADIOTHERAPY AND ONCOLOGY, 2005, 75 (03) : 355 - 365
  • [7] Chin LS, 2015, PRINCIPLES PRACTICE, P41
  • [8] Trajectory modulated prone breast irradiation: A LINAC-based technique combining intensity modulated delivery and motion of the couch
    Fahimian, Benjamin
    Yu, Victoria
    Horst, Kathleen
    Xing, Lei
    Hristov, Dimitre
    [J]. RADIOTHERAPY AND ONCOLOGY, 2013, 109 (03) : 475 - 481
  • [9] RT_Image: An open-source tool for investigating PET in radiation oncology
    Graves, Edward E.
    Quon, Andrew
    Loo, Billy W., Jr.
    [J]. TECHNOLOGY IN CANCER RESEARCH & TREATMENT, 2007, 6 (02) : 111 - 121
  • [10] VMC++ versus BEAMnrc: A comparison of simulated linear accelerator heads for photon beams
    Hasenbalg, F.
    Fix, M. K.
    Born, E. J.
    Mini, R.
    Kawrakow, I.
    [J]. MEDICAL PHYSICS, 2008, 35 (04) : 1521 - 1531