The Influence of Head Size on Corrosion and Fretting Behaviour at the Head-Neck Interface of Artificial Hip Joints

被引:142
作者
Dyrkacz, Richard M. R. [1 ]
Brandt, Jan-M. [1 ,2 ]
Ojo, Olanrewaju A. [1 ]
Turgeon, Thomas R. [2 ]
Wyss, Urs P. [1 ,2 ]
机构
[1] Univ Manitoba, Dept Mech & Mfg Engn, Winnipeg, MB R3T 5V6, Canada
[2] Concordia Hip & Knee Inst, Concordia Joint Replacement Grp, Winnipeg, MB R2K 2M9, Canada
关键词
corrosion; fretting; head size; modular; hip; METAL-ON-METAL; JOHN CHARNLEY; ION LEVELS; FOLLOW-UP; ARTHROPLASTY; PROSTHESES; DIAMETER; REPLACEMENTS; PSEUDOTUMOR; COBALT;
D O I
10.1016/j.arth.2012.10.017
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
The primary goal of this study was to determine if head size affects corrosion and fretting behaviour at the head-neck taper interface of modular hip prostheses. Seventy-four implants were retrieved that featured either a 28 mm or a 36 mm head with a metal-on-polyethylene articulation. The bore of the heads and the neck of the stems were divided into eight regions each and graded by three observers for corrosion and fretting damage separately using modified criteria as reported in the literature. The 36 mm head size featured a significant difference in the corrosion head scores (p=0.022) in comparison to the 28 mm heads. This may be attributed to a greater torque acting along the taper interface due to activities of daily living. (C) 2013 Elsevier Inc. All rights reserved.
引用
收藏
页码:1036 / 1040
页数:5
相关论文
共 24 条
[11]  
Goldberg JR, 2002, CLIN ORTHOP RELAT R, P149
[12]   Differences in the fretting corrosion of metal-metal and ceramic-metal modular junctions of total hip replacements [J].
Hallab, NJ ;
Messina, C ;
Skipor, A ;
Jacobs, JJ .
JOURNAL OF ORTHOPAEDIC RESEARCH, 2004, 22 (02) :250-259
[13]  
JACOBS JJ, 1995, CLIN ORTHOP RELAT R, P94
[14]   Corrosion of a Hip Stem With a Modular Neck Taper Junction A Retrieval Study of 16 Cases [J].
Kop, Alan M. ;
Swarts, Eric .
JOURNAL OF ARTHROPLASTY, 2009, 24 (07) :1019-1023
[15]   Taper junction failure in large-diameter metal-on-metal bearings [J].
Langton, D. J. ;
Sidaginamale, R. ;
Lord, J. K. ;
Nargol, A. V. F. ;
Joyce, T. J. .
BONE & JOINT RESEARCH, 2012, 1 (04) :56-63
[16]   Comparison of Whole-Blood Metal Ion Levels in Four Types of Metal-on-Metal Large-Diameter Femoral Head Total Hip Arthroplasty: The Potential Influence of the Adapter Sleeve [J].
Lavigne, Martin ;
Belzile, Etienne L. ;
Roy, Alain ;
Morin, Francois ;
Amzica, Traian ;
Vendittoli, Pascal-Andre .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2011, 93A :128-136
[17]  
LIEBERMAN JR, 1994, CLIN ORTHOP RELAT R, P162
[18]   Clinical Outcome Study and Radiological Findings of Zweymuller Metal on Metal Total Hip Arthroplasty. A Follow-up of 6 to 15 Years [J].
Paleochorlidis, Ilias S. ;
Badras, Leonidas S. ;
Skretas, Efstathios F. ;
Georgaklis, Vasilios A. ;
Karachalios, Theofilos S. ;
Malizos, Konstantinos N. .
HIP INTERNATIONAL, 2009, 19 (04) :301-308
[19]   Pseudotumours associated with metal-onmetal hip resurfacings [J].
Pandit, H. ;
Glyn-Jones, S. ;
McLardy-Smith, P. ;
Gundle, R. ;
Whitwell, D. ;
Gibbons, C. L. M. ;
Ostlere, S. ;
Athanasou, N. ;
Gill, H. S. ;
Murray, D. W. .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2008, 90B (07) :847-851
[20]   Influence of the clearance on in-vitro tribology of large diameter metal-on-metal articulations pertaining to resurfacing hip implants [J].
Rieker, CB ;
Schön, R ;
Konrad, R ;
Liebentritt, G ;
Gnepf, P ;
Shen, M ;
Roberts, P ;
Grigoris, P .
ORTHOPEDIC CLINICS OF NORTH AMERICA, 2005, 36 (02) :135-+