Direct Measurement of Mammographic X-Ray Spectra with a Digital CdTe Detection System

被引:23
作者
Abbene, Leonardo [1 ]
Gerardi, Gaetano [1 ]
Principato, Fabio [1 ]
Del Sordo, Stefano [2 ]
Raso, Giuseppe [1 ]
机构
[1] Univ Palermo, Dipartimento Fis, I-90128 Palermo, Italy
[2] INAF IASF Palermo, I-90146 Palermo, Italy
关键词
X-ray spectroscopy; high photon counting rate; CdTe detectors; digital pulse processing; mammography; SPECTROSCOPY; MOLYBDENUM; COMPUTATION; PROGRESS;
D O I
10.3390/s120608390
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
In this work we present a detection system, based on a CdTe detector and an innovative digital pulse processing (DPP) system, for high-rate X-ray spectroscopy in mammography (1-30 keV). The DPP system performs a height and shape analysis of the detector pulses, sampled and digitized by a 14-bit, 100 MHz ADC. We show the results of the characterization of the detection system both at low and high photon counting rates by using monoenergetic X-ray sources and a nonclinical X-ray tube. The detection system exhibits excellent performance up to 830 kcps with an energy resolution of 4.5% FWHM at 22.1 keV. Direct measurements of clinical molybdenum X-ray spectra were carried out by using a pinhole collimator and a custom alignment device. A comparison with the attenuation curves and the half value layer values, obtained from the measured and simulated spectra, from an ionization chamber and from a solid state dosimeter, also shows the accuracy of the measurements. These results make the proposed detection system a very attractive tool for both laboratory research, calibration of dosimeters and advanced quality controls in mammography.
引用
收藏
页码:8390 / 8404
页数:15
相关论文
共 47 条
[1]   Spectroscopic response of a CdZnTe multiple electrode detector [J].
Abbene, L. ;
Del Sordo, S. ;
Fauci, F. ;
Gerardi, G. ;
La Manna, A. ;
Raso, G. ;
Cola, A. ;
Perillo, E. ;
Raulo, A. ;
Gostilo, V. ;
Stumbo, S. .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2007, 583 (2-3) :324-331
[2]  
Abbene L, 2011, IEEE NUCL SCI CONF R, P4636, DOI 10.1109/NSSMIC.2011.6154749
[3]   Performance enhancements of compound semiconductor radiation detectors using digital pulse processing techniques [J].
Abbene, L. ;
Gerardi, G. .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2011, 654 (01) :340-348
[4]   High-rate x-ray spectroscopy in mammography with a CdTe detector: A digital pulse processing approach [J].
Abbene, L. ;
Gerardi, G. ;
Principato, F. ;
Del Sordo, S. ;
Ienzi, R. ;
Raso, G. .
MEDICAL PHYSICS, 2010, 37 (12) :6147-6156
[5]   Performance of a digital CdTe X-ray spectrometer in low and high counting rate environment [J].
Abbene, L. ;
Gerardi, G. ;
Del Sordo, S. ;
Raso, G. .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2010, 621 (1-3) :447-452
[6]   Hard x-ray response of pixellated CdZnTe detectors [J].
Abbene, L. ;
Del Sordo, S. ;
Caroli, E. ;
Gerardi, G. ;
Raso, G. ;
Caccia, S. ;
Bertuccio, G. .
JOURNAL OF APPLIED PHYSICS, 2009, 105 (12)
[7]  
Abbene L., 2008, IEEE NUCL SCI S, P478
[8]   X-ray spectroscopy and dosimetry with a portable CdTe device [J].
Abbene, Leonardo ;
La Manna, Angelo ;
Fauci, Francesco ;
Gerardi, Gaetano ;
Stumbo, Simone ;
Raso, Giuseppe .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2007, 571 (1-2) :373-377
[9]  
[Anonymous], 2007, 457 IAEA TRS
[10]   Detectors for the future of X-ray imaging [J].
Aslund, M. ;
Fredenberg, E. ;
Telman, M. ;
Danielsson, M. .
RADIATION PROTECTION DOSIMETRY, 2010, 139 (1-3) :327-333