Analytical performance of a new two-step ADVIA Centaur® estradiol immunoassay during ovarian stimulation

被引:3
作者
Massart, C
Gibassier, J
Laurent, MC
Le Lannou, D
机构
[1] CHU Pontchaillou, Lab Gent Mol & Hormonol, F-35043 Rennes, France
[2] Univ Rennes 1, UPRES, EA 3889, Rennes, France
[3] CHU Pontchaillou, Serv Gynecol & Obstet, Rennes, France
[4] CHU Pontchaillou, Lab Biol Reprod Cecos, Rennes, France
关键词
estradiol; immunoassay; ovarian stimulation; sensitivity;
D O I
10.1515/CCLM.2006.020
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Measurement of estradiol is useful in women undergoing ovarian stimulation for in vitro fertilization and embryo transfer (IVF-ET). The analytical performance of a new two-step estradiol assay (ADVIA Centaur (R) estradiol-6 III from Bayer Diagnostics) was evaluated in 41 sera from 11 women undergoing ovarian stimulation. The results were compared to those obtained with two radioimmunoassays (RIAs; RIA Estradiol Immunotech IM 1663 from Beckmann Coulter and Coat-A-Count Estradiol from Diagnostic Products Corporation) and with one chemiluminescent immunoassay (CLIA; ADVIA Centaur (R) estradiol-6). The ADVIA Centaur (R) estradiol-6 III assay was the most sensitive assay, with a functional sensitivity of 55 pmol/L. Within- and between- run coefficients of variation calculated for the new ADVIA Centaur (R) assay ranged from 3.3% to 9%, which was better than the precision obtained for the other assays. A dilution test showed serum interferences when estradiol was measured in non- diluted samples. No statistical difference was observed between the estradiol results obtained in diluted sera with the new two-step ADVIA Centaur (R) assay and those measured with the Immunotech RIA and the other CLIA. In conclusion, this new, two-step estradiol assay performed on the ADVIA Centaur (R) system displays suitable sensitivity, precision and intermethod agreement with the Immunotech RIA for the measurement of serum estradiol concentrations in women undergoing ovarian stimulation and IVF-ET. For correct linearity, estradiol measurement should be performed on diluted samples.
引用
收藏
页码:105 / 109
页数:5
相关论文
共 16 条
[1]   Reference intervals for serum sex steroids and gonadotropins in regularly menstruating women [J].
Anttila, Leena ;
Koskinen, Pertti ;
Irjala, Kerttu ;
Kaihola, Hanna-Leena .
ACTA OBSTETRICIA ET GYNECOLOGICA SCANDINAVICA, 1991, 70 (06) :475-481
[2]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[3]   PHILOSOPHY OF MEASUREMENT BY MEANS OF IMMUNOASSAYS [J].
BUTTNER, J .
SCANDINAVIAN JOURNAL OF CLINICAL & LABORATORY INVESTIGATION, 1991, 51 :11-20
[4]  
FOSSUM GT, 1990, OBSTET GYNECOL, V75, P407
[5]  
HUGHES EG, 1992, FERTIL STERIL, V58, P888
[6]  
LEE CS, 1991, J REPROD MED, V36, P156
[7]   Results of serial measurement of estradiol in serum with six different methods during ovarian stimulation [J].
Mikkelsen, AL ;
Borggaard, B ;
Lebech, PE .
GYNECOLOGIC AND OBSTETRIC INVESTIGATION, 1996, 41 (01) :35-40
[8]  
POTISCHMAN N, 1994, CANCER RES, V54, P5363
[9]  
SCHIOLER V, 1988, CLIN CHEM, V34, P949
[10]  
Spencer CA, 1996, CLIN CHEM, V42, P140