Precision (repeatability and reproducibility) of ocular parameters obtained by the Tomey OA-2000 biometer compared to the IOLMaster in healthy eyes

被引:38
作者
Hua, Yanjun [1 ]
Qiu, Wei [1 ]
Xiao, Qiuyi [1 ]
Wu, Qiang [1 ]
机构
[1] Shanghai Jiao Tong Univ, Affiliated Peoples Hosp 6, Dept Ophthalmol, Shanghai, Peoples R China
关键词
PARTIAL-COHERENCE INTERFEROMETRY; STATISTICAL-METHODS; CATARACT EYES; AXIAL LENGTH; AGREEMENT; MULTICENTER; TOMOGRAPHY; COMPONENTS; DEVICES; ERROR;
D O I
10.1371/journal.pone.0193023
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
Purpose To assess the precision (repeatability and reproducibility) of ocular parameters measured by the Tomey OA-2000 biometer, and to compare them with those measured by the IOLMaster. Methods In this prospective study, the right eyes of 108 healthy subjects were included. Three consecutive scans were obtained by 2 observers using the Tomey OA-2000, and in the same session one observer used the IOLMaster (version 5.4.4.0006) for the measurements. About 1 week later, 3 scans were obtained by one observer using the Tomey OA-2000. The axial length (AL), central corneal thickness (CCT), anterior chamber depth (ACD), lens thickness (LT), keratometer readings, pupil diameter (PD) and corneal diameter (CD) values measured by the Tomey OA-2000 and IOLMaster were analyzed. The coefficient of variation (CoV), intraclass correlation coefficient (ICC), within subject standard deviation (Sw) and 2.77Sw were calculated to assess the repeatability and reproducibility. The paired t test and Bland-Altman plots were used to analyze the differences and agreements of parameters measured by the two devices, respectively. Results Intraobserver repeatability, and interobserver and intersession reproducibility of the AL, COT, ACD, LT, Kf, Ks, Kin, PD and CID values measured by the Tomey OA-2000 biometer showed a CoV of less than 1% except that for PD, and an ICC of more than 0.97 except that for PD and CD. The AL, Kf, Ks, Km and CD values measured by the Tomey OA-2000 were 0.058 +/- 0.094 mm, 0.088 +/- 0.150 diopters (D), 0.163 +/- 0.170 D, 0.127 +/- 0.117 D and 0.171 +/- 0.217 mm lower than those measured by the IOLMaster, respectively (all Ps < 0.05). However, the ACID values from the two devices were comparable (P = 0.169). The 95% linite of agreement (LoA) of the AL, ACD, CD and all keratometer readings were no more than 0.24 mm, 0.14 mm 0.60 mm and 0.5 D, respectively. Conclusion Except for the PD and CD, the ocular parameters measured by the Tomey OA-2000 were highly repeatable and reproducible. Except for the CD value, there was good agreement of ocular parameters measured by the Tomey OA-2000 and the IOLMaster in healthy eyes.
引用
收藏
页数:11
相关论文
共 35 条
[1]  
Bland JM, 1996, BRIT MED J, V313, P744
[2]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[3]   A new optical low coherence reflectometry device for ocular biometry in cataract patients [J].
Buckhurst, P. J. ;
Wolffsohn, J. S. ;
Shah, S. ;
Naroo, S. A. ;
Davies, L. N. ;
Berrow, E. J. .
BRITISH JOURNAL OF OPHTHALMOLOGY, 2009, 93 (07) :949-953
[4]   Relationship among Corneal Biomechanics, Refractive Error, and Axial Length [J].
Bueno-Gimeno, Inmaculada ;
Espana-Gregori, Enrique ;
Gene-Sampedro, Andres ;
Lanzagorta-Aresti, Aitor ;
Pinero-Llorens, David P. .
OPTOMETRY AND VISION SCIENCE, 2014, 91 (05) :507-513
[5]   The Cataract National Dataset electronic multicentre audit of 55 567 operations: when should IOLMaster biometric measurements be rechecked? [J].
Cartwright, N. E. Knox ;
Johnston, R. L. ;
Jaycock, P. D. ;
Tole, D. M. ;
Sparrow, J. M. .
EYE, 2010, 24 (05) :894-900
[6]   Evaluation of the Lenstar LS 900 non-contact biometer [J].
Cruysberg, L. P. J. ;
Doors, M. ;
Verbakel, F. ;
Berendschot, T. T. J. M. ;
De Brabander, J. ;
Nuijts, R. M. M. A. .
BRITISH JOURNAL OF OPHTHALMOLOGY, 2010, 94 (01) :106-110
[7]   Keratometry obtained by corneal mapping versus the IOLMaster in the prediction of postoperative refraction in routine cataract surgery [J].
Dulku, Simon ;
Smith, Henry B. ;
Antcliff, Richard J. .
CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2013, 41 (01) :12-18
[8]   The impact of cataract severity on measurement acquisition with the IOLMaster [J].
Freeman, G ;
Pesudovs, K .
ACTA OPHTHALMOLOGICA SCANDINAVICA, 2005, 83 (04) :439-442
[9]   Comparison of 3 biometry devices in cataract patients [J].
Goebels, Susanne ;
Pattmoeller, Max ;
Eppig, Timo ;
Cayless, Alan ;
Seitz, Berthold ;
Langenbucher, Achim .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2015, 41 (11) :2387-2393
[10]   Reproducibility of the Optical Biometer OA-1000 (Tomey) [J].
Goebels, Susanne Christiane ;
Seitz, Berthold ;
Langenbucher, Achim .
BIOMED RESEARCH INTERNATIONAL, 2014, 2014