Intercomparison of phosphorus-32, technetium-99m and iodine-131 activity measurement in Pakistani nuclear medical centers and supplier

被引:1
作者
Abbas, S. [2 ]
Mushtaq, A. [1 ]
Qazi, Q. M. [1 ]
机构
[1] Pakistan Inst Nucl Sci & Technol PINSTECH, Isotope Prod Div, Islamabad, Pakistan
[2] LINAR, Larkana, Sind, Pakistan
关键词
Activity meter; Patient dose; Iodine-131; Phosphorus-32; Technetium-99m; Elution efficiency; RADIOPHARMACEUTICALS;
D O I
10.1007/s10967-012-1797-8
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Technetium-99m and iodine-131 are the leading isotopes used in diagnosis and radiotherapy, respectively in Pakistani nuclear medical centers. Both radionuclides are regularly supplied by Isotope Production Division (IPD), PINSTECH to nuclear medical centers in Pakistan. The packages of P-32, Mo-99 -> Tc-99m generators, and I-131, are prepared at IPD and dispatched by air, rail or road according to the desired activity at reference day to nuclear medical centers. An accurate determination of the activity to be administered to the patient plays key role on the quality of the clinical result of a medical investigation. Activity meters are employed for quantification of radioactivity of radionuclides at supplier site as well as by the users. In order to investigate the performance on the quality of activity measurements and elution efficiency of PAKGEN Tc-99m generators, a pilot program of intercomparison was organized between the supplier and the users. The results of intercomparison of the activity measurements showed a maximum difference of +/- 13 % between supplier and nuclear medical centers.
引用
收藏
页码:197 / 203
页数:7
相关论文
共 17 条
  • [1] [Anonymous], 1991, IAEATECDOC602
  • [2] Calibration of the LNMRI Secondary Standard Ionization Chamber for 131I capsules used in nuclear medicine
    Chaves, Taina Olivieri
    Iwahara, Akira
    Tauhata, Luiz
    Rezende, Eduarda Alexandre
    Correia, Amanda Ribeiro
    de Oliveira, Estela Maria
    [J]. JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY, 2012, 293 (01) : 211 - 216
  • [3] Activity characterization of 192Ir brachytherapy wires
    Damasceno, A. L. O.
    Iwahara, A.
    Silva, M. A. L.
    Estrada, J. J. S.
    [J]. JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY, 2006, 269 (02) : 317 - 323
  • [4] European Pharmacopoeia, 2002, EUR DIR QUAL MED EDQ
  • [5] Radiopharmaceuticals in nuclear medicine diagnostic procedures and the problem of radiation protection
    Ftacnikova, S
    [J]. JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY-ARTICLES, 1996, 209 (02): : 355 - 360
  • [6] Gansbeke BV, 1985, J RADIOANAL NUCL CHE, V92, P323
  • [7] *INT ATOM EN AG, 2006, TECHN REP SER INT AT, V454
  • [8] Proficiency test for radioactivity measurements in nuclear medicine
    Iwahara, A.
    Tauhata, L.
    Oliveira, A. E.
    Nicoli, I. G.
    Alabarse, F. G.
    Xavier, A. M.
    Oliveira, M. L.
    Koskinas, M. F.
    Almeida, M. C. M.
    [J]. JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY, 2009, 281 (01) : 3 - 6
  • [9] Labelled compounds as radiopharmaceuticals for radiosynoviorthesis
    Melichar, F.
    Kropacek, M.
    Srank, J.
    Beran, M.
    Mirzajevova, M.
    Zimova, J.
    Eigner-Henke, K.
    Forsterova, M.
    [J]. JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY, 2009, 280 (02) : 353 - 358
  • [10] MRA, 1999, MUT REC NAT MEAS STA