Effect of hoop stress fracture on micromotion of textured ingrowth stems for radial head replacement

被引:16
作者
Chanlalit, Cholawish [1 ,2 ]
Shukla, Dave R. [1 ]
Fitzsimmons, James S. [1 ]
An, Kai-Nan [1 ]
O'Driscoll, Shawn W. [1 ]
机构
[1] Mayo Clin, Dept Orthoped Surg, Biomech Lab, Rochester, MN 55905 USA
[2] Srinakharinwirot Univ, HRH Princess Maha Chakri Sirindhorn Med Ctr, Fac Med, Dept Orthoped, Bangkok, Thailand
关键词
Radial head prosthesis; fracture; hoop stress; elbow stability; micromotion; TOTAL HIP-ARTHROPLASTY; FEMORAL FRACTURES; BONE; IMPLANTS; MANAGEMENT; FEMUR;
D O I
10.1016/j.jse.2011.05.001
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: Successful bone ingrowth around cementless implants requires adequate initial stability. Hoop stress fractures during stem insertion can potentially hinder prosthesis stability. Hypothesis: We hypothesized that an oversized radial head prosthetic stem (1 mm "too large" and causing a hoop stress fracture during insertion) would result in an unacceptable amount of micromotion. Materials and methods: Grit-blasted radial head prosthetic stems were implanted into cadaveric radii. Rasp and stem insertion energies were measured, along with micromotion at the stem tip. The sizes were increased until a fracture developed in the radial neck. Results: Prosthetic radial head stems that were oversized by 1 mm caused small cracks in the radial neck. Micromotion of oversized stems (42 +/- 7 mu m) was within the threshold conducive for bone ingrowth (<100 mu m) and not significantly different from that for the maximum sized stems (50 +/- 12 mu m) (P >= .4). Discussion: Contrary to our hypothesis, hoop stress fractures caused by implantation of a stem oversized by 1 mm did not result in loss of stability. Stem micromotion remained within the range for bone ingrowth and was not significantly diminished after the fracture. This suggests that if a crack occurs during the final stages of stem insertion, it may be acceptable to leave the stem in place without adding a cerclage wire. Conclusion: A small radial neck fracture occurring during insertion of a radial head prosthetic stem oversized by 1 mm does not necessarily compromise initial stability. Level of evidence: Basic Science Study, Biomechanical Study. (C) 2012 Journal of Shoulder and Elbow Surgery Board of Trustees.
引用
收藏
页码:949 / 954
页数:6
相关论文
共 12 条
[1]  
BOBYN JD, 1980, CLIN ORTHOP RELAT R, P263
[2]   OSTEOGENIC PHENOMENA ACROSS ENDOSTEAL BONE-IMPLANT SPACES WITH POROUS SURFACED INTRA-MEDULLARY IMPLANTS [J].
BOBYN, JD ;
PILLIAR, RM ;
CAMERON, HU ;
WEATHERLY, GC .
ACTA ORTHOPAEDICA SCANDINAVICA, 1981, 52 (02) :145-153
[3]   Micromotion of plasma spray versus grit-blasted radial head prosthetic stem surfaces [J].
Chanlalit, Cholawish ;
Fitzsimmons, James S. ;
Shukla, Dave R. ;
An, Kai-Nan ;
O'Driscoll, Shawn W. .
JOURNAL OF SHOULDER AND ELBOW SURGERY, 2011, 20 (05) :717-722
[4]  
CHRISTENSEN CM, 1989, CLIN ORTHOP RELAT R, P177
[5]   Intraoperative periprosthetic fractures during total hip arthroplasty - Evaluation and management [J].
Davidson, Darin ;
Pike, Jeffrey ;
Garbuz, Donald ;
Duncan, Clive P. ;
Masri, Bassam A. .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2008, 90A (09) :2000-2012
[6]   Radial head arthroplasty with a modular metal spacer to treat acute traumatic elbow instability [J].
Doornberg, Job N. ;
Parisien, Robert ;
van Duijn, P. Joppe ;
Ring, David .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2007, 89A (05) :1075-1080
[7]  
ENGH CA, 1992, CLIN ORTHOP RELAT R, P13
[8]   Femoral fractures in revision hip surgeries with impacted bone allograft [J].
Farfalli, German L. ;
Buttaro, Martin A. ;
Piccaluga, Francisco .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2007, (462) :130-136
[9]  
FITZGERALD RH, 1988, CLIN ORTHOP RELAT R, P61
[10]   Intraoperative fracture of the femur in revision total hip arthroplasty with a diaphyseal fitting stem [J].
Meek, RMD ;
Garbuz, DS ;
Masri, BA ;
Greidanus, NV ;
Duncan, CP .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2004, 86A (03) :480-485