Protein C Assay Performance An Analysis of North American Specialized Coagulation Laboratory Association Proficiency Testing Results

被引:17
作者
Baron, Jason M. [1 ]
Johnson, Stephen M. [2 ]
Ledford-Kraemer, Marlies R. [3 ]
Hayward, Catherine P. [4 ,5 ]
Meijer, Piet [6 ]
Van Cott, Elizabeth M. [1 ]
机构
[1] Harvard Univ, Dept Pathol, Massachusetts Gen Hosp, Sch Med, Boston, MA 02114 USA
[2] Tufts Med Ctr, Dept Pathol, Boston, MA USA
[3] CLOT ED, Islamorada, FL USA
[4] McMaster Univ, Dept Pathol & Mol Med, Hamilton, ON, Canada
[5] McMaster Univ, Dept Med, Hamilton, ON, Canada
[6] ECAT Fdn, Leiden, Netherlands
关键词
Protein C; Quality assurance; Hypercoagulability;
D O I
10.1309/AJCP8MWU4QSTCLPU
中图分类号
R36 [病理学];
学科分类号
100104 ;
摘要
To determine the performance and frequency of protein C reagents currently used by clinical laboratories, we analyzed North American Specialized Coagulation Laboratory Association (NASCOLA) protein C proficiency testing data from 6 surveys conducted in 2009 and 2010 (2009-1 to 2009-3 and 2010-1 to 2010-3). Interlaboratory coefficients of variation (CV) for commonly used reagents on a survey with normal protein C ranged from 8% to 12% for antigenic assays, from 4% to 7% for chromogenic activity assays, and from 7% to 22% for clot-based activity assays. CVs for commonly used reagents on specimens with abnormal protein C ranged from 15% to 24% for antigenic, 4% to 11% for chromogenic, and 10% to 17% for clot-based assays (averaged across 3 surveys). Some reagents were used by relatively few laboratories and therefore additional study may be needed for those reagents. For all commonly used reagents, biases were usually small and often not statistically significant. All assessed reagents were clinically accurate, and were considered acceptable options for a specialized coagulation laboratory.
引用
收藏
页码:909 / 915
页数:7
相关论文
共 8 条
[1]  
[Anonymous], 2008, ALGORITHMIC APPROACH
[2]   Marked discrepancy between coagulometric Protein C activity assays with the pro-thrombotic Protein C Asn2Ile substitution [J].
Cooper, P. C. ;
Siddiq, S. ;
Morse, C. ;
Nightingale, J. ;
Mumford, A. D. .
INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2011, 33 (05) :451-456
[3]  
Eby CS, 2008, ALGORITHMIC APPROACH, P125
[4]   Laboratory tests for protein C deficiency [J].
Khor, Bernard ;
Van Cott, Elizabeth M. .
AMERICAN JOURNAL OF HEMATOLOGY, 2010, 85 (06) :440-442
[5]   Laboratory Evaluation of Hypercoagulability [J].
Khor, Bernard ;
Van Cott, Elizabeth M. .
CLINICS IN LABORATORY MEDICINE, 2009, 29 (02) :339-+
[6]  
Lefkowitz J.B., 2008, An Algorithmic Approach to Hemostasis Testing, P3
[7]   The long-term within- and between-laboratory variability for assay of antithrombin, and proteins C and S: results derived from the external quality assessment program for thrombophilia screening of the ECAT Foundation [J].
Meijer, P ;
Kluft, C ;
Haverkate, F ;
De Maat, MPM .
JOURNAL OF THROMBOSIS AND HAEMOSTASIS, 2003, 1 (04) :748-753
[8]   Protein S assays - An analysis of North American Specialized Coagulation Laboratory Association proficiency testing [J].
Van Cott, EM ;
Ledford-Kraemer, M ;
Meijer, P ;
Nichols, WL ;
Johnson, SM ;
Peerschke, EIB .
AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 2005, 123 (05) :778-785