Does implantoplasty affect the failure strength of narrow and regular diameter implants? A laboratory study

被引:22
作者
Bertl, Kristina [1 ,2 ]
Isidor, Flemming [3 ]
von Steyern, Per Vult [4 ]
Stavropoulos, Andreas [1 ,5 ,6 ]
机构
[1] Univ Malmo, Fac Odontol, Dept Periocdontol, Malmo, Sweden
[2] Med Univ Vienna, Univ Clin Dent, Div Oral Surg, Vienna, Austria
[3] Aarhus Univ, Sect Prosthet Dent, Dept Dent, Aarhus, Denmark
[4] Univ Malmo, Fac Odontol, Dept Dent Mat Sci & Technol, Malmo, Sweden
[5] Med Univ Vienna, Univ Clin Dent, Div Conservat Dent & Periodontol, Vienna, Austria
[6] CUMD Univ Geneva, Div Regenerat Dent Med & Periodontol, Geneva, Switzerland
关键词
Implantoplasty; Dental implant; Implant failure; In vitro laboratory study; Mechanical complication; Dynamic loading; COMBINED SURGICAL THERAPY; 3-YEAR CLINICAL-TRIAL; SHAPED ORAL IMPLANTS; EVALUATING; METHODS; MAXIMUM BITE FORCE; PERI-IMPLANTITIS; FOLLOW-UP; EDENTULOUS PATIENTS; CONNECTION DESIGNS; RESECTIVE SURGERY;
D O I
10.1007/s00784-020-03534-8
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Objective To assess whether the impact of implantoplasty (IP) on the maximum implant failure strength depends on implant type/design, diameter, or material. Methods Fourteen implants each of different type/design [bone (BL) and tissue level (TL)], diameter [narrow (3.3 mm) and regular (4.1 mm)], and material [titanium grade IV (Ti) and titanium-zirconium alloy (TiZr)] of one company were used. Half of the implants were subjected to IP in a computerized torn. All implants were subjected to dynamic loading prior to loading until failure to simulate regular mastication. Multiple linear regression analyses were performed with maximum implant failure strength as dependent variable and IP, implant type/design, diameter, and material as predictors. Results Implants subjected to IP and TL implants showed statistically significant reduced implant failure strength irrespective of the diameter compared with implants without IP and BL implants, respectively. Implant material had a significant impact for TL implants and for regular diameter implants, with TiZr being stronger than Ti. During dynamic loading, 1 narrow Ti TL implant without IP, 4 narrow Ti TL implants subjected to IP, and 1 narrow TiZr TL implant subjected to IP were fractured. Conclusion IP significantly reduced the maximum implant failure strength, irrespective implant type/design, diameter, or material, but the maximum implant failure strength of regular diameter implants and of narrow BL implants remained high.
引用
收藏
页码:2203 / 2211
页数:9
相关论文
共 31 条
[1]   Maximum bite force following unilateral implant-supported prosthetic treatment: within-subject comparison to opposite dentate side [J].
AL-Omiri, M. K. ;
Sghaireen, M. G. ;
Alhijawi, M. M. ;
Alzoubi, I. A. ;
Lynch, C. D. ;
Lynch, E. .
JOURNAL OF ORAL REHABILITATION, 2014, 41 (08) :624-629
[2]   Analysis of Implant Strength After Implantoplasty in Three Implant-Abutment Connection Designs: An In Vitro Study [J].
Alexandre Gehrke, Sergio ;
Aramburu Junior, Jaime Sarda ;
Dedavid, Berenice Anina ;
Shibli, Jamil Awad .
INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2016, 31 (03) :E65-E70
[3]   Two to six-year disease resolution and marginal bone stability rates of a modified resective-implantoplasty therapy in 32 peri-implantitis cases [J].
Bianchini, Marco A. ;
Galarraga-Vinueza, Maria E. ;
Apaza-Bedoya, Karin ;
De Souza, Jose M. ;
Magini, Ricardo ;
Schwarz, Frank .
CLINICAL IMPLANT DENTISTRY AND RELATED RESEARCH, 2019, 21 (04) :758-765
[4]   Implantoplasty Enhancing Peri-implant Bone Stability Over a 3-Year Follow-up: A Case Series [J].
Bianchini, Marco Aurelio ;
Galarraga-Vinueza, Maria Elisa ;
Bedoya, Karin Apaza ;
Correa, Bruna B. ;
Magini, Ricardo de Souza ;
Schwarz, Frank .
INTERNATIONAL JOURNAL OF PERIODONTICS & RESTORATIVE DENTISTRY, 2020, 40 (01) :E1-E8
[5]   Impact of Implantoplasty on Strength of the Implant-Abutment Complex [J].
Chan, Hsun-Liang ;
Oh, Won-Suck ;
Ong, Hwei Sze ;
Fu, Jia-Hui ;
Steigmann, Marius ;
Sierraalta, Marianella ;
Wang, Hom-Lay .
INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2013, 28 (06) :1530-1535
[6]   Fracture strength of implant abutments after fatigue testing: A systematic review and a meta-analysis [J].
Coray, Rafaela ;
Zeltner, Marco ;
Ozcan, Mutlu .
JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2016, 62 :333-346
[7]   Effect of implantoplasty on fracture resistance and surface roughness of standard diameter dental implants [J].
Costa-Berenguer, Xavier ;
Garcia-Garcia, Marta ;
Sanchez-Torres, Alba ;
Sanz-Alonso, Mariano ;
Figueiredo, Rui ;
Valmaseda-Castellon, Eduard .
CLINICAL ORAL IMPLANTS RESEARCH, 2018, 29 (01) :46-54
[8]   Influence of bone insertion level of the implant on the fracture strength of different connection designs: an in vitro study [J].
Gehrke, Sergio Alexandre ;
dos Santos Vianna, Miriam Souza ;
Dedavid, Berenice Anina .
CLINICAL ORAL INVESTIGATIONS, 2014, 18 (03) :715-720
[9]   Influence of superstructure geometry on the mechanical behavior of zirconia implant abutments: a finite element analysis [J].
Geringer, Alexander ;
Diebels, Stefan ;
Nothdurft, Frank P. .
BIOMEDICAL ENGINEERING-BIOMEDIZINISCHE TECHNIK, 2014, 59 (06) :501-506
[10]   Functional aspects of treatment with implant-supported single crowns: a quality control study in subjects with tooth agenesis [J].
Goshima, Kenichi ;
Lexner, Michala O. ;
Thomsen, Carsten Eckhart ;
Miura, Hiroyuki ;
Gotfredsen, Klaus ;
Bakke, Merete .
CLINICAL ORAL IMPLANTS RESEARCH, 2010, 21 (01) :108-114