Optimizing 4D cone beam computed tomography acquisition by varying the gantry velocity and projection time interval

被引:33
作者
O'Brien, Ricky T. [1 ]
Cooper, Benjamin J. [1 ,2 ]
Keall, Paul J. [1 ]
机构
[1] Univ Sydney, Sydney Med Sch, Radiat Phys Lab, Sydney, NSW 2006, Australia
[2] Canberra Hosp, Dept Med Phys, Canberra, ACT 2605, Australia
基金
澳大利亚国家健康与医学研究理事会; 英国医学研究理事会;
关键词
LUNG; CT; RADIOTHERAPY; DELIVERY;
D O I
10.1088/0031-9155/58/6/1705
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Four dimensional cone beam computed tomography (4DCBCT) is an emerging clinical image guidance strategy for tumour sites affected by respiratory motion. In current generation 4DCBCT techniques, both the gantry rotation speed and imaging frequency are constant and independent of the patient's breathing which can lead to projection clustering. We present a mixed integer quadratic programming (MIQP) model for respiratory motion guided-4DCBCT (RMG-4DCBCT) which regulates the gantry velocity and projection time interval, in response to the patient's respiratory signal, so that a full set of evenly spaced projections can be taken in a number of phase, or displacement, bins during the respiratory cycle. In each respiratory bin, an image can be reconstructed from the projections to give a 4D view of the patient's anatomy so that the motion of the lungs, and tumour, can be observed during the breathing cycle. A solution to the full MIQP model in a practical amount of time, 10 s, is not possible with the leading commercial MIQP solvers, so a heuristic method is presented. Using parameter settings typically used on current generation 4DCBCT systems (4 min image acquisition, 1200 projections, 10 respiratory bins) and a sinusoidal breathing trace with a 4 s period, we show that the root mean square (RMS) of the angular separation between projections with displacement binning is 2.7 degrees using existing constant gantry speed systems and 0.6 degrees using RMG-4DCBCT. For phase based binning the RMS is 2.7 degrees using constant gantry speed systems and 2.5 degrees using RMG-4DCBCT. The optimization algorithm presented is a critical step on the path to developing a system for RMG-4DCBCT.
引用
收藏
页码:1705 / 1723
页数:19
相关论文
共 18 条
[1]   Phase and amplitude binning for 4D-CT imaging [J].
Abdelnour, A. F. ;
Nehmeh, S. A. ;
Pan, T. ;
Humm, J. L. ;
Vernon, P. ;
Schoder, H. ;
Rosenzweig, K. E. ;
Mageras, G. S. ;
Yorke, E. ;
Larson, S. M. ;
Erdi, Y. E. .
PHYSICS IN MEDICINE AND BIOLOGY, 2007, 52 (12) :3515-3529
[2]  
[Anonymous], 2008, WORLD CANC REPORT
[3]  
[Anonymous], 2009, METAHEURISTICS DESIG
[4]   Evaluation of sparse-view reconstruction from flat-panel-detector cone-beam CT [J].
Bian, Junguo ;
Siewerdsen, Jeffrey H. ;
Han, Xiao ;
Sidky, Emil Y. ;
Prince, Jerry L. ;
Pelizzari, Charles A. ;
Pan, Xiaochuan .
PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (22) :6575-6599
[5]   The use of a realistic VMAT delivery emulator to optimize dynamic machine parameters for improved treatment efficiency [J].
Boylan, C. J. ;
Rowbottom, C. G. ;
Mackay, R. I. .
PHYSICS IN MEDICINE AND BIOLOGY, 2011, 56 (13) :4119-4133
[6]   The role of radiotherapy in cancer treatment - Estimating optimal utilization from a review of evidence-based clinical guidelines [J].
Delaney, G ;
Jacob, S ;
Featherstone, C ;
Barton, M .
CANCER, 2005, 104 (06) :1129-1137
[7]   Dosimetric effect of respiration-gated beam on IMRT delivery [J].
Duan, J ;
Shen, S ;
Fiveash, JB ;
Brezovich, IA ;
Popple, RA ;
Pareek, PN .
MEDICAL PHYSICS, 2003, 30 (08) :2241-2252
[8]   PRACTICAL CONE-BEAM ALGORITHM [J].
FELDKAMP, LA ;
DAVIS, LC ;
KRESS, JW .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1984, 1 (06) :612-619
[9]  
Kini Vijay R, 2003, Med Dosim, V28, P7, DOI 10.1016/S0958-3947(02)00136-X
[10]   The comparative performance of four respiratory motion predictors for real-time tumour tracking [J].
Krauss, A. ;
Nill, S. ;
Oelfke, U. .
PHYSICS IN MEDICINE AND BIOLOGY, 2011, 56 (16) :5303-5317