Absolute and relative dose measurements with Gafchromic™ EBT film for high energy electron beams with different doses per pulse

被引:11
作者
Fiandra, Christian [1 ]
Ragona, Riccardo [1 ]
Ricardi, Umberto [1 ]
Anglesio, Silvia [2 ]
Giglioli, Francesca Romana [2 ]
机构
[1] Univ Turin, Dept Radiat Oncol, I-10126 Turin, Italy
[2] S Giovanni Battista Hosp, SC Fis Sanitaria 1, I-10126 Turin, Italy
关键词
dosimetry; ionisation chambers; radiation therapy; thin films;
D O I
10.1118/1.3005975
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The authors have evaluated the accuracy, in absolute and relative dose measurements, of the Gafchromic (TM) EBT film in pulsed high-energy electron beams. Typically, the electron beams used in radiotherapy have a dose-per-pulse value of less than 0.1 mGy/pulse. However, very high dose-per-pulse electron beams are employed in certain linear accelerators dedicated to intraoperatory radiation therapy (IORT). In this study, the absorbed dose measurements with Gafchromic (TM) EBT in both low (less than 0.3 mGy per pulse) and high (30 and 70 mGy per pulse) dose-per-pulse electron beams were compared with ferrous sulfate chemical Fricke dosimetry (operated by the Italian Primary Standard Dosimetry Laboratory), a method independent of the dose per pulse. A summary of Gafchromic (TM) EBT in relative and absolute beam output determination is reported. This study demonstrates the independence of Gafchromic (TM) EBT absorption as a function of dose per pulse at different dose levels. A good agreement (within 3%) was found with Fricke dosimeters for plane-base IORT applicators. Comparison with a diode detector is presented for relative dose measurements, showing acceptable agreement both in the steep dose falloff zone and in the homogeneous dose region. This work also provides experimental values for recombination correction factor (K-sat) of a Roos (plane parallel) ionization chamber calculated on the basis of theoretical models for charge recombination.
引用
收藏
页码:5463 / 5470
页数:8
相关论文
共 18 条
[1]   AAPM's TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams [J].
Almond, PR ;
Biggs, PJ ;
Coursey, BM ;
Hanson, WF ;
Huq, MS ;
Nath, R ;
Rogers, DWO .
MEDICAL PHYSICS, 1999, 26 (09) :1847-1870
[2]   Intraoperative radiation therapy using mobile electron linear accelerators: Report of AAPM radiation therapy committee task group no. 72 [J].
Beddar, A. Sam ;
Biggs, Peter J. ;
Chang, Sha ;
Ezzell, Gary A. ;
Faddegon, Bruce A. ;
Hensley, Frank W. ;
Mills, Michael D. .
MEDICAL PHYSICS, 2006, 33 (05) :1476-1489
[3]   The effect of free-electron collection on the recombination correction to ionization measurements of pulsed radiation [J].
Boag, JW ;
Hochhauser, E ;
Balk, OA .
PHYSICS IN MEDICINE AND BIOLOGY, 1996, 41 (05) :885-897
[4]   Energy dependence of response of new high sensitivity radiochromic films for megavoltage and kilovoltage radiation energies [J].
Chiu-Tsao, ST ;
Ho, Y ;
Shankar, R ;
Wang, L ;
Harrison, LB .
MEDICAL PHYSICS, 2005, 32 (11) :3350-3354
[5]   On measuring the output of an IORT mobile dedicated accelerator [J].
De Angelis, C. ;
Soriani, A. ;
Benassi, M. ;
Onori, S. .
RADIATION PROTECTION DOSIMETRY, 2006, 120 (1-4) :221-225
[6]   Precise radiochromic film dosimetry using a flat-bed document scanner [J].
Devic, S ;
Seuntjens, J ;
Sham, E ;
Podgorsak, EB ;
Schmidtlein, CR ;
Kirov, AS ;
Soares, CG .
MEDICAL PHYSICS, 2005, 32 (07) :2245-2253
[7]   Sensitivity of linear CCD array based film scanners used for film dosimetry [J].
Devic, Slobodan ;
Wang, Yi-Zhen ;
Tomic, Nada ;
Podgorsak, Ervin B. .
MEDICAL PHYSICS, 2006, 33 (11) :3993-3996
[8]   Ion recombination correction for very high dose-per-pulse high-energy electron beams [J].
Di Martino, F ;
Giannelli, M ;
Traino, AC ;
Lazzeri, M .
MEDICAL PHYSICS, 2005, 32 (07) :2204-2210
[9]   Clinical use of EBT model Gafchromic™ film in radiotherapy [J].
Fiandra, Christian ;
Ricardi, Umberto ;
Ragona, Riccardo ;
Anglesio, Silvia ;
Giglioli, Francesca Romana ;
Calamia, Elisa ;
Lucio, Francesco .
MEDICAL PHYSICS, 2006, 33 (11) :4314-4319
[10]   Characteristics of the absorbed dose to water standard at ENEA [J].
Guerra, AS ;
Laitano, RF ;
Pimpinella, M .
PHYSICS IN MEDICINE AND BIOLOGY, 1996, 41 (04) :657-674