Calibration and operation of a neutron time-of-flight scintillator array on Nova

被引:2
作者
Chrien, RE
Klare, KA
Murphy, TJ
Rauber, LA
机构
[1] Los Alamos National Laboratory, Los Alamos
关键词
TEMPERATURE-MEASUREMENTS; SPECTRA;
D O I
10.1016/S0920-3796(96)00592-3
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
The Tion diagnostic is a 1020-channel scintillator array on the Nova laser at Livermore. It measures the arrival time of single D-D or D-T neutrons at 27 m from the target with time resolution of 4 ns. The effect of neutron scattering has been obtained from Monte-Carlo neutron transport simulations and the scattered response is convolved with a Gaussian source to fit the data. The data analysis has been checked using targets which produce roughly equal numbers of D-T and D-D counts. The temperature obtained from Tion and from a current-mode neutron time-of-flight detector are in agreement for neutron yields above 1 x 10(9). (C) 1997 Elsevier Science S.A.
引用
收藏
页码:577 / 580
页数:4
相关论文
共 11 条
[1]   THE RATIO OF D-T TO D-D REACTIONS AS A MEASURE OF THE FUEL DENSITY-RADIUS PRODUCT IN INITIALLY TRITIUM-FREE INERTIAL CONFINEMENT FUSION-TARGETS [J].
BLUE, TE ;
HARRIS, DB .
NUCLEAR SCIENCE AND ENGINEERING, 1981, 77 (04) :463-469
[2]  
Briesmeister JF, 1993, LA12625M LOS AL NAT
[3]   FUSION NEUTRON ENERGIES AND SPECTRA [J].
BRYSK, H .
PLASMA PHYSICS AND CONTROLLED FUSION, 1973, 15 (07) :611-617
[4]   NEUTRON-SPECTRA FROM INERTIAL CONFINEMENT FUSION-TARGETS FOR MEASUREMENT OF FUEL AREAL DENSITY AND CHARGED-PARTICLE STOPPING POWERS [J].
CABLE, MD ;
HATCHETT, SP .
JOURNAL OF APPLIED PHYSICS, 1987, 62 (06) :2233-2236
[6]   NEUTRON TIME-OF-FLIGHT ION TEMPERATURE DIAGNOSTIC FOR INERTIAL-CONFINEMENT-FUSION EXPERIMENTS [J].
CHRIEN, RE ;
SIMMONS, DF ;
HOLMBERG, DL .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1992, 63 (10) :4886-4888
[7]   DETECTOR DISTANCE SELECTION FOR NEUTRON TIME-OF-FLIGHT TEMPERATURE-MEASUREMENTS [J].
LERCHE, RA ;
REMINGTON, BA .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1990, 61 (10) :3131-3133
[8]  
MURPHY T, 1994, COMMUNICATION APR
[9]   LANSA - A LARGE NEUTRON SCINTILLATOR ARRAY FOR NEUTRON SPECTROSCOPY AT NOVA [J].
NELSON, MB ;
CABLE, MD .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1992, 63 (10) :4874-4876
[10]  
PLATTS D, 1995, IN PRESS P 10 IEE IN