Real-Time Iterative Monitoring of Radiofrequency Ablation Tumor Therapy with 15O-Water PET Imaging

被引:13
作者
Bao, Ande [1 ,2 ]
Goins, Beth [1 ]
Dodd, Gerald D., III [1 ]
Soundararajan, Anuradha [1 ]
Santoyo, Cristina [1 ]
Otto, Randal A. [2 ]
Davis, Michael D. [3 ]
Phillips, William T. [1 ]
机构
[1] Univ Texas Hlth Sci Ctr San Antonio, Dept Radiol, San Antonio, TX 78229 USA
[2] Univ Texas Hlth Sci Ctr San Antonio, Dept Otolaryngol Head & Neck Surg, San Antonio, TX 78229 USA
[3] Univ Texas Hlth Sci Ctr San Antonio, Res Imaging Ctr, San Antonio, TX 78229 USA
基金
美国国家卫生研究院;
关键词
RF ablation; real-time monitoring; O-15-water; PET; cancer;
D O I
10.2967/jnumed.108.052886
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A method that provides real-time image-based monitoring of solid tumor therapy to ensure complete tumor eradication during image-guided interventional therapy would be a valuable tool. The short, 2-min half-life of 150 makes it possible to perform repeated PET imaging at 20-min intervals at multiple time points before and after image-guided therapy. In this study, O-15-water PET was evaluated as a tool to provide real-time feedback and iterative image guidance to rapidly monitor the intratumoral coverage of radiofrequency (RF) ablation therapy. Methods: Tumor RF ablation therapy was performed on head and neck squamous cell carcinoma (SCC) xenograft tumors (length, similar to 23 mm) in 6 nude rats. The tumor in each animal was ablated with RF (1-cm active size ablation catheter, 70 C for 5 min) twice in 2 separate tumor regions with a 20-min separation. The 150-water PET images were acquired before RF ablation and after the first RF and second RF ablations using a small-animal PET scanner. In each PET session, approximately 100 MBq of O-15-water in 1.0 mL of saline were injected intravenously into each animal. List-mode PET images were acquired for 7 min starting 20 s before injection. PET images were reconstructed by 2-dimensional ordered-subset expectation maximization into single-frame images and dynamic images at 10 s/frame. PET images were displayed and analyzed with software. Results: Pre-RF ablation images demonstrate that 150-water accumulates in tumors with 150 activity reaching peak levels immediately after administration. After RF ablation, the ablated region had almost zero activity, whereas the unablated tumor tissue continued to have a high 150-water accumulation. Using image feedback, the RF probe was repositioned to a tumor region with residual 150-water uptake and then ablated. The second RF ablation in this new region of the tumor resulted in additional ablation of the solid tumor, with a corresponding decrease in activity on the 150-water PET image. Conclusion: 150-water PET clearly demonstrated the ablated tumor region, whereas the unablated tumor continued to show high 150-water accumulation. 150-water imaging shows promise as a tool for on-site, real-time monitoring of image-guided interventional cancer therapy.
引用
收藏
页码:1723 / 1729
页数:7
相关论文
共 33 条
[1]  
[Anonymous], 1996, GUIDE CARE USE LAB A
[2]  
Antoch G, 2005, J NUCL MED, V46, P520
[3]   Setup and characterization of a human head and neck squamous cell carcinoma xenograft model in nude rats [J].
Bao, Ande ;
Phillips, William T. ;
Goins, Beth ;
McGuff, Howard S. ;
Zheng, Xiangpeng ;
Woolley, F. Ross ;
Natarajan, Mohan ;
Santoyo, Cristina ;
Miller, Frank R. ;
Otto, Randal A. .
OTOLARYNGOLOGY-HEAD AND NECK SURGERY, 2006, 135 (06) :853-857
[4]   MR-guided ablation of hepatocellular carcinoma aided by gadoxetic acid [J].
Bathe, Oliver F. ;
Mahallati, Houman .
JOURNAL OF SURGICAL ONCOLOGY, 2007, 95 (08) :670-673
[5]   Radiofrequency, cryoablation, and other modalities for breast cancer ablation [J].
Bland, Keiva L. ;
Gass, Jennifer ;
Klimberg, V. Suzanne .
SURGICAL CLINICS OF NORTH AMERICA, 2007, 87 (02) :539-+
[6]   Prediction of subtle thermal histopathological change using a novel analysis of Gd-DTPA kinetics [J].
Cheng, HLM ;
Purcell, CM ;
Bilbao, JM ;
Plewes, DB .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2003, 18 (05) :585-598
[7]   Magnetic resonance imaging for hepatic radiofrequency ablation [J].
Clasen, S. ;
Boss, A. ;
Schmidt, D. ;
Fritz, J. ;
Schraml, C. ;
Claussen, C. D. ;
Pereira, P. L. .
EUROPEAN JOURNAL OF RADIOLOGY, 2006, 59 (02) :140-148
[8]   MR thermometry for monitoring tumor ablation [J].
de Senneville, Baudouin Denis ;
Mougenot, Charles ;
Quesson, Bruno ;
Dragonu, Iulius ;
Grenier, Nicolas ;
Moonen, Chrit T. W. .
EUROPEAN RADIOLOGY, 2007, 17 (09) :2401-2410
[9]   Radiofrequency thermal ablation: Computer analysis of the size of the thermal injury created by overlapping ablations [J].
Dodd, GD ;
Frank, MS ;
Aribandi, M ;
Chopra, S ;
Chintapalli, KN .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2001, 177 (04) :777-782
[10]   Minimally invasive treatment of malignant hepatic tumors: At the threshold of a major breakthrough [J].
Dodd, GD ;
Soulen, MC ;
Kane, RA ;
Livraghi, T ;
Lees, WR ;
Yamashita, Y ;
Gillams, AR ;
Karahan, OI ;
Rhim, H .
RADIOGRAPHICS, 2000, 20 (01) :9-27