Towards an inline reconstruction architecture for micro-CT systems

被引:28
作者
Brasse, D
Humbert, B
Mathelin, C
Rio, MC
Guyonnet, JL
机构
[1] Univ Strasbourg 1, CNRS, Inst Rech Subatom, IN2P3, Strasbourg, France
[2] Inst Genet & Biol Mol & Cellulaire, Illkirch Graffenstaden, France
关键词
D O I
10.1088/0031-9155/50/24/003
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Recent developments in micro-CT have revolutionized the ability to examine in vivo living experimental animal models such as mouse with a spatial resolution less than 50 mu m. The main requirements of in vivo imaging for biological researchers are a good spatial resolution, a low dose induced to the animal during the full examination and a reduced acquisition and reconstruction time for screening purposes. We introduce inline acquisition and reconstruction architecture to obtain in real time the 3D attenuation map of the animal fulfilling the three previous requirements. The micro-CT system is based on commercially available x-ray detector and micro-focus x-ray source. The reconstruction architecture is based on a cluster of PCs where a dedicated communication scheme combining serial and parallel treatments is implemented. In order to obtain high performance transmission rate between the detector and the reconstruction architecture, a dedicated data acquisition system is also developed. With the proposed solution, the time required to filter and backproject a projection of 2048 x 2048 pixels inside a volume of 140 mega voxels using the Feldkamp algorithm is similar to 500 ms, the time needed to acquire the same projection.
引用
收藏
页码:5799 / 5811
页数:13
相关论文
共 22 条
[1]  
Andarawewa KL, 2003, CANCER RES, V63, P5844
[2]   USE OF TRANSPUTERS IN A 3-D POSITRON EMISSION TOMOGRAPH [J].
ATKINS, MS ;
MURRAY, D ;
HARROP, R .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1991, 10 (03) :276-283
[3]  
CORIC S, 10 AIM INT S FPGA
[4]   Distance-driven projection and backprojection in three dimensions [J].
De Man, B ;
Basu, S .
PHYSICS IN MEDICINE AND BIOLOGY, 2004, 49 (11) :2463-2475
[5]  
ELSTER AC, 2002, SIMS 43 C SIM MOD
[6]   PRACTICAL CONE-BEAM ALGORITHM [J].
FELDKAMP, LA ;
DAVIS, LC ;
KRESS, JW .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1984, 1 (06) :612-619
[7]   The design and implementation of FFTW3 [J].
Frigo, M ;
Johnson, SG .
PROCEEDINGS OF THE IEEE, 2005, 93 (02) :216-231
[8]   Fast Feldkamp reconstruction based on focus of attention and distributed computing [J].
Gregor, J ;
Gleason, SS ;
Paulus, MJ ;
Cates, J .
INTERNATIONAL JOURNAL OF IMAGING SYSTEMS AND TECHNOLOGY, 2002, 12 (06) :229-234
[9]   Analysis of an exact inversion algorithm for spiral cone-beam CT [J].
Katsevich, A .
PHYSICS IN MEDICINE AND BIOLOGY, 2002, 47 (15) :2583-2597
[10]   Exact and approximate algorithms for helical cone-beam CT [J].
Kudo, H ;
Rodet, T ;
Noo, F ;
Defrise, M .
PHYSICS IN MEDICINE AND BIOLOGY, 2004, 49 (13) :2913-2931