Endocardial impedance mapping during circumferential pulmonary vein ablation of atrial fibrillation differentiates between atrial and venous tissue

被引:17
作者
Lang, CCE
Gugliotta, F
Santinelli, V
Mesas, C
Tomita, T
Vicedomini, G
Augello, G
Gulletta, S
Mazzone, P
De Cobelli, F
Del Maschio, A
Pappone, C [1 ]
机构
[1] Hosp San Raffaele, Div Arrhythmol & Cardiac Electropphysiol, I-20132 Milan, Italy
[2] Hosp San Raffaele, Dept Diagnost Radiol, I-20132 Milan, Italy
关键词
ablation; atrial fibrillation; impedance; mapping;
D O I
10.1016/j.hrthm.2005.10.017
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
BACKGROUND Circumferential pulmonary vein ablation (CPVA) is an effective treatment for atrial fibrillation (AF). Accurate left atrial (LA) mapping is essential for creating lesions at the LA-pulmonary vein (PV) junction, avoiding PV stenosis. OBJECTIVES The purpose of this study was to establish whether endocardial impedance varies within the LA and PVs and whether it is a useful tool for mapping and ablation. METHODS Pilot Phase: Three-dimensional LA maps were created using CARTO. Impedance (Z) was measured using a radiofrequency generator at multiple points in the LA, PV ostia (PVO), and deep PVs in 79 patients undergoing their first AF ablation (group 1) and 29 patients undergoing repeat CPVA (group 2). Prospective Phase: In an additional 20 patients, using pilot phase data, one operator defined catheter tip location as either LA or PVO based on CARTO and fluoroscopy. A second operator blinded to CARTO simultaneously did the same based oil impedance at 15 +/- 4 points per patient. RESULTS Group 1: Z(LA) was 99.4 +/- 9.0 Omega. Z(PVO) was higher (109.2 +/- 8.5 Omega), rising further as the catheter advanced into deep PV (137 Omega +/- 18). Z(PVO) differed from Z(LA) by 9 +/- 4 Q. Group 2 had a lower Z(LA) and Z(PVO) compared with group 1 (P < .05). Impedance monitoring differentiated between LA and PVO, with 91% specificity and sensitivity, 96% positive predictive value, and 81% negative predictive value. At 3-month follow-up, no patients had evidence of PV stenosis on magnetic resonance imaging. CONCLUSION Impedance mapping reliably identifies the LA-PV transitional zone, facilitating AF ablation, and its use is associated with a low incidence of PV stenosis.
引用
收藏
页码:171 / 178
页数:8
相关论文
共 25 条
  • [1] Using electrical impedance to predict catheter-endocardial contact during RF cardiac ablation
    Cao, H
    Tungjitkusolmun, S
    Choy, YB
    Tsai, JZ
    Vorperian, VR
    Webster, JG
    [J]. IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2002, 49 (03) : 247 - 253
  • [2] FEM analysis of predicting electrode-myocardium contact from RF cardiac catheter ablation system impedance
    Cao, H
    Speidel, MA
    Tsai, JZ
    Van Lysel, MS
    Vorperian, VR
    Webster, JG
    [J]. IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2002, 49 (06) : 520 - 526
  • [3] Detection of inadvertent catheter movement into a pulmonary vein during radiofrequency catheter ablation by real-time impedance monitoring
    Cheung, P
    Hall, B
    Chugh, A
    Good, E
    Lemola, K
    Han, J
    Tamirisa, K
    Pelosi, F
    Morady, F
    Oral, H
    [J]. JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2004, 15 (06) : 674 - 678
  • [4] Changes in myocardial electrical impedance induced by coronary artery occlusion in pigs with and without preconditioning - Correlation with local ST-segment potential and ventricular arrhythmias
    Cinca, J
    Warren, M
    Carreno, A
    Tresanchez, M
    Armadans, L
    Gomez, P
    SolerSoler, J
    [J]. CIRCULATION, 1997, 96 (09) : 3079 - 3086
  • [5] ELECTRICAL RESISTANCES OF INTERSTITIAL AND MICROVASCULAR SPACE AS DETERMINANTS OF THE EXTRACELLULAR ELECTRICAL-FIELD AND VELOCITY OF PROPAGATION IN VENTRICULAR MYOCARDIUM
    FLEISCHHAUER, J
    LEHMANN, L
    KLEBER, AG
    [J]. CIRCULATION, 1995, 92 (03) : 587 - 594
  • [6] Mechanistic insights into very slow conduction in branching cardiac tissue - A model study
    Kucera, JP
    Rudy, Y
    [J]. CIRCULATION RESEARCH, 2001, 89 (09) : 799 - 806
  • [7] Leite L, 2002, PACING CLIN ELECTROP, V25, P559
  • [8] Assessment of pulmonary vein anatomic variability by magnetic resonance imaging: Implications for catheter ablation techniques for atrial fibrillation
    Mansour, M
    Holmvang, G
    Sosnovik, D
    Migrino, R
    Abbara, S
    Ruskin, J
    Keane, D
    [J]. JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2004, 15 (04) : 387 - 393
  • [9] Moubarak JB, 2000, PACE, V23, P1836
  • [10] A new approach for catheter ablation of atrial fibrillation: Mapping of the electrophysiologic substrate
    Nademanee, K
    McKenzie, J
    Kosar, E
    Schwab, M
    Sunsaneewitayakul, B
    Vasavakul, T
    Khunnawat, C
    Ngarmukos, T
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2004, 43 (11) : 2044 - 2053