The Moxus Modular metabolic system evaluated with two sensors for ventilation against the Douglas bag method

被引:15
作者
Rosdahl, Hans [1 ]
Lindberg, Thomas [1 ]
Edin, Fredrik [1 ]
Nilsson, Johnny [1 ]
机构
[1] Swedish Sch Sport & Hlth Sci GIH, Stockholm, Sweden
关键词
Moxus Modular metabolic system; Douglas bag method; Validity; Reliability; Turbine flowmeter; Pneumotachometer; Oxygen uptake; Carbon dioxide production; Ventilation; Respiratory exchange ratio; MAXIMAL OXYGEN INTAKE; GAS-ANALYSIS SYSTEMS; HERITAGE FAMILY; SKELETAL-MUSCLE; EXERCISE; VALIDITY; RELIABILITY; REPRODUCIBILITY; PNEUMOTACHOMETER; PERFORMANCE;
D O I
10.1007/s00421-012-2551-1
中图分类号
Q4 [生理学];
学科分类号
071003 ;
摘要
This study evaluated the Moxus metabolic system with the Douglas bag method (DBM) as criterion. Reliability and validity were investigated in a wide range of ventilation and oxygen uptake and two sensors for determining ventilation were included. Thirteen well-trained athletes participated in one pre-test and four tests for data collection, exercising on a cycle ergometer at five submaximal powers (50-263 W) and at . Gas exchange variables were measured simultaneously using a serial setup with data collected on different days in an order randomized between Moxus with pneumotachometer (MP) and turbine flowmeter (MT) sensors for ventilation. Reliability with both sensors was comparable to the DBM. Average CV (%) of all exercise intensities were with MP: 3.0 +/- A 1.3 for VO2, 3.8 +/- A 1.5 for VCO2, 3.1 +/- A 1.2 for the respiratory exchange ratio (RER) and 4.2 +/- A 0.8 for V (E). The corresponding values with MT were: 2.7 +/- A 0.3 for VO2, 4.7 +/- A 0.4 for VCO2, 3.3 +/- A 0.9 for RER and 4.8 +/- A 1.4 for V (E). Validity was acceptable except for small differences related to the determination of ventilation. The relative differences in relation to DBM at the powers including were similar for both sensors with the ranges being: +4 to -2 % for V (E), +5 to -3 % for VO2 and +5 to -4 % for VCO2 while RER did not differ at any power. The Moxus metabolic system shows high and adequate reliability and reasonable validity over a wide measurement range. At a few exercise levels, V (E) differed slightly from DBM, resulting in concomitant changes in VO2 and VCO2.
引用
收藏
页码:1353 / 1367
页数:15
相关论文
共 9 条
  • [1] The Moxus Modular metabolic system evaluated with two sensors for ventilation against the Douglas bag method
    Hans Rosdahl
    Thomas Lindberg
    Fredrik Edin
    Johnny Nilsson
    European Journal of Applied Physiology, 2013, 113 : 1353 - 1367
  • [2] Examination of the Moxus Modular Metabolic System by the Douglas-bag technique
    Medbo, Jon Ingulf
    Mamen, Asgeir
    Beltrami, Fernando G.
    APPLIED PHYSIOLOGY NUTRITION AND METABOLISM, 2012, 37 (05) : 860 - 871
  • [3] Evaluation of the Oxycon Mobile metabolic system against the Douglas bag method
    Rosdahl, Hans
    Gullstrand, Lennart
    Salier-Eriksson, Jane
    Johansson, Patrik
    Schantz, Peter
    EUROPEAN JOURNAL OF APPLIED PHYSIOLOGY, 2010, 109 (02) : 159 - 171
  • [4] Evaluation of the Oxycon Mobile metabolic system against the Douglas bag method
    Hans Rosdahl
    Lennart Gullstrand
    Jane Salier-Eriksson
    Patrik Johansson
    Peter Schantz
    European Journal of Applied Physiology, 2010, 109 : 159 - 171
  • [5] The validity of the Moxus Modular metabolic system during incremental exercise tests: impacts on detection of small changes in oxygen consumption
    Beltrami, Fernando G.
    Froyd, Christian
    Mamen, Asgeir
    Noakes, Timothy D.
    EUROPEAN JOURNAL OF APPLIED PHYSIOLOGY, 2014, 114 (05) : 941 - 950
  • [6] Breathing resistance in automated metabolic systems is high in comparison with the Douglas Bag method and previous recommendations
    Ainegren, Mats
    Jensen, Kurt
    Rosdahl, Hans
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART P-JOURNAL OF SPORTS ENGINEERING AND TECHNOLOGY, 2018, 232 (02) : 122 - 130
  • [7] The validity of the Moxus Modular metabolic system during incremental exercise tests: impacts on detection of small changes in oxygen consumption
    Fernando G. Beltrami
    Christian Froyd
    Asgeir Mamen
    Timothy D. Noakes
    European Journal of Applied Physiology, 2014, 114 : 941 - 950
  • [8] The Ventilation-Corrected ParvoMedics TrueOne 2400 Provides a Valid and Reliable Assessment of Resting Metabolic Rate (RMR) in Athletes Compared With the Douglas Bag Method
    Woods, Amy L.
    Garvican-Lewis, Laura A.
    Rice, Anthony J.
    Thompson, Kevin G.
    INTERNATIONAL JOURNAL OF SPORT NUTRITION AND EXERCISE METABOLISM, 2016, 26 (05) : 454 - 463
  • [9] A STUDY OF EVALUATION METHOD OF ENERGY CONSERVATION UNDER TWO DIFFERENT CONTROL MODES FOR THE TUNNEL VENTILATION SYSTEM
    Fang, Biao
    Zhang, Quan
    Zhang, Guoqiang
    He, Dejun
    FIFTH INTERNATIONAL WORKSHOP ON ENERGY AND ENVIRONMENT OF RESIDENTIAL BUILDINGS AND THIRD INTERNATIONAL CONFERENCE ON BUILT ENVIRONMENT AND PUBLIC HEALTH, VOL I AND II, PROCEEDINGS, 2009, : 710 - 717