Degrees of Reality Airway Anatomy of High-fidelity Human Patient Simulators and Airway Trainers

被引:106
作者
Schebesta, Karl [1 ]
Huepfl, Michael [1 ]
Roessler, Bernhard [1 ]
Ringl, Helmut [2 ]
Mueller, Michael P. [3 ]
Kimberger, Oliver [4 ]
机构
[1] Med Univ Vienna, Med Simulat & Emergency Management Res Grp, Dept Anesthesia Gen Intens Care & Pain Management, A-1090 Vienna, Austria
[2] Med Univ Vienna, Dept Radiol, A-1090 Vienna, Austria
[3] Univ Hosp Dresden, Interdisciplinary Med Simulat Ctr Dresden, Dept Anesthesiol, Dresden, Germany
[4] Med Univ Vienna, Div Gen Anesthesia & Intens Care, Dept Anesthesia Gen Intens Care & Pain Management, A-1090 Vienna, Austria
关键词
3-DIMENSIONAL COMPUTED-TOMOGRAPHY; TRAINING MANNEQUINS; DIFFICULT INTUBATION; EQUIPMENT MAN; MANAGEMENT; PERFORMANCE; MACINTOSH; EDUCATION; INSERTION; HISTORY;
D O I
10.1097/ALN.0b013e318254cf41
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Background: Human patient simulators and airway training manikins are widely used to train airway management skills to medical professionals. Furthermore, these patient simulators are employed as standardized "patients" to evaluate airway devices. However, little is known about how realistic these patient simulators and airway-training manikins really are. This trial aimed to evaluate the upper airway anatomy of four high-fidelity patient simulators and two airway trainers in comparison with actual patients by means of radiographic measurements. The volume of the pharyngeal airspace was the primary outcome parameter. Methods: Computed tomography scans of 20 adult trauma patients without head or neck injuries were compared with computed tomography scans of four high-fidelity patient simulators and two airway trainers. By using 14 predefined distances, two cross-sectional areas and three volume parameters of the upper airway, the manikins' similarity to a human patient was assessed. Results: The pharyngeal airspace of all manikins differed significantly from the patients' pharyngeal airspace. The HPS Human Patient Simulator (METI (R), Sarasota, FL) was the most realistic high-fidelity patient simulator (6/19 [32%] of all parameters were within the 95% CI of human airway measurements). Conclusion: The airway anatomy of four high-fidelity patient simulators and two airway trainers does not reflect the upper airway anatomy of actual patients. This finding may impact airway training and confound comparative airway device studies.
引用
收藏
页码:1204 / 1209
页数:6
相关论文
共 40 条
  • [1] Tracheal intubation following training with the GlideScope® compared to direct laryngoscopy
    Ayoub, C. M.
    Kanazi, G. E.
    Al Alami, A.
    Rameh, C.
    El-Khatib, M. F.
    [J]. ANAESTHESIA, 2010, 65 (07) : 674 - 678
  • [2] The history of simulation in medical education and possible future directions
    Bradley, P
    [J]. MEDICAL EDUCATION, 2006, 40 (03) : 254 - 262
  • [3] Fiberoptic Oral Intubation The Effect of Model Fidelity on Training for Transfer to Patient Care
    Chandra, Deven B.
    Savoldelli, Georges L.
    Joo, Hwan S.
    Weiss, Israel D.
    Naik, Viren N.
    [J]. ANESTHESIOLOGY, 2008, 109 (06) : 1007 - 1013
  • [4] Chou HC, 2002, ANESTH ANALG, V94, P478
  • [5] Large hypopharyngeal tongue: A shared anatomic abnormality for difficult mask ventilation, difficult intubation, and obstructive sleep apnea?
    Chou, HC
    Wu, TL
    [J]. ANESTHESIOLOGY, 2001, 94 (05) : 936 - 937
  • [6] Evaluation of four airway training manikins as patient simulators for the insertion of single use laryngeal mask airways
    Cook, T. M.
    Green, C.
    McGrath, J.
    Srivastava, R.
    [J]. ANAESTHESIA, 2007, 62 (07) : 713 - 718
  • [7] Cook TM, 2011, ANAESTHESIA, V66, P529, DOI 10.1111/j.1365-2044.2011.06726.x
  • [8] Litigation related to airway and respiratory complications of anaesthesia: an analysis of claims against the NHS in England 1995-2007
    Cook, T. M.
    Scott, S.
    Mihai, R.
    [J]. ANAESTHESIA, 2010, 65 (06) : 556 - 563
  • [9] Novel airway devices: Spoilt for choice?
    Cook, TM
    [J]. ANAESTHESIA, 2003, 58 (02) : 107 - 110
  • [10] Cooper JB, 2008, POSTGRAD MED J, V84, P563, DOI 10.1136/qshc.2004.009886