Label-free Detection of Microcystin-LR in Waters Using Real-Time Potentiometric Biosensors Based on Single-Walled Carbon Nanotubes Imprinted Polymers

被引:4
作者
Queiros, Raquel B. [2 ]
Noronha, J. P. [3 ]
Marques, P. V. S. [2 ]
Sales, M. G. F. [1 ]
机构
[1] BioMark Sensor Res, Rua Dr Antonio Bernardino de Almeida, P-4200072 Oporto, Portugal
[2] INESC TEC, P-4150179 Oporto, Portugal
[3] Univ Nova Lisboa, REQUIMTE, CQFB, Fac Ciencias & Tecnol, P-1200 Lisbon, Portugal
来源
26TH EUROPEAN CONFERENCE ON SOLID-STATE TRANSDUCERS, EUROSENSOR 2012 | 2012年 / 47卷
关键词
Biosensors; MC-LR; CNTs; Potenciometric;
D O I
10.1016/j.proeng.2012.09.258
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Microcystin-LR (MC-LR) is a dangerous toxin found in environmental waters, quantified by high performance liquid chromatography and/or enzyme-linked immunosorbent assays. Quick, low cost and on-site analysis is thus required to ensure human safety and wide screening programs. This work proposes label-free potentiometric sensors made of solid-contact electrodes coated with a surface imprinted polymer on the surface of Multi-Walled Carbon NanoTubes (CNTs) incorporated in a polyvinyl chloride membrane. The imprinting effect was checked by using non-imprinted materials. The MC-LR sensitive sensors were evaluated, characterized and applied successfully in spiked environmental waters. The presented method offered the advantages of low cost, portability, easy operation and suitability for adaptation to flow methods. (C) 2012 Elsevier Ltd....Selection and/or peer-review under responsibility of the Symposium Cracoviense Sp. z.o.o.
引用
收藏
页码:758 / 761
页数:4
相关论文
共 5 条
[1]  
[Anonymous], 2007, REPUBLIC J, V164, P5747
[2]   EXTRACTION AND HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC METHOD FOR THE DETERMINATION OF MICROCYSTINS IN RAW AND TREATED WATERS [J].
LAWTON, LA ;
EDWARDS, C ;
CODD, GA .
ANALYST, 1994, 119 (07) :1525-1530
[3]   CYANOBACTERIAL MICROCYSTIN-LR IS A POTENT AND SPECIFIC INHIBITOR OF PROTEIN PHOSPHATASE-1 AND PHOSPHATASE-2A FROM BOTH MAMMALS AND HIGHER-PLANTS [J].
MACKINTOSH, C ;
BEATTIE, KA ;
KLUMPP, S ;
COHEN, P ;
CODD, GA .
FEBS LETTERS, 1990, 264 (02) :187-192
[4]  
Mao Y, 2001, BIOSENS BIOELECTRON, V28, P291
[5]   SELECTIVITY COEFFICIENTS FOR ION-SELECTIVE ELECTRODES - RECOMMENDED METHODS FOR REPORTING K-A,B(POT) VALUES - (TECHNICAL REPORT) [J].
UMEZAWA, Y ;
UMEZAWA, K ;
SATO, H .
PURE AND APPLIED CHEMISTRY, 1995, 67 (03) :507-518