Estimation of Mean Blood Pressure from Oscillometric and Manual Methods

被引:4
作者
Zheng, D. [1 ]
Murray, A. [1 ]
机构
[1] Univ Newcastle, Freeman Hosp, Dept Reg Med Phys, Newcastle Upon Tyne NE7 7DN, Tyne & Wear, England
来源
COMPUTERS IN CARDIOLOGY 2008, VOLS 1 AND 2 | 2008年
关键词
D O I
10.1109/CIC.2008.4749198
中图分类号
TP [自动化技术、计算机技术];
学科分类号
0812 ;
摘要
The aim of this study was to estimate mean blood pressure from oscillometric and manual methods. Ten healthy subjects were studied with three repeat blood pressure measurements. Manual systolic and diastolic blood pressures (SBP and DBP) were obtained by two trained observers. During the measurement the oscillometric cuff pressure waveform was recorded digitally. The cuff pressure corresponding to the largest oscillometric pulse pressure was taken as the mean arterial pressure (AMP). AMP was also estimated from manual DBP plus one third the pressure change from DBP to SBP. Blood pressure measurement variability and the difference between automated and manual MAP were quantified. The overall coefficients of variabilily for manual SBP and DBP were 2.2% and 4.5%. Corresponding values for manual and automated MAP were 3.1% and 3.7%. The automated MAP, as a percentage of the difference between manual DBP and SBP was 28 +/- 14% (mean SD), which was lower than the classically assumed mean value of 33%. In conclusion, the relationship between MAP and SBP and DBP is complex with a large between-subject SD variability of 14%.
引用
收藏
页码:941 / 944
页数:4
相关论文
共 50 条
  • [21] Oscillometric blood pressure measurements: differences between measured and calculated mean arterial pressure
    Kiers, H. D.
    Hofstra, J. M.
    Wetzels, J. F. M.
    NETHERLANDS JOURNAL OF MEDICINE, 2008, 66 (11) : 474 - 479
  • [22] Novel methods of testing and calibration of oscillometric blood pressure monitors
    Celler, Branko G.
    Argha, Ahmadreza
    Phu Ngoc Le
    Ambikairajah, Eliathamby
    PLOS ONE, 2018, 13 (08):
  • [23] Coefficient-Free Blood Pressure Estimation Based on Arterial Lumen Area Oscillations in Oscillometric Methods
    Koohi, Iraj
    Batkin, Izmail
    Groza, Voicu Z.
    Shirmohammadi, Shervin
    Dajani, Hilmi R.
    Ahmad, Saif
    2015 IEEE 9TH INTERNATIONAL SYMPOSIUM ON INTELLIGENT SIGNAL PROCESSING (WISP), 2015, : 165 - 170
  • [24] Instruments and Methods for Calibration of Oscillometric Blood Pressure Measurement Devices
    Balestrieri, Eulalia
    Rapuano, Sergio
    IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2010, 59 (09) : 2391 - 2404
  • [25] Mean arterial pressure estimation method using morphological changes in oscillometric waveform
    Song, S. H.
    Kim, D. K.
    Lee, J. S.
    Chee, Y. J.
    Kim, I. Y.
    CINC: 2009 36TH ANNUAL COMPUTERS IN CARDIOLOGY CONFERENCE, 2009, 36 : 737 - +
  • [26] Oscillometric blood pressure measurements on smartphones using vibrometric force estimation
    Barry, Colin
    Xuan, Yinan
    Fascetti, Ava
    Moore, Alison
    Wang, Edward Jay
    SCIENTIFIC REPORTS, 2024, 14 (01):
  • [27] Performance Analysis of Oscillometric Blood Pressure Estimation Techniques in Cardiac Patients
    Sidhu, Ekambir
    Yoshida, Masayoshi
    Groza, Voicu Z.
    Dajani, Hilmi R.
    Bolic, Miodrag
    2019 IEEE INTERNATIONAL SYMPOSIUM ON MEDICAL MEASUREMENTS AND APPLICATIONS (MEMEA), 2019,
  • [28] Deep Boltzmann Regression With Mimic Features for Oscillometric Blood Pressure Estimation
    Lee, Soojeong
    Chang, Joon-Hyuk
    IEEE SENSORS JOURNAL, 2017, 17 (18) : 5982 - 5993
  • [29] Deep learning ensemble with asymptotic techniques for oscillometric blood pressure estimation
    Lee, Soojeong
    Chang, Joon-Hyuk
    COMPUTER METHODS AND PROGRAMS IN BIOMEDICINE, 2017, 151 : 1 - 13
  • [30] Artificial Intelligence Based Blood Pressure Estimation From Auscultatory and Oscillometric Waveforms: A Methodological Review
    Argha, Ahmadreza
    Celler, Branko G.
    Lovell, Nigel H.
    IEEE REVIEWS IN BIOMEDICAL ENGINEERING, 2022, 15 : 152 - 168