Characterization of a prototype rapid kilovoltage x-ray image guidance system designed for a ring shape radiation therapy unit

被引:78
作者
Cai, Bin [1 ]
Laugeman, Eric [1 ]
Mazur, Thomas R. [1 ]
Park, Justin C. [1 ]
Henke, Lauren E. [1 ]
Kim, Hyun [1 ]
Hugo, Geoffrey D. [1 ]
Mutic, Sasa [1 ]
Li, Hua [1 ]
机构
[1] Washington Univ, Dept Radiat Oncol, St Louis, MO 63110 USA
关键词
fast CBCT imaging; KV-IGRT; ring gantry LINAC; QUALITY-ASSURANCE; RADIOTHERAPY; QA;
D O I
10.1002/mp.13396
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose This study aims to characterize the performance of a prototype rapid kilovoltage (kV) x-ray image guidance system onboard the newly released Halcyon 2.0 linear accelerator (Varian Medical Systems, Palo Alto, CA) by use of conventional and innovatively designed testing procedures. Methods Basic imaging system performance tests and radiation dose measurements were performed for all eleven kV-cone beam computed tomography (CBCT) imaging protocols available on a preclinical Halcyon 2.0 LINAC. Both conventional CBCT reconstruction using the Feldkamp-Davis-Kress (FDK) algorithm and a novel, advanced iterative reconstruction (iCBCT) available on this platform were evaluated. Standard image quality metrics, including slice thickness accuracy, high-contrast resolution, low-contrast resolution, regional uniformity and noise, and CT Hounsfield unit (HU) number accuracy and linearity were evaluated using a manufacturer-supplied QUART phantom (GmbH, Zorneding, Germany) and an independent image quality phantom (Catphan 500, The Phantom Laboratory, New York, NY). Due to the simplified design of the QUART phantom, we developed surrogate and clinically feasible strategies for measuring slice thickness and high- and low-contrast resolution. Imaging dose delivered by these eleven protocols was measured using a computed tomography dose index phantom and pencil chamber with commonly accepted methods and procedures. A subset of measurements were repeated on a conventional C-arm LINAC (TrueBeam and Trilogy, Varian Medical System) for comparison. Clinical patient images of pelvic and abdominal regions are also presented for qualitative assessment as part of a feasibility study for clinical implementation. Results Image acquisition time was 17-42 s on the Halcyon system compared with 60 s on the C-arm LINAC systems. The kV imager projection offset, imaging and treatment isocenter coincidence and the couch three-dimensional match movement all achieved less than1 mm mechanical accuracy. All major image quality metrics were within either the national guideline or vendor-recommended tolerances. The designed surrogate approach with the QUART phantom showed a range of 0.24-0.35 cycles/mm for spatial resolution, a contrast-to-noise ratio (CNR) of 2-20 for FDK reconstruction and a tolerance of 0.5 mm for slice thickness. Other metrics derived from the Catphan images obtained on the Halcyon and C-arm LINACs showed comparable values for the FDK reconstruction. The iterative reconstruction tended to reduce noise, as evidenced by a higher CNR ratio. The fast scan pelvis protocols for Halcyon resulted in 50% lower dose compared to the standard scans, and the thorax fast protocol similarly delivered 10% lower dose than the standard thoracic scan. Preliminary patient images indicated that rapid kV CBCT with breath-hold is feasible, with improved imaging quality compared to free-breathing scans. Conclusion Independent and comprehensive characterization of the kV imaging guidance system on the Halcyon 2.0 system demonstrated acceptable image quality for clinical use. The imaging unit onboard the Halcyon meets vendor specifications and satisfies requirements for routine clinical use. The fast kV imaging system enables the potential for volumetric CBCT acquisition during a single breath-hold and the iterative reconstruction tends to reduce the noise therefore has the potential to improve the CNR for normal size patient. (c) 2019 American Association of Physicists in Medicine
引用
收藏
页码:1355 / 1370
页数:16
相关论文
共 19 条
  • [1] [Anonymous], CANC STAT
  • [2] Evaluation of the truebeam machine performance check (MPC) geometric checks for daily IGRT geometric accuracy quality assurance
    Barnes, Michael P.
    Greer, Peter B.
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2017, 18 (03): : 200 - 206
  • [3] Estimating the demand for radiotherapy from the evidence: A review of changes from 2003 to 2012
    Barton, Michael B.
    Jacob, Susannah
    Shafiq, Jesmin
    Wong, Karen
    Thompson, Stephen R.
    Hanna, Timothy P.
    Delaney, Geoff P.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2014, 112 (01) : 140 - 144
  • [4] A quality assurance program for image quality of cone-beam CT guidance in radiation therapy
    Bissonnette, Jean-Pierre
    Moseley, Douglas J.
    Jaffray, David A.
    [J]. MEDICAL PHYSICS, 2008, 35 (05) : 1807 - 1815
  • [5] Quality assurance for image-guided radiation therapy utilizing CT-based technologies: A report of the AAPM TG-179
    Bissonnette, Jean-Pierre
    Balter, Peter A.
    Dong, Lei
    Langen, Katja M.
    Lovelock, D. Michael
    Miften, Moyed
    Moseley, Douglas J.
    Pouliot, Jean
    Sonke, Jan-Jakob
    Yoo, Sua
    [J]. MEDICAL PHYSICS, 2012, 39 (04) : 1946 - 1963
  • [6] Online adaptive radiotherapy of the bladder: Small bowel irradiated-volume reduction
    Burridge, Nichola
    Amer, Ali
    Marchant, Tom
    Sykes, Jonathan
    Stratford, Julie
    Henry, Ann
    McBain, Catherine
    Price, Pat
    Moore, Chris
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2006, 66 (03): : 892 - 897
  • [7] Evaluation of the Machine Performance Check application for TrueBeam Linac
    Clivio, Alessandro
    Vanetti, Eugenio
    Rose, Steven
    Nicolini, Giorgia
    Belosi, Maria F.
    Cozzi, Luca
    Baltes, Christof
    Fogliata, Antonella
    [J]. RADIATION ONCOLOGY, 2015, 10
  • [8] Evaluation of an algorithm for the assessment of the MTF using an edge method
    Greer, PB
    van Doorn, T
    [J]. MEDICAL PHYSICS, 2000, 27 (09) : 2048 - 2059
  • [9] Simulated Online Adaptive Magnetic Resonance-Guided Stereotactic Body Radiation Therapy for the Treatment of Oligometastatic Disease of the Abdomen and Central Thorax: Characterization of Potential Advantages
    Henke, Lauren
    Kashani, Rojano
    Yang, Deshan
    Zhao, Tianyu
    Green, Olga
    Olsen, Lindsey
    Rodriguez, Vivian
    Wooten, H. Omar
    Li, H. Harold
    Hu, Yanle
    Bradley, Jeffrey
    Robinson, Clifford
    Parikh, Parag
    Michalski, Jeff
    Mutic, Sasa
    Olsen, Jeffrey R.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2016, 96 (05): : 1078 - 1086
  • [10] Task Group 142 report: Quality assurance of medical accelerators
    Klein, Eric E.
    Hanley, Joseph
    Bayouth, John
    Yin, Fang-Fang
    Simon, William
    Dresser, Sean
    Serago, Christopher
    Aguirre, Francisco
    Ma, Lijun
    Arjomandy, Bijan
    Liu, Chihray
    Sandin, Carlos
    Holmes, Todd
    [J]. MEDICAL PHYSICS, 2009, 36 (09) : 4197 - 4212