Contrast-to-Noise Ratio and Low-Contrast Object Resolution on Full- and Low-Dose MDCT: SAFIRE Versus Filtered Back Projection in a Low-Contrast Object Phantom and in the Liver
被引:152
作者:
Baker, Mark E.
[1
,2
]
Dong, Frank
[1
]
Primak, Andrew
[3
]
Obuchowski, Nancy A.
[4
]
Einstein, David
[1
]
Gandhi, Namita
[1
]
Herts, Brian R.
[1
]
Purysko, Andrei
[1
]
Remer, Erick
[1
]
Vachani, Neil
[1
]
机构:
[1] Cleveland Clin, Imaging Inst, Cleveland, OH 44195 USA
[2] Cleveland Clin, Inst Digest Dis, Cleveland, OH 44106 USA
[3] Siemens Healthcare, Malvern, PA USA
[4] Cleveland Clin, Dept Quantitat Hlth Sci, Cleveland, OH 44106 USA
关键词:
contrast-to-noise ratio;
iterative reconstruction;
liver;
low-contrast object detection;
MDCT;
STATISTICAL ITERATIVE RECONSTRUCTION;
CORONARY CT ANGIOGRAPHY;
IMAGE-QUALITY;
COMPUTED-TOMOGRAPHY;
DETAIL CURVES;
REDUCTION;
PERFORMANCE;
ALGORITHM;
SELECTION;
CHEST;
D O I:
10.2214/AJR.11.7421
中图分类号:
R8 [特种医学];
R445 [影像诊断学];
学科分类号:
1002 ;
100207 ;
1009 ;
摘要:
OBJECTIVE. The purpose of this article is to evaluate the effect of sinogram-affirmed iterative reconstruction (SAFIRE) on contrast-to-noise ratio (CNR) compared with filtered back projection (FBP) and to determine whether SAFIRE improves low-contrast object detection or conspicuity in a low-contrast object phantom and in the liver on full-and low-dose examinations. SUBJECTS AND METHODS. A low-contrast object phantom was scanned at 100%, 70%, 50%, and 30% dose using a single-source made of a dual-source MDCT scanner, with the raw data reconstructed with SAFIRE and FBP. Unenhanced liver CT scans in 22 patients were performed using a dual-source MDCT. The raw data from both tubes (100% dose) were reconstructed using FBP, and data from one tube (50% dose) were reconstructed using both FBP and SAFIRE. CNR was measured in the phantom and in the liver. Noise, contrast, and CNR were compared using paired Student t tests. Six readers assessed sphere detection and conspicuity in the phantom and liver-inferior vena cava conspicuity in the patient data. The phantom and patient data were assessed using multiple-variable logistic regression. RESULTS. The phantom at 70% and 50% doses with SAFIRE had decreased noise and increased CNR compared with the 100% dose with FBP. In the liver, the mean CNR improvement at 50% dose with SAFIRE compared with FBP was 31.4% and 88% at 100% and 50% doses, respectively (p < 0.001). Sphere object detection and conspicuity improved with SAFIRE (p < 0.001). However, smaller spheres were obscured on both FBP and SAFIRE images at lower doses. Liver-vessel conspicuity improved with SAFIRE over 50%-dose FBP in 67.4% of cases (p < 0.001), and versus 100%-dose FBP, improved in 38.6% of cases (p = 0.085). As a predictor for detection, CNR alone had a discriminatory ability (c-index, 0.970) similar to that of the model that analyzed dose, lesion size, attenuation difference, and reconstruction technique (c-index, 0.978). CONCLUSION. Lower dose scans reconstructed with SAFIRE have a higher CNR. The ability of SAFIRE to improve low-contrast object detection and conspicuity depends on the radiation dose level. At low radiation doses, low-contrast objects are invisible, regardless of reconstruction technique.