Contrast-to-Noise Ratio and Low-Contrast Object Resolution on Full- and Low-Dose MDCT: SAFIRE Versus Filtered Back Projection in a Low-Contrast Object Phantom and in the Liver

被引:152
作者
Baker, Mark E. [1 ,2 ]
Dong, Frank [1 ]
Primak, Andrew [3 ]
Obuchowski, Nancy A. [4 ]
Einstein, David [1 ]
Gandhi, Namita [1 ]
Herts, Brian R. [1 ]
Purysko, Andrei [1 ]
Remer, Erick [1 ]
Vachani, Neil [1 ]
机构
[1] Cleveland Clin, Imaging Inst, Cleveland, OH 44195 USA
[2] Cleveland Clin, Inst Digest Dis, Cleveland, OH 44106 USA
[3] Siemens Healthcare, Malvern, PA USA
[4] Cleveland Clin, Dept Quantitat Hlth Sci, Cleveland, OH 44106 USA
关键词
contrast-to-noise ratio; iterative reconstruction; liver; low-contrast object detection; MDCT; STATISTICAL ITERATIVE RECONSTRUCTION; CORONARY CT ANGIOGRAPHY; IMAGE-QUALITY; COMPUTED-TOMOGRAPHY; DETAIL CURVES; REDUCTION; PERFORMANCE; ALGORITHM; SELECTION; CHEST;
D O I
10.2214/AJR.11.7421
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
OBJECTIVE. The purpose of this article is to evaluate the effect of sinogram-affirmed iterative reconstruction (SAFIRE) on contrast-to-noise ratio (CNR) compared with filtered back projection (FBP) and to determine whether SAFIRE improves low-contrast object detection or conspicuity in a low-contrast object phantom and in the liver on full-and low-dose examinations. SUBJECTS AND METHODS. A low-contrast object phantom was scanned at 100%, 70%, 50%, and 30% dose using a single-source made of a dual-source MDCT scanner, with the raw data reconstructed with SAFIRE and FBP. Unenhanced liver CT scans in 22 patients were performed using a dual-source MDCT. The raw data from both tubes (100% dose) were reconstructed using FBP, and data from one tube (50% dose) were reconstructed using both FBP and SAFIRE. CNR was measured in the phantom and in the liver. Noise, contrast, and CNR were compared using paired Student t tests. Six readers assessed sphere detection and conspicuity in the phantom and liver-inferior vena cava conspicuity in the patient data. The phantom and patient data were assessed using multiple-variable logistic regression. RESULTS. The phantom at 70% and 50% doses with SAFIRE had decreased noise and increased CNR compared with the 100% dose with FBP. In the liver, the mean CNR improvement at 50% dose with SAFIRE compared with FBP was 31.4% and 88% at 100% and 50% doses, respectively (p < 0.001). Sphere object detection and conspicuity improved with SAFIRE (p < 0.001). However, smaller spheres were obscured on both FBP and SAFIRE images at lower doses. Liver-vessel conspicuity improved with SAFIRE over 50%-dose FBP in 67.4% of cases (p < 0.001), and versus 100%-dose FBP, improved in 38.6% of cases (p = 0.085). As a predictor for detection, CNR alone had a discriminatory ability (c-index, 0.970) similar to that of the model that analyzed dose, lesion size, attenuation difference, and reconstruction technique (c-index, 0.978). CONCLUSION. Lower dose scans reconstructed with SAFIRE have a higher CNR. The ability of SAFIRE to improve low-contrast object detection and conspicuity depends on the radiation dose level. At low radiation doses, low-contrast objects are invisible, regardless of reconstruction technique.
引用
收藏
页码:8 / 18
页数:11
相关论文
共 49 条
  • [1] Amis E Stephen Jr, 2007, J Am Coll Radiol, V4, P272, DOI 10.1016/j.jacr.2007.03.002
  • [2] [Anonymous], 2009, J RADIOLOGICAL PROTE
  • [3] Barrett H.H., 1996, Radiological imaging: the theory of image formation, detection, and processing
  • [4] Current concepts - Computed tomography - An increasing source of radiation exposure
    Brenner, David J.
    Hall, Eric J.
    [J]. NEW ENGLAND JOURNAL OF MEDICINE, 2007, 357 (22) : 2277 - 2284
  • [5] A review of image denoising algorithms, with a new one
    Buades, A
    Coll, B
    Morel, JM
    [J]. MULTISCALE MODELING & SIMULATION, 2005, 4 (02) : 490 - 530
  • [6] Chao EH, 2000, MEDICINE, V217, P262
  • [7] Projected Cancer Risks From Computed Tomographic Scans Performed in the United States in 2007
    de Gonzalez, Amy Berrington
    Mahesh, Mahadevappa
    Kim, Kwang-Pyo
    Bhargavan, Mythreyi
    Lewis, Rebecca
    Mettler, Fred
    Land, Charles
    [J]. ARCHIVES OF INTERNAL MEDICINE, 2009, 169 (22) : 2071 - 2077
  • [8] Reducing the Radiation Dose for CT Colonography Using Adaptive Statistical Iterative Reconstruction: A Pilot Study
    Flicek, Kristina T.
    Hara, Amy K.
    Silva, Alvin C.
    Wu, Qing
    Peter, Mary B.
    Johnson, C. Daniel
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 2010, 195 (01) : 126 - 131
  • [9] Ghetti C., 2011, PHYS MED
  • [10] FACTORS RELATED TO LOW CONTRAST RESOLUTION IN CT SCANNERS
    GOODENOUGH, DJ
    WEAVER, KE
    [J]. COMPUTERIZED RADIOLOGY, 1984, 8 (05): : 297 - 308