PURPOSE: To compare visual,: refractive, and safety outcomes among a non-wavefront scanning-slit laser (NIDEK EC-5000) and wavefront-driven lasers (Alcon CustomCornea and VISX CustomVue). METHODS: A retrospective comparison of outcomes for 290 eyes that underwent LASIK for myopia and myopic astigmatism with either a conventional or custom ablation excimer laser system were compared. The preoperative refractive error and, age of the patients were matched. Outcomes were tested for, statistically significant differences among the conventional laser and each of the custom ablation lasers. A P value <.05 was considered statistically significant. Data with 3-month follow-up are reported. RESULTS: Postoperatively, mean manifest refraction spherical equivalent was -0.12 +/- 0.31 diopters (D) (range: -1.50 to 0.75 D) for the NIDEK group, -0.13 +/- 0.39 D (range: -1.88 to 0.75 D) for the VISX group, and -0.06 +/- 0.26 D (range: -0.75 to 0.75 D) for the Alcon group. Eighty-nine percent of eyes in the NIDEK group, 88% of eyes in the VISX group, and 92% of eyes in the Alcon group were within a half-diopter of intended correction. None of the NIDEK and VISX eyes and 3% of Alcon eyes lost 2 or more lines of best spectacle-corrected visual acuity. No statistically significant differences in any of the outcomes studied. between conventional and custom ablation treatments were noted (P >.05). CONCLUSIONS: No statistically. significant differences were observed in safety, efficacy, or predictability of:eyes that underwent conventional ablation, with the NIDEK EC-5000 scanning slit laser compared to custom ablation with the VISX CustomVue or Alcon CustomCornea laser systems. [J Refract,Surg. 2008;24:891-896.]
引用
收藏
页码:891 / 896
页数:6
相关论文
共 7 条
  • [1] Bilateral comparison of conventional versus topographic-guided customized ablation for myopic LASIK with the NIDEK EC-5000
    Du, Chi-xin
    Yang, Ya-bo
    Shen, Ye
    Wang, Yang
    Dougherty, Paul J.
    [J]. JOURNAL OF REFRACTIVE SURGERY, 2006, 22 (07) : 642 - 646
  • [2] Comparison of LASIK using the NIDEK EC-5000 optimized aspheric transition zone (OATz) and conventional ablation profile
    Hori-Komai, Yoshiko
    Toda, Ikuko
    Asano-Kato, Naoko
    Ito, Mitsutoshi
    Yamamoto, Takahiro
    Tsubota, Kazuo
    [J]. JOURNAL OF REFRACTIVE SURGERY, 2006, 22 (06) : 546 - 555
  • [3] Kim TI, 2004, J REFRACT SURG, V20, P432
  • [4] Wavefront-guided ablation: Evidence for efficacy compared to traditional ablation
    Netto, MV
    Dupps, W
    Wilson, SE
    [J]. AMERICAN JOURNAL OF OPHTHALMOLOGY, 2006, 141 (02) : 360 - 368
  • [5] Wavefront-guided versus standard laser in situ keratomileusis to correct low to moderate myopia
    Nuijts, RMMA
    Nabar, VA
    Hament, WJ
    Eggink, FAGJ
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2002, 28 (11) : 1907 - 1913
  • [6] Visual experience and patient satisfaction with wavefront-guided laser in situ keratomileusis
    Tuan, KMA
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2006, 32 (04) : 577 - 583
  • [7] Waheed Samra, 2003, Curr Opin Ophthalmol, V14, P198, DOI 10.1097/00055735-200308000-00005