A Nonocclusive, Inexpensive Pediatric Pulsatile Roller Pump for Cardiopulmonary Bypass, Extracorporeal Life Support, and Left/Right Ventricular Assist Systems

被引:5
作者
Wang, Shigang [1 ]
Durandy, Yves [2 ]
Kunselman, Allen R. [3 ]
Uendar, Akif [1 ]
机构
[1] Penn State Hershey Childrens Hosp, Penn State Hershey Coll Med, Penn State Hershey Pediat Cardiovasc Res Ctr, Dept Pediat,Dept Surg & Bioengn, Hershey, PA USA
[2] Penn State Hershey Childrens Hosp, Penn State Hershey Coll Med, Penn State Milton S Hershey Med Ctr, Hershey, PA USA
[3] Inst Cardiovasc Paris Sud, Massy, France
关键词
Nonocclusive roller pump; Pulsatile flow; Cardiopulmonary bypass; Extracorporeal life support; Ventricular assist system; Pediatrics; CLINICAL-OUTCOMES; LUNG SUPPORT; FLOW; PERFUSION; CIRCUIT; IMPACT;
D O I
10.1111/aor.12026
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A simple, inexpensive pediatric pulsatile roller blood pump has been utilized for routine cardiopulmonary bypass (CPB) procedures, extracorporeal life support (ECLS), and left/right ventricular assist systems (LVAS/RVAS) for decades in France. This particular nonocclusive pulsatile system has many advantages including several safety features for patients as well as an extremely lower cost. The objective of this study is to evaluate the performance of this particular system for CPB, ECLS, and LVAS/RVAS in pulsatile mode. This pediatric nonocclusive system was evaluated with pump flow rates of 500, 750, and 1000?mL/min under normothermic (35 degrees C) and hypothermic (25 degrees C) conditions in CPB, ECLS, and LVAS/RVAS circuits using clinical disposables and settings. Energy equivalent pressure (EEP), surplus homodynamic energy (SHE), and total hemodynamic energy (THE) were calculated for each experimental stage. The pump generated near physiological quality of pulsatile flow without backflow in the three simulated pediatric circuits. With increased flow rates, more hemodynamic energy was delivered to the pseudo patient. This particular nonocclusive pediatric pulsatile system performed well during all of the experimental conditions and generated adequate quality pulsatile pressure-flow waveforms using CPB, ECLS, and LVAS/RVAS circuitry. Although this novel concept was first introduced in the 1990s, we believe that there is still need for this technology (with engineering modifications) because of significant advantages including safety and cost.
引用
收藏
页码:48 / 56
页数:9
相关论文
共 16 条
[1]   Evaluation of Perfusion Modes on Vital Organ Recovery and Thyroid Hormone Homeostasis in Pediatric Patients Undergoing Cardiopulmonary Bypass [J].
Akcevin, Atif ;
Alkan-Bozkaya, Tijen ;
Qiu, Feng ;
Uendar, Akif .
ARTIFICIAL ORGANS, 2010, 34 (11) :879-884
[2]   Impact of Pulsatile Perfusion on Clinical Outcomes of Neonates and Infants With Complex Pathologies Undergoing Cardiopulmonary Bypass Procedures [J].
Alkan-Bozkaya, Tijen ;
Akcevin, Atif ;
Turkoglu, Halil ;
Uendar, Akif .
ARTIFICIAL ORGANS, 2013, 37 (01) :82-86
[3]   EXTRACORPOREAL LUNG SUPPORT FOR NEONATAL ACUTE RESPIRATORY-FAILURE - PRELIMINARY-REPORT [J].
CHEVALIER, JY ;
DURANDY, Y ;
BATISSE, A ;
MATHE, JC ;
COSTIL, J .
LANCET, 1990, 335 (8702) :1364-1366
[4]  
DURANDY Y, 1990, J THORAC CARDIOV SUR, V99, P404
[5]   Comparison of Parameters for Detection of Splanchnic Hypoxia in Children Undergoing Cardiopulmonary Bypass With Pulsatile Versus Nonpulsatile Normothermia or Hypothermia During Congenital Heart Surgeries [J].
Karaci, Ali Riza ;
Sasmazel, Ahmet ;
Aydemir, Numan Ali ;
Saritas, Turkay ;
Harmandar, Bugra ;
Tuncel, Zeliha ;
Uendar, Akif .
ARTIFICIAL ORGANS, 2011, 35 (11) :1010-1017
[6]   Extracorporeal life support systems: alternative vs. conventional circuits [J].
Khan, Sameer ;
Vasavada, Rahul ;
Qiu, Feng ;
Kunselman, Allan ;
Uendar, Akif .
PERFUSION-UK, 2011, 26 (03) :191-198
[7]   Pulsatile flow during cardiopulmonary bypass preserves postoperative microcirculatory perfusion irrespective of systemic hemodynamics [J].
Koning, Nick J. ;
Vonk, Alexander B. A. ;
van Barneveld, Lerau J. ;
Beishuizen, Albertus ;
Atasever, Bektas ;
van den Brom, Charissa E. ;
Boer, Christa .
JOURNAL OF APPLIED PHYSIOLOGY, 2012, 112 (10) :1727-1734
[8]  
O'Neil MP, 2012, ANN THORAC SURG
[9]   Evaluation of two pediatric polymethyl pentene membrane oxygenators with pulsatile and non-pulsatile perfusion [J].
Qiu, Feng ;
Khan, Sameer ;
Talor, Jonathan ;
Kunselman, Allan ;
Uendar, Akif .
PERFUSION-UK, 2011, 26 (03) :229-237
[10]  
SHEPARD RB, 1966, ARCH SURG-CHICAGO, V93, P730