Case Report: Failure of Rotating-hinge Total Knee Prosthesis by Disengagement of the Hinge-post Extension

被引:10
作者
Biswas, Debdut [1 ]
Haughom, Bryan [1 ]
Mayle, Robert E., Jr. [1 ]
Della Valle, Craig J. [1 ]
机构
[1] Rush Univ Med Ctr, Dept Orthopaed Surg, Sect Adult Reconstruct, Chicago, IL 60612 USA
关键词
ARTHROPLASTY; REPLACEMENT; DISLOCATION;
D O I
10.1007/s11999-012-2736-2
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Although modern-generation rotating-hinge knee implants have been modified to mitigate earlier complications related to aseptic loosening with these devices in revision TKAs, there are few, if any, reports of failure related to the hinge mechanism in these prostheses. We present a case of disengagement of the hinge-post extension in a contemporary rotating-hinge knee prosthesis originally implanted during revision surgery for instability and extensor mechanism insufficiency. We hypothesize the screw home kinematics of the knee resulted in a counterclockwise torque, contributing to the hinge post locking screw disengagement in this case. Although some authors have reported complications and failures related to aseptic loosening with rotating-hinge knee prostheses, there have been no reports regarding complications related to the hinge mechanism, which simultaneously links and permits rotation between the femoral and tibial components. Arthroplasty surgeons who use rotating-hinge knee prostheses need to be aware of this potential mode of failure. Design modifications, including more secure locking mechanisms and side-specific implants, may avoid this complication.
引用
收藏
页码:1389 / 1392
页数:4
相关论文
共 21 条
  • [1] Barrack RL, 2001, CLIN ORTHOP RELAT R, P292
  • [2] The use of a modular rotating hinge component in salvage revision total knee arthroplasty
    Barrack, RL
    Lyons, TR
    Ingraham, RQ
    Johnson, JC
    [J]. JOURNAL OF ARTHROPLASTY, 2000, 15 (07) : 858 - 866
  • [3] Beauchamp CP, 2011, J ARTHROPLASTY, V26
  • [4] Rotating-hinge Total Knee for Revision Total Knee Arthroplasty
    Bistolfi, Alessandro
    Massazza, Giuseppe
    Rosso, Federica
    Crova, Maurizio
    [J]. ORTHOPEDICS, 2012, 35 (03) : E325 - E330
  • [5] Browne JA, 2011, J BONE JOINT SURG AM, V93A, P1137, DOI [10.2106/JBJSJ.01036, 10.2106/JBJS.J.01036]
  • [6] Burnett R Stephen J, 2005, J Bone Joint Surg Am, V87 Suppl 1, P175, DOI 10.2106/JBJS.E.00442
  • [7] The role of the rotating hinge prosthesis in the salvage arthroplasty setting
    Deehan, David John
    Murray, James
    Birdsall, Paul D.
    Holland, James P.
    Pinder, Ian M.
    [J]. JOURNAL OF ARTHROPLASTY, 2008, 23 (05) : 683 - 688
  • [8] A multicenter analysis of axial femorotibial rotation after total knee arthroplasty
    Dennis, DA
    Komistek, RD
    Mahfouz, MR
    Walker, SA
    Tucker, A
    [J]. CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2004, (428) : 180 - 189
  • [9] Failure rate of a rotating hinge knee design due to yoke fracture of the hinged tibial insert: a retrospective data analysis and review of the literature
    Friesenbichler, Joerg
    Schwarzkopf, Ran
    Sadoghi, Patrick
    Marwin, Scott E.
    Glehr, Mathias
    Maurer-Ertl, Werner
    Leithner, Andreas
    [J]. INTERNATIONAL ORTHOPAEDICS, 2012, 36 (05) : 993 - 998
  • [10] Hinged Total Knee Arthroplasty in the Presence of Ligamentous Deficiency
    Hernandez-Vaquero, Daniel
    Sandoval-Garcia, Manuel A.