Instrument-free argentometric determination of chloride via trapezoidal distance-based microfluidic paper devices

被引:53
作者
Rahbar, Mohammad [1 ]
Paull, Brett [1 ]
Macka, Mirek [1 ,2 ,3 ]
机构
[1] Univ Tasmania, Sch Nat Sci, ACROSS, Hobart, Tas 7001, Australia
[2] Mendel Univ Brno, Dept Chem & Biochem, Zemedelska 1, CZ-61300 Brno, Czech Republic
[3] Brno Univ Technol, Cent European Inst Technol, Purkynova 123, CZ-61200 Brno, Czech Republic
基金
澳大利亚研究理事会;
关键词
Microfluidic paper-based analytical device; Argentometric determination; Distance-based measurement; Chloride determination; Asymmetric trapezoidal microfluidic channels; MU-PADS; FABRICATION; ASSAY;
D O I
10.1016/j.aca.2019.02.048
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Chloride (Cl-) is an inorganic anion present in a broad range of samples (e.g. biological, environmental, food, water, etc.), the determination of which is of widespread significance. In this work, we translate the well-established traditional argentometric method (Mohr's precipitation titration) into a small, simple, portable, and low-cost paper-based microfluidic diagnostic device, which provides rapid and quantitative analysis. The developed device enables the determination of chloride sample volumes as small as 5 mu L. A distance-based detection method is implemented providing fully instrument-free quantitation. The beneficial effects of channel geometry (variable widths with constant heights) on analytical parameters were investigated. Trapezoidal channels (channel width changes linearly with height) were used to create a gradient of paper surface (titrant) available for the reaction, compared to the typical uniform rectangular channels (constant channel width). The trapezoid with increasing width offered higher sensitivity and lower detection limits (i.e. 0.05 mM vs 0.1 mM from the rectangular channel) for chloride determination across the concentration range of 0.05-25 mM. In addition, the effect of concentration of the deposited reagent on the obtained distance signals was investigated using varying concentrations of titrant (AgNO3), which allowed determination of chloride across a wider dynamic range (up to 200 mM). The utility of the paper devices was demonstrated by determination of chloride in a variety of matrices including body fluids (sweat, serum, and urine) and water samples (drinking, mineral, river water). (C) 2019 Elsevier B.V. All rights reserved.
引用
收藏
页码:1 / 8
页数:8
相关论文
共 37 条
[1]  
Ann Terry Carol, 1993, U. S. Patent, Patent No. 5July201993229
[2]   Chloride: The queen of electrolytes? [J].
Berend, Kenrick ;
van Hulsteijn, Leonard Hendrik ;
Gans, Rijk O. B. .
EUROPEAN JOURNAL OF INTERNAL MEDICINE, 2012, 23 (03) :203-211
[3]  
Brunzel NA., 2016, Fundamentals of Urine and Body Fluid Analysis, V4th
[4]   Multiplexed paper analytical device for quantification of metals using distance-based detection [J].
Cate, David M. ;
Noblitt, Scott D. ;
Volckens, John ;
Henry, Charles S. .
LAB ON A CHIP, 2015, 15 (13) :2808-2818
[5]   Recent Developments in Paper-Based Microfluidic Devices [J].
Cate, David M. ;
Adkins, Jaclyn A. ;
Mettakoonpitak, Jaruwan ;
Henry, Charles S. .
ANALYTICAL CHEMISTRY, 2015, 87 (01) :19-41
[6]   Simple, distance-based measurement for paper analytical devices [J].
Cate, David M. ;
Dungchai, Wijitar ;
Cunningham, Josephine C. ;
Volckens, John ;
Henry, Charles S. .
LAB ON A CHIP, 2013, 13 (12) :2397-2404
[7]   Distance-based carcinoembryonic antigen assay on microfluidic paper immunodevice [J].
Chen, Ying ;
Chu, Weiru ;
Liu, Wei ;
Guo, Xiaoyan .
SENSORS AND ACTUATORS B-CHEMICAL, 2018, 260 :452-459
[8]  
Crompton T. R., 2014, DETERMINATION ANIONS
[9]  
Federation W. E., 2005, STANDARD METHODS EXA
[10]   Selective Distance-Based K+ Quantification on Paper-Based Microfluidics [J].
Gerold, Chase T. ;
Bakker, Eric ;
Henry, Charles S. .
ANALYTICAL CHEMISTRY, 2018, 90 (07) :4894-4900