Effect of Optic Disc Size and Disease Severity on the Diagnostic Capability of Glaucoma Imaging Technologies in an Indian Population

被引:11
作者
Garudadri, Chandra S. [1 ]
Rao, Harsha L. [1 ]
Parikh, Rajul S. [1 ,2 ,3 ]
Jonnadula, Ganesh B. [1 ]
Selvaraj, Prabhakaran [1 ]
Nutheti, Rishita [1 ]
Thomas, Ravi [1 ,4 ,5 ]
机构
[1] LV Prasad Eye Inst, Hyderabad 500034, Andhra Pradesh, India
[2] Bombay City Eye Inst, Bombay, Maharashtra, India
[3] Res Ctr, Bombay, Maharashtra, India
[4] Queensland Eye Inst, Brisbane, Qld, Australia
[5] Univ Queensland, Brisbane, Qld 4072, Australia
关键词
stratus OCT; GDx VCC; HRT; glaucoma imaging devices; NERVE-FIBER LAYER; RACIAL-DIFFERENCES; SENSITIVITY; DIAMETER;
D O I
10.1097/IJG.0b013e31821829f1
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To evaluate the influence of optic disc size and disease severity on the diagnostic validity of optical coherence tomography (Stratus OCT), scanning laser polarimetry [GDx variable corneal compensator (VCC)], and confocal scanning laser ophthalmoscopy [Heidelberg retina tomograph II (HRT II)] in Indian eyes with glaucoma. Methods: Ninety-five normal and 125 glaucoma patients underwent imaging with Stratus OCT, GDx VCC, and HRT II. One eye of each person was randomly selected for analysis. Using disc area determined on HRT II, discs were classified as small (< 2 mm(2)), moderate (2 to 3 mm(2)), and large (> 3 mm(2)). The parameter with the best sensitivity for each device, at a fixed specificity, was compared for different disc sizes. Logistic marginal regression was used to study the effect of disc size and disease severity (mean deviation on standard automated perimetry) on the diagnostic performance of these imaging devices. Results: At a fixed specificity of 84.2%, the sensitivity of HRT II was significantly different for varying disc sizes (P = 0.0004). The sensitivities for dissimilar disc sizes were not significantly different for the GDx VCC (P = 0.928) or Stratus OCT (P = 0.381). Logistic marginal regression also showed that sensitivity of HRT II increased with increasing disc size, whereas sensitivity of OCT and GDx were independent of the disc size. The sensitivity of all 3 technologies increased with increasing disease severity (decreasing mean deviation). Conclusions: Optic disc size affects the diagnostic capability of HRT II but not that of GDx VCC or Stratus OCT. The sensitivity of all 3 imaging technologies increased with increasing disease severity.
引用
收藏
页码:475 / 480
页数:6
相关论文
共 31 条
[1]  
Agarwal Harish C, 2003, Indian J Ophthalmol, V51, P25
[2]  
*AM AC OPHTH GLAUC, 2003, PREF PRACT PATT PRIM
[3]  
Anderson DRPV, 1999, Automated Static Perimetry, V2nd, P121
[4]   Optic disk and nerve fiber layer Imaging to detect glaucoma [J].
Badala, Federico ;
Nouri-Mahdavi, Kouros ;
Raoof, Duna A. ;
Leeprechanon, Narakorn ;
Law, Simon K. ;
Caprioli, Joseph .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 2007, 144 (05) :724-732
[5]   Sensitivity and specificity of the StratusOCT for perimetric glaucoma [J].
Budenz, DL ;
Michael, A ;
Chang, RT ;
McSoley, J ;
Katz, J .
OPHTHALMOLOGY, 2005, 112 (01) :3-9
[6]   RACIAL-DIFFERENCES IN OPTIC-NERVE HEAD PARAMETERS [J].
CHI, T ;
RITCH, R ;
STICKLER, D ;
PITMAN, B ;
TSAI, C ;
HSIEH, FY .
ARCHIVES OF OPHTHALMOLOGY, 1989, 107 (06) :836-839
[7]   THE LENS OPACITIES CLASSIFICATION SYSTEM-III [J].
CHYLACK, LT ;
WOLFE, JK ;
SINGER, DM ;
LESKE, MC ;
BULLIMORE, MA ;
BAILEY, IL ;
FRIEND, J ;
MCCARTHY, D ;
WU, SY .
ARCHIVES OF OPHTHALMOLOGY, 1993, 111 (06) :831-836
[8]  
Healey PRJ, 1997, AUST NZ J OPHTHALMOL, V25, pS99
[9]   OPTICAL COHERENCE TOMOGRAPHY [J].
HUANG, D ;
SWANSON, EA ;
LIN, CP ;
SCHUMAN, JS ;
STINSON, WG ;
CHANG, W ;
HEE, MR ;
FLOTTE, T ;
GREGORY, K ;
PULIAFITO, CA ;
FUJIMOTO, JG .
SCIENCE, 1991, 254 (5035) :1178-1181
[10]  
JONAS JB, 1992, INVEST OPHTH VIS SCI, V33, P2012