Quality of the Threshold Algorithm in Age-Related Macular Degeneration: Stratus versus Cirrus OCT

被引:23
作者
Krebs, Ilse [1 ,2 ]
Falkner-Radler, Christiane [1 ,2 ]
Hagen, Stefan [2 ]
Haas, Paulina [1 ,2 ]
Brannath, Werner [3 ]
Lie, Shilla [2 ]
Ansari-Shahrezaei, Siamak [1 ,2 ]
Binder, Susanne [1 ,2 ]
机构
[1] Rudolf Fdn Clin, Dept Ophthalmol, A-1030 Vienna, Austria
[2] Ludwig Boltzmann Inst Retinol & Biomicroscop Lase, Vienna, Austria
[3] Med Univ, Core Unit Med Stat & Informat, Vienna, Austria
关键词
OPTICAL COHERENCE TOMOGRAPHY; THICKNESS MEASUREMENTS; REPEATABILITY; DOMAIN; BEVACIZUMAB;
D O I
10.1167/iovs.08-2617
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
PURPOSE. Automatically generated measurements of the retinal volume or the central retinal thickness are based on correctly set threshold lines on the retinal surface and the retinal pigment epithelium. The purpose of this study was to compare the accuracy of threshold algorithm lines of Stratus optical coherence tomography (OCT) with those of Cirrus OCT. METHODS. A consecutive series of patients at least 50 years of age with exudative age-related macular degeneration was included. Stratus OCT (retinal thickness program) and Cirrus OCT (macular cube 512 x 128) were performed by the same examiner, the sequence of the examinations was randomized. Two independent examiners evaluated the positioning of the threshold algorithm lines and performed a grading of the failures. Logistic regression analysis was applied for evaluation of the failure rate. RESULTS. One hundred four patients were included. For the entire OCT examination (6 scans Stratus OCT, 128 scans Cirrus OCT) algorithm line failures were detected in 69.2% of the Stratus OCT and in 25% of the Cirrus OCT examinations, with the difference reaching statistical significance (P < 0.001). The median failure grade was 1 (0-6) for Stratus and 0 (0-5.15) for Cirrus OCT. Age, measurement sequence, and investigator did not influence the error rates. CONCLUSIONS. With Cirrus OCT automatically performed and therefore objective measurements of central retinal thickness and retinal volume were provided correctly in 69.2% of the scans. Furthermore, this latest software version offers the possibility of manual correction of false positioned algorithm lines (ClinicalTrials. gov number, NCT00568191). (Invest Ophthalmol Vis Sci. 2009; 50: 995-1000) DOI: 10.1167/iovs.08-2617
引用
收藏
页码:995 / 1000
页数:6
相关论文
共 20 条
[1]   Automatic segmentation in three-dimensional analysis of fibrovascular pigmentepithelial detachment using high-definition optical coherence tomography [J].
Ahlers, C. ;
Simader, C. ;
Geitzenauer, W. ;
Stock, G. ;
Stetson, P. ;
Dastmalchi, S. ;
Schmidt-Erfurth, U. .
BRITISH JOURNAL OF OPHTHALMOLOGY, 2008, 92 (02) :197-203
[2]   High-resolution spectral domain-OCT imaging in geographic atrophy associated with age-related macular degeneration [J].
Fleckenstein, Monika ;
Issa, Peter Charbel ;
Helb, Hans-Martin ;
Schmitz-Valckenberg, Steffen ;
Finger, Robert P. ;
Scholl, Hendrik P. N. ;
Loeffler, Karin U. ;
Holz, Frank G. .
INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2008, 49 (09) :4137-4144
[3]   An optical coherence tomography-guided, variable dosing regimen with intravitreal ranibizumab (lucentis) for neovascular age-related macular degeneration [J].
Fung, Anne E. ;
Lalwani, Geeta A. ;
Rosenfeld, Philip J. ;
Dubovy, Sander R. ;
Michels, Stephan ;
Feuer, William J. ;
Puliafito, Carmen A. ;
Davis, Janet L. ;
Flynn, Harry W., Jr. ;
Esquiabro, Maria .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 2007, 143 (04) :566-583
[4]   Quantitative comparison of optical coherence tomography after pegaptanib or bevacizumab in neovascular age-related macular degeneration [J].
Joeres, Sandra ;
Kaplowitz, Kevin ;
Brubaker, Jacob W. ;
Updike, Paul G. ;
Collins, Allyson T. ;
Walsh, Alexander C. ;
Romano, Peggy W. ;
Sadda, Srinivas R. .
OPHTHALMOLOGY, 2008, 115 (02) :347-354
[5]  
KO TH, 2005, OPHTHALMOLOGY, V112
[6]  
Koozekanani D, 2000, INVEST OPHTH VIS SCI, V41, P1486
[7]   Optical coherence tomography: limits of the retinal-mapping program in age-related macular degeneration [J].
Krebs, I. ;
Haas, P. ;
Zeiler, F. ;
Binder, S. .
BRITISH JOURNAL OF OPHTHALMOLOGY, 2008, 92 (07) :933-935
[8]   Optical coherence tomography guided retreatment of photodynamic therapy [J].
Krebs, I ;
Binder, S ;
Stolba, U ;
Schmid, K ;
Glittenberg, C ;
Brannath, W ;
Goll, A .
BRITISH JOURNAL OF OPHTHALMOLOGY, 2005, 89 (09) :1184-1187
[9]   Activity of neovascular lesions treated with bevacizumab: comparison between optical coherence tomography and fluorescein angiography [J].
Krebs, Ilse ;
Ansari-Shahrezaei, Siamak ;
Goll, Alexandra ;
Binder, Susanne .
GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2008, 246 (06) :811-815
[10]   Comparison of Macular Thickness Measurements between Time Domain and Spectral Domain Optical Coherence Tomography [J].
Leung, Christopher Kai-shun ;
Cheung, Carol Yim-lui ;
Weinreb, Robert N. ;
Lee, Gary ;
Lin, Dusheng ;
Pang, Chi Pui ;
Lam, Dennis S. C. .