Radiotherapy electron beams collimated by small tubular applicators: characterization by silicon and diamond diodes

被引:13
作者
Bagala, P. [1 ]
Di Venanzio, C. [2 ]
Falco, M. D. [1 ]
Guerra, A. S. [3 ]
Marinelli, Marco [2 ]
Milani, E. [2 ]
Pimpinella, M. [3 ]
Pompili, F. [2 ]
Prestopino, G. [2 ]
Santoni, R. [1 ]
Tonnetti, A. [2 ]
Verona, C. [2 ]
Verona-Rinati, G. [2 ]
机构
[1] Tor Vergata Univ Gen Hosp, Dept Diagnost Imaging Mol Imaging Intervent Radio, I-00133 Rome, Italy
[2] Univ Roma Tor Vergata, Dipartimento Ingn Ind, Ist Nazl Fis Nucl, I-00133 Rome, Italy
[3] ENEA INMRI CR Casaccia, Ist Nazl Metrol Radiaz Ionizzanti, I-00123 Rome, Italy
关键词
D-W MEASUREMENTS; DOSIMETRIC CHARACTERISTICS; OUTPUT FACTORS; PHOTON BEAMS; DOSE-RATE; DETECTOR; ENERGY;
D O I
10.1088/0031-9155/58/22/8121
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
High-energy electron beams generated by linear accelerators, typically in the range 6 to 20 MeV, are used in small field sizes for radiotherapy of localized superficial tumors. Unshielded silicon diodes (Si-D) are commonly considered suitable detectors for relative dose measurements in small electron fields due to their high spatial resolution. Recently, a novel synthetic single crystal diamond diode (SCDD) showed suitable properties for standard electron beams and small photon beams dosimetry. The aim of the present study is twofold: to characterize 6 to 15 MeV small electron beams shaped by using commercial tubular applicators with 2, 3, 4 and 5 cm diameter and to assess the dosimetric performance under such irradiation conditions of the novel SCDD dosimeter by comparison with commercially available dosimeters, namely a Si-D and a plane-parallel ionization chamber. Percentage depth dose curves, beam profiles and output factors (OFs) were measured. A good agreement among the dosimeters was observed in all of the performed measurements. As for the tubular applicators, two main effects were evidenced: (i) OFs larger than unity were measured for a number of field sizes and energies, with values up to about 1.3, that is an output 30% greater than that obtained at the 10 x 10cm(2) reference field; (ii) for each diameter of the tubular applicator a noticeable increase of the OF values was observed with increasing beam energy, up to about 100% in the case of the smaller applicator. This OF behavior is remarkably different from what typically observed for small blocked fields having the same size and energy as those used in this study. OFs for tubular applicators depend considerably on the field size, so interpolation is unadvisable to predict the linear accelerator output for such applicators whereas reliable high-resolution detectors, as the silicon and diamond diodes used in this work allow OF measurements with uncertainties of about 1%.
引用
收藏
页码:8121 / 8133
页数:13
相关论文
共 29 条
[1]   Synthetic single crystal diamond diodes for radiotherapy dosimetry [J].
Almaviva, S. ;
Marinelli, Marco ;
Milani, E. ;
Tucciarone, A. ;
Verona-Rinati, G. ;
Consorti, R. ;
Petrucci, A. ;
De Notaristefani, F. ;
Ciancaglioni, I. .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2008, 594 (02) :273-277
[2]   Chemical vapor deposition diamond based multilayered radiation detector: Physical analysis of detection properties [J].
Almaviva, S. ;
Marinelli, Marco ;
Milani, E. ;
Prestopino, G. ;
Tucciarone, A. ;
Verona, C. ;
Verona-Rinati, G. ;
Angelone, M. ;
Pillon, M. ;
Dolbnya, I. ;
Sawhney, K. ;
Tartoni, N. .
JOURNAL OF APPLIED PHYSICS, 2010, 107 (01)
[3]  
[Anonymous], 1984, 35 ICRU
[4]  
Björk P, 2000, MED PHYS, V27, P2580, DOI 10.1118/1.1315317
[5]   Measurements of output factors with different detector types and Monte Carlo calculations of stopping-power ratios for degraded electron beams [J].
Björk, P ;
Knöös, T ;
Nilsson, P .
PHYSICS IN MEDICINE AND BIOLOGY, 2004, 49 (19) :4493-4506
[6]   Dosimetric characterization of a synthetic single crystal diamond detector in clinical radiation therapy small photon beams [J].
Ciancaglioni, I. ;
Marinelli, Marco ;
Milani, E. ;
Prestopino, G. ;
Verona, C. ;
Verona-Rinati, G. ;
Consorti, R. ;
Petrucci, A. ;
De Notaristefani, F. .
MEDICAL PHYSICS, 2012, 39 (07) :4493-4501
[7]   Characterization of a synthetic single crystal diamond Schottky diode for radiotherapy electron beam dosimetry [J].
Di Venanzio, C. ;
Marinelli, Marco ;
Milani, E. ;
Prestopino, G. ;
Verona, C. ;
Verona-Rinati, G. ;
Falco, M. D. ;
Bagala, P. ;
Santoni, R. ;
Pimpinella, M. .
MEDICAL PHYSICS, 2013, 40 (02)
[8]   Commissioning a p-type silicon diode for use in clinical electron beams [J].
Eveling, JN ;
Morgan, AM ;
Pitchford, WG .
MEDICAL PHYSICS, 1999, 26 (01) :100-107
[9]   Recommendations for clinical electron beam dosimetry: Supplement to the recommendations of Task Group 25 [J].
Gerbi, Bruce J. ;
Antolak, John A. ;
Deibel, F. Christopher ;
Followill, David S. ;
Herman, Michael G. ;
Higgins, Patrick D. ;
Huq, M. Saiful ;
Mihailidis, Dimitris N. ;
Yorke, Ellen D. ;
Hogstrom, Kenneth R. ;
Khan, Faiz M. .
MEDICAL PHYSICS, 2009, 36 (07) :3239-3279
[10]   Dosimetric characteristics of a new unshielded silicon diode and its application in clinical photon and electron beams [J].
Griessbach, I ;
Lapp, M ;
Bohsung, J ;
Gademann, G ;
Harder, D .
MEDICAL PHYSICS, 2005, 32 (12) :3750-3754