Proton pencil beam scanning treatment with feedback based voluntary moderate breath hold

被引:1
作者
Wang, Peng [1 ]
Tang, Shikui [2 ]
Leach, Karla [2 ]
Mangona, Victor [2 ]
Simone, Charles B., II [3 ]
Langen, Katja [4 ]
Chang, Chang [5 ]
机构
[1] Inova Hlth Syst, Dept Radiat Oncol, Falls Church, VA 22031 USA
[2] Texas Ctr Proton Therapy, Irving, TX USA
[3] New York Proton Ctr, New York, NY USA
[4] Emory Univ, Atlanta, GA 30322 USA
[5] Calif Protons Ca Therapy Ctr, San Diego, CA USA
关键词
Proton therapy; Moderate deep inspiration breath hold; Hodgkin's lymphoma; Cone beam CT; Deformable image registration; DEFORMABLE IMAGE REGISTRATION; DEEP INSPIRATION; LUNG DENSITY; COMPUTED-TOMOGRAPHY; RADIATION-THERAPY; MOTION MITIGATION; HODGKINS-DISEASE; CONTROL ABC; CT; IMMOBILIZATION;
D O I
10.1016/j.meddos.2019.11.003
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Introduction The aim of this article is to introduce a novel protocol for proton pencil beam scanning treatment with moderate deep inspiration breath hold (mDIBH) and report on our clinical implementation results. Methods Three computed tomography (CT) scannings to build the patient's anatomy model were performed during the patient's voluntary mDIBH. All 3 CT scans were used in the optimization during the treatment planning process. Both orthogonal kV imaging and cone-beam computed tomography (CBCT) were implemented for patient alignment with BH prior to the treatment. The BH CBCT images were analyzed for BH reproducibility and the virtual total dose (VTD) retrospectively. To find the VTD, a series of deformable image registrations (DIR) were performed between CBCT and pCT. The effect of the variation of lung density on the dose distribution was also analyzed in the study. Results The values of the mean, standard deviation, maximum, and minimum of the tumor location difference between the CBCT and pCT were 1.9, 1.6, 4.7, and 0.0 mm, respectively. The percentage difference in D99% of CTVs between VTD and the nominal plan was within 1.5%. Conclusions The feedback-based voluntary moderate BH proton PBS treatment was successfully performed in our clinic. This study shows that there is a potential to implement the BH treatment widely in proton centers. (C) 2019 American Association of Medical Dosimetrists. Published by Elsevier Inc. All rights reserved.
引用
收藏
页码:E10 / E15
页数:6
相关论文
共 48 条
[1]   Dosimetric evaluation of lung tumor immobilization using breath hold at deep inspiration [J].
Barnes, EIA ;
Murray, BR ;
Robinson, DM ;
Underwood, LJ ;
Hanson, J ;
Roa, WHY .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2001, 50 (04) :1091-1098
[2]   The UK HeartSpare Study: Randomised evaluation of voluntary deep-inspiratory breath-hold in women undergoing breast radiotherapy [J].
Bartlett, Frederick R. ;
Colgan, Ruth M. ;
Carr, Karen ;
Donovan, Ellen M. ;
McNair, Helen A. ;
Locke, Imogen ;
Evans, Philip M. ;
Haviland, Joanne S. ;
Yarnold, John R. ;
Kirby, Anna M. .
RADIOTHERAPY AND ONCOLOGY, 2013, 108 (02) :242-247
[3]   4D treatment planning for scanned ion beams [J].
Bert, Christoph ;
Rietzel, Eike .
RADIATION ONCOLOGY, 2007, 2 (1)
[4]   RESULTS OF A MULTI-INSTITUTION DEFORMABLE REGISTRATION ACCURACY STUDY (MIDRAS) [J].
Brock, Kristy K. .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2010, 76 (02) :583-596
[5]   Consensus Guidelines for Implementing Pencil-Beam Scanning Proton Therapy for Thoracic Malignancies on Behalf of the PTCOG Thoracic and Lymphoma Subcommittee [J].
Chang, Joe Y. ;
Zhang, Xiaodong ;
Knopf, Antje ;
Li, Heng ;
Mori, Shinichiro ;
Dong, Lei ;
Lu, Hsiao-Ming ;
Liu, Wei ;
Badiyan, Shahed N. ;
Both, Stephen ;
Meijers, Arturs ;
Lin, Liyong ;
Flampouri, Stella ;
Li, Zuofeng ;
Umegaki, Kikuo ;
Simone, Charles B., II ;
Zhu, Xiaorong R. .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2017, 99 (01) :41-50
[6]   Robustness of the Voluntary Breath-Hold Approach for the Treatment of Peripheral Lung Tumors Using Hypofractionated Pencil Beam Scanning Proton Therapy [J].
Dueck, Jenny ;
Knopf, Antje-Christin ;
Lomax, Antony ;
Albertini, Francesca ;
Persson, Gitte F. ;
Josipovic, Mirjana ;
Aznar, Marianne ;
Weber, Damien C. ;
af Rosenschold, Per Munck .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2016, 95 (01) :534-541
[7]   Minimax optimization for handling range and setup uncertainties in proton therapy [J].
Fredriksson, Albin ;
Forsgren, Anders ;
Hardemark, Bjorn .
MEDICAL PHYSICS, 2011, 38 (03) :1672-1684
[8]   Validation of a deformable image registration produced by a commercial treatment planning system in head and neck [J].
Garcia-Molla, Rafael ;
de Marco-Blancas, Noelia ;
Bonaque, Jorge ;
Vidueira, Laura ;
Lopez-Tarjuelo, Juan ;
Perez-Calatayud, Jose .
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2015, 31 (03) :219-223
[9]   A Voluntary Breath-Hold Treatment Technique for the Left Breast With Unfavorable Cardiac Anatomy Using Surface Imaging [J].
Gierga, David P. ;
Turcotte, Julie C. ;
Sharp, Gregory C. ;
Sedlacek, Daniel E. ;
Cotter, Christopher R. ;
Taghian, Alphonse G. .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2012, 84 (05) :E663-E668
[10]   Continuous Positive Airway Pressure for Motion Management in Stereotactic Body Radiation Therapy to the Lung: A Controlled Pilot Study [J].
Goldstein, Jeffrey D. ;
Lawrence, Yaacov R. ;
Appel, Sarit ;
Landau, Efrat ;
Ben-David, Merav A. ;
Rabin, Tatiana ;
Benayun, Maoz ;
Dubinski, Sergey ;
Weizman, Noam ;
Alezra, Dror ;
Gnessin, Hila ;
Goldstein, Adam M. ;
Baidun, Khader ;
Segel, Michael J. ;
Peled, Nir ;
Symon, Zvi .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2015, 93 (02) :391-399