Mechanical stress: An independent determinant of early bioprosthetic calcification in humans

被引:18
作者
Liao, Kenneth K.
Li, Xiaohuan
John, Ranjit
Amatya, Devesh M.
Joyce, Lyle D.
Park, Soon J.
Bianco, Richard
Bolman, R. Morton, III
机构
[1] Univ Minnesota, Div Cardiothorac Surg, Minneapolis, MN 55455 USA
[2] Mayo Clin, Div Cardiac Surg, Rochester, MN USA
[3] Brigham & Womens Hosp, Div Cardiac Surg, Boston, MA 02115 USA
关键词
D O I
10.1016/j.athoracsur.2008.03.061
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background. Mechanical stress is one of the contributing factors for bioprosthetic calcification. A HeartMate XVE (Thoratec, CA) left ventricular assist device (LVAD) has two identical porcine valves, one for inflow and the other for outflow. The inflow valve endures a higher closing pressure than the outflow valve; thus, an implanted LVAD offers an ideal human model to study the independent effect of stress on calcification. Methods. Sixty-four pairs of LVAD inflow and outflow valves underwent gross examination, histologic study, and x-ray imaging. X-ray films were converted to digital images, and the calcification area was calculated. The distribution of calcium deposits was documented. The frequency and degree of calcification in both valves were analyzed (paired t test). Calcification of both valves in relationship to the duration of LVAD implantation and to the patient's age was also analyzed (linear regression). Results. The mean age of patients supported with LVAD was 55 +/- 12 years (range, 17 to 77 years). The mean duration of LVAD implantation was 265 +/- 151 days (range, 3 to 630 days). Calcification developed more commonly in inflow valves. The calcification area (CA) was larger in the inflow valves (21.6 +/- 30.7 mm(2)) than in the outflow valves (15.1 +/- 26.2 mm(2), p < 0.05). There was a positive relationship between CA and days of implantation for both valves (inflow CA = 4.96 +/- 0.063 days; outflow, CA = 2.39 +/- 0.047 days, linear regression; p < 0.05 for both). Conclusions. Mechanical stress is an independent determinant of early bioprosthetic calcification in humans.
引用
收藏
页码:491 / 495
页数:5
相关论文
共 14 条
  • [1] BERNACCA GM, 1992, J BIOMED MATER RES, V26, P959
  • [2] A 3-DIMENSIONAL ANALYSIS OF A BIOPROSTHETIC HEART-VALVE
    BLACK, MM
    HOWARD, IC
    HUANG, X
    PATTERSON, EA
    [J]. JOURNAL OF BIOMECHANICS, 1991, 24 (09) : 793 - &
  • [3] Mechanical heart valve prostheses: identification and evaluation (vol 12, pg 322, 2003)
    Butany, J
    Ahluwalia, MS
    Munroe, C
    Fayet, C
    Ahn, C
    Blit, P
    Kepron, C
    Cusimano, RJ
    Leask, RL
    [J]. CARDIOVASCULAR PATHOLOGY, 2003, 12 (06) : 322 - 344
  • [4] Biological replacement heart valves - Identification and evaluation
    Butany, J
    Fayet, C
    Ahluwalia, MS
    Blit, P
    Ahn, C
    Munroe, C
    Israel, N
    Cusimano, RJ
    Leask, RL
    [J]. CARDIOVASCULAR PATHOLOGY, 2003, 12 (03) : 119 - 139
  • [5] STRESS-DISTRIBUTION ON THE CUSPS OF A POLYURETHANE TRILEAFLET HEART-VALVE PROSTHESIS IN THE CLOSED POSITION
    CHANDRAN, KB
    KIM, SH
    HAN, G
    [J]. JOURNAL OF BIOMECHANICS, 1991, 24 (06) : 385 - 395
  • [6] In vitro testing of bioprostheses: Influence of mechanical stresses and lipids on calcification
    Deiwick, M
    Glasmacher, B
    Baba, HA
    Roeder, N
    Reul, H
    von Bally, G
    Scheld, HH
    [J]. ANNALS OF THORACIC SURGERY, 1998, 66 (06) : S206 - S211
  • [7] Deiwick M, 1998, J HEART VALVE DIS, V7, P419
  • [8] Glasmacher B, 1998, J HEART VALVE DIS, V7, P415
  • [9] A new in vitro test method for calcification of bioprosthetic heart valves
    Glasmacher, B
    Deiwick, M
    Reul, H
    Knesch, H
    Keus, D
    Rau, G
    [J]. INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 1997, 20 (05) : 267 - 271
  • [10] NUMERICAL-SIMULATION OF LEAFLET FLEXURE IN BIOPROSTHETIC VALVES MOUNTED ON RIGID AND EXPANSILE STENTS
    KRUCINSKI, S
    VESELY, I
    DOKAINISH, MA
    CAMPBELL, G
    [J]. JOURNAL OF BIOMECHANICS, 1993, 26 (08) : 929 - 943