Development of a Computerized 4-D MRI Phantom for Liver Motion Study

被引:6
|
作者
Wang, Chunhao [1 ]
Yin, Fang-Fang [1 ,2 ]
Segars, W. P. [3 ]
Chang, Zheng [1 ]
Ren, Lei [1 ,2 ]
机构
[1] Duke Univ, Med Ctr, Dept Radiat Oncol, Durham, NC 27710 USA
[2] Duke Kunshan Univ, Med Phys Grad Program, Kunshan, Jiangsu, Peoples R China
[3] Duke Univ, Med Ctr, Dept Radiol, Durham, NC 27710 USA
基金
美国国家卫生研究院;
关键词
MRI; computerized phantom; liver; organ motion; deformable registration; PHARMACOKINETIC ANALYSIS; ORGAN MOTION; XCAT; SIMULATION; THERAPY; HEAD;
D O I
10.1177/1533034617723753
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: To develop a 4-dimensional computerized magnetic resonance imaging phantom with image textures extracted from real patient scans for liver motion studies. Methods: The proposed phantom was developed based on the current version of 4-dimensional extended cardiac-torso computerized phantom and a clinical magnetic resonance scan. Initially, the extended cardiac-torso phantom was voxelized in abdominal-chest region at the end of exhalation phase. Structures/tissues were classified into 4 categories: (1) Seven key textured organs, including liver, gallbladder, spleen, stomach, heart, kidneys, and pancreas, were mapped from a clinical T1-weighted liver magnetic resonance scan using deformable registration. (2) Large textured soft tissue volumes were simulated via an iterative pattern generation method using the same magnetic resonance scan. (3) Lung and intestine structures were generated by assigning uniform intensity with proper noise modeling. (4) Bony structures were generated by assigning the magnetic resonance values. A spherical hypointensity tumor was inserted into the liver. Other respiratory phases of the 4-dimensional phantom were generated using the backward deformation vector fields exported by the extended cardiac-torso program, except that bony structures were generated separately for each phase. A weighted image filtering process was utilized to improve the overall tissue smoothness at each phase. Results: Three 4-dimensional series with different motion amplitudes were generated. The developed motion phantom produced good illustrations of abdominal-chest region with anatomical structures in key organs and texture patterns in large soft tissue volumes. In a standard series, the tumor volume was measured as 13.90 0.11 cm(3) in a respiratory cycle and the tumor's maximum center-of-mass shift was 2.95 cm/1.84 cm on superior-inferior/anterior-posterior directions. The organ motion during the respiratory cycle was well rendered. The developed motion phantom has the flexibility of motion pattern variation, organ geometry variation, and tumor modeling variation. Conclusions: A 4-D computerized phantom was developed and could be used to produce image series with synthetic magnetic resonance textures for magnetic resonance imaging research of liver motion.
引用
收藏
页码:1051 / 1059
页数:9
相关论文
共 50 条
  • [1] Development of a 4-D digital mouse phantom for molecular imaging research
    Segars, WP
    Tsui, BMW
    Frey, EC
    Johnson, GA
    Berr, SS
    MOLECULAR IMAGING AND BIOLOGY, 2004, 6 (03) : 149 - 159
  • [2] Motion modeling from 4D MR images of liver simulating phantom
    Kavaluus, Henna
    Koivula, Lauri
    Salli, Eero
    Seppala, Tiina
    Saarilahti, Kauko
    Tenhunen, Mikko
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2022, 23 (07):
  • [3] Phantom-Based Evaluation of Isotropic Reconstruction of 4-D MRI Volumes using Super-Resolution
    Van Reeth, Eric
    Tham, Ivan W. K.
    Tan, Cher Heng
    Poh, Chueh Loo
    PROCEEDINGS OF THE 2013 FOURTH INTERNATIONAL WORKSHOP ON COMPUTATIONAL INTELLIGENCE IN MEDICAL IMAGING (CIMI), 2013, : 6 - 13
  • [4] Adaptive 4-D scan for real-time intraoperative MRI
    Tokuda, J.
    Morikawa, S.
    Haque, H. A.
    Tsukamoto, T.
    Matsumiya, K.
    Liao, H.
    Masamune, K.
    Dohi, T.
    INTERNATIONAL JOURNAL OF COMPUTER ASSISTED RADIOLOGY AND SURGERY, 2006, 1 : 290 - 292
  • [5] Comparative study of sub-second temporal resolution 4D-MRI and 4D-CT for target motion assessment in a phantom model
    Wang, Tianyuan
    Sofue, Keitaro
    Shimada, Ryuji
    Ishihara, Takeaki
    Yada, Ryuichi
    Miyamoto, Masanori
    Sasaki, Ryohei
    Murakami, Takamichi
    SCIENTIFIC REPORTS, 2023, 13 (01)
  • [6] MCAT to XCAT: The Evolution of 4-D Computerized Phantoms for Imaging Research
    Segars, W. Paul
    Tsui, Benjamin M. W.
    PROCEEDINGS OF THE IEEE, 2009, 97 (12) : 1954 - 1968
  • [7] Speckle tracking in a phantom and feature-based tracking in liver in the presence of respiratory motion using 4D ultrasound
    Harris, Emma J.
    Miller, Naomi R.
    Bamber, Jeffrey C.
    Symonds-Tayler, J. Richard N.
    Evans, Philip M.
    PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (12): : 3363 - 3380
  • [8] Gastrointestinal 4D MRI with respiratory motion correction
    Johansson, Adam
    Balter, James M.
    Cao, Yue
    MEDICAL PHYSICS, 2021, 48 (05) : 2521 - 2527
  • [9] Motion tracking in the liver: Validation of a method based on 4D ultrasound using a nonrigid registration technique
    Vijayan, Sinara
    Klein, Stefan
    Hofstad, Erlend Fagertun
    Lindseth, Frank
    Ystgaard, Brynjulf
    Lango, Thomas
    MEDICAL PHYSICS, 2014, 41 (08) : 562 - 569
  • [10] Extension of the NCAT phantom for the investigation of intra-fraction respiratory motion in IMRT using 4D Monte Carlo
    McGurk, Ross
    Seco, Joao
    Riboldi, Marco
    Wolfgang, John
    Segars, Paul
    Paganetti, Harald
    PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (05): : 1475 - 1490