Electron dosimetry based on the absorbed dose to water concept:: A comparison of the AAPM TG-51 and DIN 6800-2 protocols

被引:10
|
作者
Dohm, OS
Christ, G
Nüsslin, F
Schüle, E
Bruggmoser, G
机构
[1] Univ Tubingen, Klin Radioonkol, Abt Med Phys, D-72076 Tubingen, Germany
[2] PTW Freiburg, D-79115 Freiburg, Germany
[3] Univ Freiburg Klinikum, Abt Strahlentherapie, D-79106 Freiburg, Germany
关键词
electron dosimetry; absorbed dose to water; Roos chamber; perturbation correction;
D O I
10.1118/1.1414008
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The dosimetry protocols DIN 6800-2 and AAPM TG-51, both based on the absorbed dose to water concept, are compared in their theoretical background and in their application to electron dosimetry. The agreement and disagreement in correction factors and energy parameters used in both protocols will be shown and discussed. Measurements with three different types of ionization chambers were performed and evaluated according to both protocols. As a result the perturbation correction factor P-wall(60Co) for the Roos chamber was determined to 1.024 +/- 0.5%. (C) 2001 American Association of Physicists in Medicine.
引用
收藏
页码:2258 / 2264
页数:7
相关论文
共 5 条
  • [1] Determination of absorbed dose to water for high-energy photon and electron beams-comparison of the standards DIN 6800-2 (1997), IAEA TRS 398 (2000) and DIN 6800-2 (2006)
    Abu Zakaria, Golam
    Schuette, Wilhelm
    JOURNAL OF MEDICAL PHYSICS, 2007, 32 (01) : 3 - 11
  • [2] Comparison of IPSM 1990 photon dosimetry code of practice with IAEA TRS-398 and AAPM TG-51
    Castrillon, Silvia Vargas
    Henriquez, Francisco Cutanda
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2009, 10 (01): : 136 - 146
  • [3] AAPM WGTG51 Report 385: Addendum to the AAPM's TG-51 protocol for clinical reference dosimetry of high-energy electron beams
    Muir, Bryan
    Davis, Stephen
    Dhanesar, Sandeep
    Hillman, Yair
    Iakovenko, Viktor
    Kim, Grace Gwe-Ya
    Alves, Victor Gabriel Leandro
    Lei, Yu
    Lowenstein, Jessica
    Renaud, James
    Sarfehnia, Arman
    Siebers, Jeffrey
    Tantot, Laurent
    MEDICAL PHYSICS, 2024, 51 (09) : 5840 - 5857
  • [4] A Comparison of dosimetric parameters between IAEA TRS-398, AAPM TG-51 protocols and Monte-Carlo simulation
    Mahdavi, S. R. M.
    Mahdavi, M.
    Alijanzadeh, H.
    Zabihzadeh, M.
    Mostaar, A.
    IRANIAN JOURNAL OF RADIATION RESEARCH, 2012, 10 (01): : 43 - 51
  • [5] Comparison of AAPM Addendum to TG-51, IAEA TRS-398, and JS']JSMP 12: Calibration of photon beams in water
    Kinoshita, Naoki
    Oguchi, Hiroshi
    Nishimoto, Yasuhiro
    Adachi, Toshiki
    Shioura, Hiroki
    Kimura, Hirohiko
    Doi, Kunio
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2017, 18 (05): : 271 - 278