Method for measuring the refractive index of liquids using a cylindrical cell

被引:1
|
作者
Duran-Ramirez, Victor M. [1 ]
Martinez-Rios, Alejandro [2 ]
Munoz-Maciel, Jesus [1 ]
Pena-Lecona, Francisco G. [1 ]
Casillas-Rodriguez, Francisco J. [1 ]
Selvas-Aguilar, Romeo [3 ]
Mora-Gonzalez, Miguel [1 ]
机构
[1] Univ Guadalajara, Ctr Univ Lagos, Guadalajara 47460, Jalisco, Mexico
[2] Ctr Invest Opt, Guanajuato 37150, Mexico
[3] Univ Autonoma Nuevo Leon, Nuevo Leon 66450, Mexico
关键词
refractive index of liquid; cylindrical cell; principal ray; cylindrical lens; WATER; RAINBOW;
D O I
10.1117/1.OE.52.7.074101
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
A new and simple method for measuring the refractive index of liquid substances is presented. In this method, a laser beam impinges transversely on a glass tube (cylindrical cell) filled with the liquid to be measured. The laser beam incident on the cylindrical cell is deviated when it propagates through the wall of the cell and the liquid contained in it. By measuring the deviation of the principal ray of the laser beam when it emerges from the cylindrical cell, we can determine the refractive index of the liquid. To show the feasibility of the method, we measured the refractive index of pure water with a He-Ne laser. (C) 2013 Society of Photo-Optical Instrumentation Engineers (SPIE)
引用
收藏
页数:6
相关论文
共 50 条
  • [41] Application of the critical angle method to refractive index measurement of human skin in vivo under partial contact
    Yoshida, Kenichiro
    Ohkubo, Kohji
    Ojima, Nobutoshi
    Iwata, Kayoko
    JOURNAL OF BIOMEDICAL OPTICS, 2013, 18 (03)
  • [42] Pressure, temperature and refractive index determination of fluids using a single fibre optic point sensor
    Bremer, K.
    Reinsch, T.
    Leen, G.
    Roth, B.
    Lochmann, S.
    Lewis, E.
    SENSORS AND ACTUATORS A-PHYSICAL, 2017, 256 : 84 - 88
  • [43] Comparison of Transparency and Shrinkage During Clearing of Insect Brains Using Media With Tunable Refractive Index
    Bekkouche, Bo M. B.
    Fritz, Helena K. M.
    Rigosi, Elisa
    O'Carroll, David C.
    FRONTIERS IN NEUROANATOMY, 2020, 14
  • [44] DENSITY, REFRACTIVE INDEX, APPARENT VOLUMES AND EXCESS MOLAR VOLUMES OF FOUR PROTIC IONIC LIQUIDS plus WATER AT T=298.15 AND 323.15 K
    Rocha Pinto, R.
    Santos, D.
    Mattedi, S.
    Aznar, M.
    BRAZILIAN JOURNAL OF CHEMICAL ENGINEERING, 2015, 32 (03) : 671 - 682
  • [45] Trapping and controlled rotation of low-refractive-index particles using dual line optical tweezers
    S.K. Mohanty
    R.S. Verma
    P.K. Gupta
    Applied Physics B, 2007, 87 : 211 - 215
  • [46] Real-Time Measurement of Single Bacterium's Refractive Index using Optofluidic Immersion Refractometry
    Liu, P. Y.
    Chin, L. K.
    Ser, W.
    Ayi, T. C.
    Yap, P. H.
    Bourouina, T.
    Leprince-Wang, Y.
    28TH EUROPEAN CONFERENCE ON SOLID-STATE TRANSDUCERS (EUROSENSORS 2014), 2014, 87 : 356 - 359
  • [47] Measuring and predicting canopy nitrogen nutrition in wheat using a spectral index-The canopy chlorophyll content index (CCCI)
    Fitzgerald, Glenn
    Rodriguez, Daniel
    O'Leary, Garry
    FIELD CROPS RESEARCH, 2010, 116 (03) : 318 - 324
  • [48] Cylindrical cell membranes in uniform applied electric fields: Validation of a transport lattice method
    Stewart, DA
    Gowrishankar, TR
    Smith, KC
    Weaver, JC
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2005, 52 (10) : 1643 - 1653
  • [49] Utilization of cylindrical Cu/Fe galvanic cell as an effective method for the removal of humic acids
    Libecki, Bartosz
    Mikolajczyk, Tomasz
    Pierozynski, Boguslaw
    Kuczynski, Mateusz
    CHEMICAL AND PROCESS ENGINEERING-NEW FRONTIERS, 2023, 44 (03):
  • [50] Method for vector characterization of polar liquids using frequency-domain spectroscopy
    Saha, Shimul C.
    Grant, James P.
    Ma, Yong
    Khalid, A.
    Hong, Feng
    Cumming, David R. S.