Crimping and deployment of balloon-expandable valved stents are responsible for the increase in the hydraulic conductance of leaflets

被引:7
作者
Convelbo, Channing [1 ]
Guetat, Pierre [1 ]
Cambillau, Michele [2 ]
Allam, Bachir [3 ]
Bruneval, Patrick [4 ]
Lafont, Antoine [1 ,3 ,5 ]
Zegdi, Rachid [1 ,5 ,6 ]
机构
[1] INSERM, U970, Paris, France
[2] Hop Europeen Georges Pompidou, AP HP, Dept Biochem, F-75908 Paris, France
[3] Hop Europeen Georges Pompidou, AP HP, Dept Cardiol, F-75908 Paris, France
[4] Hop Europeen Georges Pompidou, AP HP, Dept Pathol, F-75908 Paris, France
[5] Univ Paris 05, Paris, France
[6] Hop Europeen Georges Pompidou, AP HP, Dept Cardiovasc Surg, F-75908 Paris, France
关键词
Pericardium; Leaflets; Percutaneous valves; Injury; Hydraulic conductance; PERICARDIUM; INJURY;
D O I
10.1093/ejcts/ezt175
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Leaflet injury has been documented to occur during the deployment of valved stents (VSs). The pathological aspects, however, of this injury are difficult to quantify. Conversely, the hydraulic conductance of a (pericardial) membrane may be easily determined. The impact of crimping and deployment of VS on this parameter was therefore investigated. Bovine pericardial square (25 x 25 mm) patches were placed within a pressure chamber and their hydraulic conductance was determined. The influence of the pressure gradient and tissue thickness on this parameter was analysed. Six balloon-expandable VS were constructed. The hydraulic conductance of their bovine pericardial leaflets was determined before and after VS crimping and deployment in four of them. Pericardial leaflets of non-crimped VS were used as controls. Hydraulic conductance increased insignificantly with the pressure level within the chamber: from 128 +/- 26.9 ml/h/m(2)/mmHg at a pressure of 50 mmHg to 232.3 +/- 51.9 ml/h/m(2)/mmHg at a pressure of 250 mmHg (P = 0.117). Hydraulic conductance was not correlated to pericardial thickness, for thickness measurements ranging from 0.34 to 0.76 mm. The hydraulic conductance of VS leaflets significantly increased immediately after crimping from 45.2 +/- 7.6 to 667.9.0 +/- 527.2 ml/h/m(2)/mmHg (P < 0.001). This increase was still observed 24 h after VS deployment. No change in hydraulic conductance occurred in the control group. The determination of the hydraulic conductance of pericardial patches was easy to perform, reproducible and not influenced by tissue thickness. The hydraulic conductance of pericardial leaflets dramatically increased after VS crimping and deployment. This parameter might be, in the future, a useful noninvasive tool in studying leaflet trauma.
引用
收藏
页码:1045 / 1050
页数:6
相关论文
共 10 条
[1]   Traumatic leaflet injury during the use of percutaneous valves: a comparative study of balloon- and self-expandable valved stents [J].
Amahzoune, Brahim ;
Bruneval, Patrick ;
Allam, Bachir ;
Lafont, Antoine ;
Fabiani, Jean-Noel ;
Zegdi, Rachid .
EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 2013, 43 (03) :488-493
[2]   Impairment of pericardial leaflet structure from balloon-expanded valved stents [J].
de Buhr, Wiebke ;
Pfeifer, Stefan ;
Slotta-Huspenina, Julia ;
Wintermantel, Erich ;
Lutter, Georg ;
Goetz, Wolfgang A. .
JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2012, 143 (06) :1417-1421
[3]   HYDRAULIC PERMEABILITY OF CANINE AND HUMAN PERICARDIUM INVITRO [J].
FINGEROTE, R ;
FUNG, HYM ;
RABKIN, SW .
BASIC RESEARCH IN CARDIOLOGY, 1980, 75 (06) :764-771
[4]  
HILBERT SL, 1992, AM J PATHOL, V140, P1195
[5]   Crimping May Affect the Durability of Transcatheter Valves: An Experimental Analysis [J].
Kiefer, Philipp ;
Gruenwald, Felix ;
Kempfert, Joerg ;
Aupperle, Heike ;
Seeburger, Joerg ;
Mohr, Friedrich Wilhelm ;
Walther, Thomas .
ANNALS OF THORACIC SURGERY, 2011, 92 (01) :155-160
[6]   Porcine vena cava as an alternative to bovine pericardium in bioprosthetic percutaneous heart valves [J].
Munnelly, Amy E. ;
Cochrane, Leonard ;
Leong, Joshua ;
Vyavahare, Naren R. .
BIOMATERIALS, 2012, 33 (01) :1-8
[7]   CORRELATION OF STRUCTURE AND VISCOELASTIC PROPERTIES IN THE PERICARDIA OF 4 MAMMALIAN-SPECIES [J].
NAIMARK, WA ;
LEE, JM ;
LIMEBACK, H ;
CHEUNG, DT .
AMERICAN JOURNAL OF PHYSIOLOGY, 1992, 263 (04) :H1095-H1106
[8]   Calcification of tissue heart valve substitutes: Progress toward understanding and prevention [J].
Schoen, FJ ;
Levy, RJ .
ANNALS OF THORACIC SURGERY, 2005, 79 (03) :1072-1080
[9]   Transcatheter valve implantation for patients with aortic stenosis:: a position statement from the European Association of Cardio-Thoracic Surgery (EACTS) and the European Society of Cardiology (ESC), in collaboration with the European Association of Percutaneous Cardiovascular Interventions (EAPCI) [J].
Vahanian, Alec ;
Alfieri, Ottavio ;
Al-Attar, Nawwar ;
Antunes, Manuel ;
Bax, Jeroen ;
Cormier, Bertrand ;
Cribier, Alain ;
De Jaegere, Peter ;
Fournial, Gerard ;
Kappetein, Arie Pieter ;
Kovac, Jan ;
Ludgate, Susanne ;
Maisano, Francesco ;
Moat, Neil ;
Mohr, Friedrich ;
Nataf, Patrick ;
Pierard, Luc ;
Pomar, Jose Luis ;
Schofer, Joachim ;
Tornos, Pilar ;
Tuzcu, Murat ;
van Hout, Ben ;
Von Segesser, Ludwig K. ;
Walther, Thomas .
EUROPEAN HEART JOURNAL, 2008, 29 (11) :1463-1470
[10]   Evidence of leaflet injury during percutaneous aortic valve deployment [J].
Zegdi, Rachid ;
Bruneval, Patrick ;
Blanchard, Didier ;
Fabiani, Jean-Noel .
EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 2011, 40 (01) :257-259