Estimation of pulsatile flow and differential pressure based on multi-layer perceptron using an axial flow blood pump

被引:0
|
作者
Dang Caixin [1 ]
Wang Shuai [1 ]
Yu Zheqin [1 ]
Wu Weiqiang [1 ]
Wu Kun [1 ]
Tan Jianping [1 ]
机构
[1] Cent South Univ, Sch Mech & Elect Engn, Changsha 410083, Peoples R China
来源
JOURNAL OF ENGINEERING-JOE | 2020年 / 2020卷 / 14期
关键词
blood vessels; blood pressure measurement; pulsatile flow; haemodynamics; multilayer perceptrons; blood; pumps; differential pressure; multilayer perceptron; axial flow blood pump; noninvasive method; pulsating flow; estimation model; pulsating conditions; pulsation waveform; flow error; pressure difference error; traditional flow; pressure prediction method; MODULATION; HEART;
D O I
10.1049/joe.2020.0045
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
This study proposes a non-invasive method for estimating the pulsating flow and pressure difference, which uses the blood pump estimation model based on a multi-layer perceptron to calculate the flow and pressure difference under pulsating conditions. The model takes 11 parameters such as the rotational speed, power, and pulsation waveform of the blood pump as the input and uses the pressure difference and flow as the output. The experimental results of 119,590 sample data show that the flow error of the training set of the blood pump estimation model is 0.14 l/min and the pressure difference error is 7.50 mmHg; the flow error of the test set is 0.14 l/min and the pressure difference error is 7.50 mmHg. Compared with the traditional flow and pressure prediction method, this method has higher precision, which will provide a certain technical accumulation for accurately estimating the flow and pressure difference of the blood pump in the pulsating conditions.
引用
收藏
页码:928 / 931
页数:4
相关论文
共 41 条
  • [21] Condition-based monitoring system for rolling element bearing using a generic multi-layer perceptron
    de Almeida, Luis F.
    Bizarria, Jose W. P.
    Bizarria, Francisco C. P.
    Mathias, Mauro H.
    JOURNAL OF VIBRATION AND CONTROL, 2015, 21 (16) : 3456 - 3464
  • [22] Cloud storage based diagnosis of breast cancer using novel transfer learning with multi-layer perceptron
    Gul Shaira Banu Jahangeer
    T. Dhiliphan Rajkumar
    International Journal of System Assurance Engineering and Management, 2024, 15 : 60 - 72
  • [23] Accelerating Multi-Layer Perceptron based Short Term Demand Forecasting Using Graphics Processing Units
    He, Ting
    Dong, ZhaoYang
    Meng, Ke
    Wang, Hua
    Oh, Y. T.
    T& D ASIA: 2009 TRANSMISSION & DISTRIBUTION CONFERENCE & EXPOSITION: ASIA AND PACIFIC, 2009, : 604 - +
  • [24] Implementation of hybrid particle swarm optimization-differential evolution algorithms coupled with multi-layer perceptron for suspended sediment load estimation
    Mohammadi, Babak
    Guan, Yiqing
    Moazenzadeh, Roozbeh
    Safari, Mir Jafar Sadegh
    CATENA, 2021, 198
  • [25] fMRI-Based Alzheimer's Disease Detection Using the SAS Method with Multi-Layer Perceptron Network
    Chelladurai, Aarthi
    Narayan, Dayanand Lal
    Divakarachari, Parameshachari Bidare
    Loganathan, Umasankar
    BRAIN SCIENCES, 2023, 13 (06)
  • [26] Cross-view and multi-view gait recognitions based on view transformation model using multi-layer perceptron
    Kusakunniran, Worapan
    Wu, Qiang
    Zhang, Jian
    Li, Hongdong
    PATTERN RECOGNITION LETTERS, 2012, 33 (07) : 882 - 889
  • [27] Measure-based modeling and FPGA Implementation of RF Power Amplifier using a multi-layer perceptron neural network
    Nunez-Perez, J. C.
    Cardenas-Valdez, J. R.
    Galaviz Aguilar, J. A.
    Gontrand, C.
    Goral, B.
    Verdier, J.
    2014 INTERNATIONAL CONFERENCE ON ELECTRONICS, COMMUNICATIONS AND COMPUTERS (CONIELECOMP), 2014, : 237 - 242
  • [28] Validation of an Axial Flow Blood Pump: Computational Fluid Dynamics Results Using Particle Image Velocimetry
    Su, Boyang
    Chua, Leok Poh
    Wang, Xikun
    ARTIFICIAL ORGANS, 2012, 36 (04) : 359 - 367
  • [29] Development and initial performance of a miniature axial flow blood pump using magnetic fluid shaft seal
    Eiji Okamoto
    Tetsuya Yano
    Kazumitsu Sekine
    Yusuke Inoue
    Yasuyuki Shiraishi
    Tomoyuki Yambe
    Yoshinori Mitamura
    Journal of Artificial Organs, 2023, 26 : 12 - 16
  • [30] Development and initial performance of a miniature axial flow blood pump using magnetic fluid shaft seal
    Okamoto, Eiji
    Yano, Tetsuya
    Sekine, Kazumitsu
    Inoue, Yusuke
    Shiraishi, Yasuyuki
    Yambe, Tomoyuki
    Mitamura, Yoshinori
    JOURNAL OF ARTIFICIAL ORGANS, 2023, 26 (01) : 12 - 16