Characterisation of the IPNE Bonner sphere spectrometer by comparison with the PTB system

被引:6
作者
Alevra, AV
Plostinaru, VD
机构
[1] Phys Tech Bundesanstalt, Sect 6 42, D-38116 Braunschweig, Germany
[2] Inst Phys & Nucl Engn, Bucharest, Romania
关键词
multisphere neutron spectrometer; response matrix; unfolding; neutron spectrometry;
D O I
10.1016/S0168-9002(01)01381-X
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
An existing set of Bonner spheres of the Institute for Physics and Nuclear Engineering in Bucharest has been modified following the PTB design for application of an He-3-filled SP9 counter. By simple interpolations of the fluence responses of the PTB spheres, a preliminary fluence response matrix could be established for the IPNE Bonner sphere set. For further adjustments of the preliminary responses and for validation of the final fluence response matrix of the Romanian Bonner sphere spectrometer, calibration measurements with a reference Cf-252 neutron source and joint measurements, including the PTB Bonner sphere spectrometer, were performed in a few neutron fields differing substantially in their spectral distributions. It is estimated that the integral neutron fluences and dose-equivalent values can now be determined with the Romanian spectrometer with uncertainties of about +/-4% and +/-8%, respectively. (C) 2002 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:21 / 25
页数:5
相关论文
共 5 条
[1]  
Alevra A. V., 2000, PTB642001
[2]  
ALEVRA AV, 1992, RADIAT PROT DOSIM, V40, P91
[3]   EXPERIMENTAL-DETERMINATION OF THE RESPONSE OF 4 BONNER SPHERE SETS TO THERMAL-NEUTRONS [J].
THOMAS, DJ ;
ALEVRA, AV ;
HUNT, JB ;
SCHRAUBE, H .
RADIATION PROTECTION DOSIMETRY, 1994, 54 (01) :25-31
[4]  
THOMAS DJ, 1992, EXT31 RSA NAT PHYS L
[5]  
Wiegel B., 1994, N21 PTB