Evaluation of Four Commercial Kits for SARS-CoV-2 Real-Time Reverse-Transcription Polymerase Chain Reaction Approved by Emergency-Use-Authorization in Korea

被引:47
作者
Hur, Kyu-Hwa [1 ]
Park, Kuenyoul [1 ]
Lim, Youngkuen [1 ]
Jeong, Yun Sil [1 ]
Sung, Heungsup [1 ]
Kim, Mi-Na [1 ]
机构
[1] Univ Ulsan, Coll Med, Asan Med Ctr, Dept Lab Med, Seoul, South Korea
关键词
severe acute respiratory syndrome coronavirus 2; coronavirus infectious disease 2019; emergency-use-authorization; real time; reverse transcription; polymerase chain reaction; EAST RESPIRATORY SYNDROME; VALIDATION; GUIDELINES; DIAGNOSIS; COVID-19;
D O I
10.3389/fmed.2020.00521
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
SARS-CoV-2 real-time reverse-transcription PCR (rRT-PCR) is the most effective testing system currently available to counter COVID-19 epidemics when potent treatments and vaccines are unavailable. Therefore, four SARS-CoV-2 rRT-PCR kits have been approved by the emergency-use-authorization (EUA) without clinical validation in Korea until March 15, 2020. This study evaluated the analytical and clinical performance of these kits. Allplex 2019-nCoV Real-time PCR (Seegene, Seoul, Korea), PowerChek 2019-nCoV (KogeneBiotech, Seoul), Real-Q 2019-nCoV Real-Time Detection (BioSewoom, Seoul), and StandardM nCoV Detection (SD BIOSENSOR, Osong, Korea) were evaluated. The limit of detection (LODs) of Allplex, PowerChek, and Real-Q was determined by testing the transcribed RNA of SARS-CoV-2Eand the RNA of SARS-CoV Frankfurt1. A total of 27 consecutive samples comprising 13 sputum, 12 nasopharyngeal swab (NPS), 1 urine and 1 stool sample were collected from 2 COVID-19 patients for sensitivity analysis. Precision was assessed via daily tests of positive and negative controls in each kit for 5 d. Reproducibility was examined by repeating 21 samples and 10-fold dilutions of 14 samples in pairs using Allplex. Specificity was evaluated with 24 other respiratory virus-positive samples. LOD of Allplex, PowerChek, and Real-Q were 153.9, 84.1, and 80.6 copies/mL, respectively. The degrees of association between Cts and log viral concentrations by Allplex and PowerChek was expressed as y = -3.319 log (x) + 42.039 (R = 0.96) and y = -3.392 log(x) + 43.113 (R = 0.98), respectively. One or more of the 4 kits detected 20 out of 27 clinical samples positive. Of the 20 positive samples, the detection rates of positives for Allplex, PowerChek, Real-Q, and StandardM were 90.0, 82.3, 75.0, and 100.0%, respectively, but those of PowerChek and Real-Q would be 100% if out-of-cutoff Cts were counted as positives. Precision was 100%. Interpretation of Allplex results was reproducible when Ct ofE <= 33. All 4 kits showed no cross-reactivity with other respiratory viruses. Performance of the 4 kits indicated the suitability of these for diagnosis and follow-up testing of COVID-19. Laboratory doctors who initially implement these EUA kits must be able to interpret quality control parameters.
引用
收藏
页数:10
相关论文
共 20 条
[1]   Purification and characterization of PCR-inhibitory components in blood cells [J].
Abu al-Soud, W ;
Rådström, P .
JOURNAL OF CLINICAL MICROBIOLOGY, 2001, 39 (02) :485-493
[2]   Detection of 2019 novel coronavirus (2019-nCoV) by real-time RT-PCR (Publication with Expression of Concern) [J].
Corman, Victor M. ;
Landt, Olfert ;
Kaiser, Marco ;
Molenkamp, Richard ;
Meijer, Adam ;
Chu, Daniel K. W. ;
Bleicker, Tobias ;
Bruenink, Sebastian ;
Schneider, Julia ;
Schmidt, Marie Luisa ;
Mulders, Daphne G. J. C. ;
Haagmans, Bart L. ;
van der Veer, Bas ;
van den Brink, Sharon ;
Wijsman, Lisa ;
Goderski, Gabriel ;
Romette, Jean-Louis ;
Ellis, Joanna ;
Zambon, Maria ;
Peiris, Malik ;
Goossens, Herman ;
Reusken, Chantal ;
Koopmans, Marion P. G. ;
Drosten, Christian .
EUROSURVEILLANCE, 2020, 25 (03) :23-30
[3]   Guidelines for Laboratory Diagnosis of Coronavirus Disease 2019 (COVID-19) in Korea [J].
Hong, Ki Ho ;
Lee, Sang Won ;
Kim, Taek Soo ;
Huh, Hee Jae ;
Lee, Jaehyeon ;
Kim, So Yeon ;
Park, Jae-Sun ;
Kim, Gab Jeong ;
Sung, Heungsup ;
Roh, Kyoung Ho ;
Kim, Jae-Seok ;
Kim, Hyun Soo ;
Lee, Seung-Tae ;
Seong, Moon-Woo ;
Ryoo, Namhee ;
Lee, Hyukmin ;
Kwon, Kye Chul ;
Yoo, Cheon Kwon .
ANNALS OF LABORATORY MEDICINE, 2020, 40 (05) :351-360
[4]   Comparison of commercial realtime reverse transcription PCR assays for the detection of SARS-CoV-2 [J].
Igloi, Zsofia ;
Leven, Margareta ;
Abou-Nouar, Zain Abdel-Karem ;
Weller, Babette ;
Matheeussen, Veerle ;
Coppens, Jasmine ;
Koopmans, Marion ;
Molenkamp, Richard .
JOURNAL OF CLINICAL VIROLOGY, 2020, 129
[5]   Re: Low-dose corticosteroid therapy does not delay viral clearance in patients with COVID-19 [J].
Jung, Jiwon ;
Oh, Dong Kyu ;
Ahn, Jee Hwan ;
Hong, Sang-Bum ;
Sung, Heungsup ;
Kim, Mi-Na ;
Kim, Sung-Han .
JOURNAL OF INFECTION, 2020, 81 (02) :E79-E81
[6]   Korean Society for Laboratory Medicine Practice Guidelines for the Molecular Diagnosis of Middle East Respiratory Syndrome During an Outbreak in Korea in 2015 [J].
Ki, Chang-Seok ;
Lee, Hyukmin ;
Sung, Heungsup ;
Kim, Sinyoung ;
Seong, Moon-Woo ;
Yong, Dongeun ;
Kim, Jae-Seok ;
Lee, Mi-Kyung ;
Kim, Mi-Na ;
Choi, Jong-Rak ;
Kim, Jeong-Ho .
ANNALS OF LABORATORY MEDICINE, 2016, 36 (03) :203-208
[7]  
Kim M, 2020, PUBLIC HLTH WEEKLY R, V13, P800
[8]   Analytical and Clinical Validation of Six Commercial Middle East Respiratory Syndrome Coronavirus RNA Detection Kits Based on Real-Time Reverse-Transcription PCR [J].
Kim, Mi-Na ;
Ko, Young Jin ;
Seong, Moon-Woo ;
Kim, Jae-Seok ;
Shin, Bo-Moon ;
Sung, Heungsup .
ANNALS OF LABORATORY MEDICINE, 2016, 36 (05) :450-456
[9]   COVID-19 Testing in South Korea: Current Status and the Need for Faster Diagnostics [J].
Kim, Young Jin ;
Sung, Heungsup ;
Ki, Chang-Seok ;
Hur, Mina .
ANNALS OF LABORATORY MEDICINE, 2020, 40 (05) :349-350
[10]  
Korea Center for Disease Control and Prevention, 2020, UPDATES COVID 19 KOR