High-performance liquid chromatography and coulometric electrode array detector in serum 25-hydroxyvitamin D3 and 25-hydroxyvitamin D2 analyses

被引:23
作者
Nurmi, Tarja [1 ]
Tuomainen, Tomi-Pekka [1 ]
Virtanen, Jyrki [1 ]
Mursu, Jaakko [1 ]
Voutilainen, Sari [1 ]
机构
[1] Univ Eastern Finland, Inst Publ Hlth & Clin Nutr, FI-70211 Kuopio, Finland
基金
芬兰科学院;
关键词
25-Hydroxyvitamin D; Serum; HPLC; Coulometric electrode array detector; VITAMIN-D METABOLITES; ASSAY; HPLC; PLASMA;
D O I
10.1016/j.ab.2012.12.010
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
The diode array detector in our high-performance liquid chromatography (HPLC) method for 25-hydroxyvitamin D-3 (25OHD(3)) and 25-hydroxyvitamin D-2 (25OHD(2)) did not perform satisfactorily for measuring human serum concentrations below 30 nM. Because of a need for a reliable self-managed method in ongoing and starting vitamin D studies of the laboratory, we decided to develop a chromatographic method applying coulometric electrode array detector (CEAD) and evaluate reliability of the method by participating in the Vitamin D External Quality Assessment Scheme (DEQAS). The limit of quantification for 25OHD(3) and 25OHD(2) of the new method was 0.36 pmol on column (3.6 nM), and linearity ranged from 5 to 2400 nM. Accuracy of the method was 90% for 25OHD(3) and 69% for 25OHD(2). The HPLC-CEAD results from five DEQAS rounds were in line with those of the other participating laboratories using HPLC methods. The HPLC-CEAD results for 25OHD(3) also corresponded to the results obtained with the Chromsystems HPLC method in a certified laboratory. The long-term coefficients of variation for 25OHD(3) were 6.2%, 7.8%, 5.2%, 6.7%, and 7.3% in concentrations of 27.5, 38.7, 48.4, 78.4, and 88.0 nM, respectively. The developed HPLC-CEAD method was shown to be applicable for determining 25OHD(3) and 25OHD(2) in human serum samples. (C) 2013 Elsevier Inc. All rights reserved.
引用
收藏
页码:1 / 9
页数:9
相关论文
共 19 条
[1]   HPLC determination of Vitamin D3 and its metabolite in human plasma with on-line sample cleanup [J].
Brunetto, MR ;
Obando, MA ;
Gallignani, M ;
Alarcón, OM ;
Nieto, E ;
Salinas, R ;
Burguera, JL ;
Burguera, M .
TALANTA, 2004, 64 (05) :1364-1370
[2]   25-Hydroxyvitamin D: A Difficult Analyte [J].
Carter, Graham D. .
CLINICAL CHEMISTRY, 2012, 58 (03) :486-488
[3]   UK Food Standards Agency Workshop Consensus Report: the choice of method for measuring 25-hydroxyvitamin D to estimate vitamin D status for the UK National Diet and Nutrition Survey [J].
de la Hunty, Anne ;
Wallace, A. Michael ;
Gibson, Sigrid ;
Viljakainen, Heli ;
Lamberg-Allardt, Christel ;
Ashwell, Margaret .
BRITISH JOURNAL OF NUTRITION, 2010, 104 (04) :612-619
[4]  
Gamache P.H., 2000, HPLC COULOMETRIC ARR, P3
[5]   Current assays overestimate 25-hydroxyvitamin D3 and underestimate 25-hydroxyvitamin D2 compared with HPLC:: need for assay-specific decision limits and metabolite-specific assays [J].
Glendenning, P ;
Taranto, M ;
Noble, JM ;
Musk, AA ;
Hammond, C ;
Goldswain, PR ;
Fraser, WD ;
Vasikaran, SD .
ANNALS OF CLINICAL BIOCHEMISTRY, 2006, 43 :23-30
[6]   Issues of methodology, standardization and metabolite recognition for 25-hydroxyvitamin D when comparing the DiaSorin radioimmunoassay and the Nichols Advantage automated chemiluminescence protein-binding assay in hip fracture cases [J].
Glendenning, P ;
Noble, JM ;
Taranto, M ;
Musk, AA ;
McGuiness, M ;
Goldswain, PR ;
Fraser, WD ;
Vasikaran, SD .
ANNALS OF CLINICAL BIOCHEMISTRY, 2003, 40 :546-551
[7]   Exogenous versus endogenous recovery of 25-hydroxyvitamins D2 and D3 in human samples using high-performance liquid chromatography and the DiaSorin LIAISON Total-D Assay [J].
Horst, Ronald L. .
JOURNAL OF STEROID BIOCHEMISTRY AND MOLECULAR BIOLOGY, 2010, 121 (1-2) :180-182
[8]   The C-3 Epimer of 25-Hydroxyvitamin D3 Is Present in Adult Serum [J].
Lensmeyer, G. ;
Poquette, M. ;
Wiebe, D. ;
Binkley, N. .
JOURNAL OF CLINICAL ENDOCRINOLOGY & METABOLISM, 2012, 97 (01) :163-168
[9]   A novel high-performance liquid chromatographic assay for vitamin D metabolites using a coulometric electrochemical detector [J].
Masuda, S ;
Okano, T ;
Kamao, M ;
Kanedai, Y ;
Kobayashi, T .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 1997, 15 (9-10) :1497-1502
[10]   Fully automatic method for the determination of fat soluble vitamins and vitamin D metabolites in serum [J].
Mata-Granados, J. M. ;
Quesada Gomez, J. M. ;
Luque de Castro, M. D. .
CLINICA CHIMICA ACTA, 2009, 403 (1-2) :126-130