Imaging and spectroscopic performance studies of pixellated CdTe Timepix detector

被引:22
|
作者
Maneuski, D. [1 ]
Astromskas, V. [1 ]
Frojdh, E. [2 ]
Frojdh, C. [2 ]
Gimenez, E. N. [3 ]
Marchal, J. [3 ]
O'Shea, V. [1 ]
Stewart, G. [1 ]
Tartoni, N. [3 ]
Wilhelm, H. [3 ]
Wraight, K. [1 ]
Zain, R. M. [1 ]
机构
[1] Univ Glasgow, Sch Phys & Astron, Glasgow G12 8QQ, Lanark, Scotland
[2] Mid Sweden Univ, S-85170 Sundsvall, Sweden
[3] Diamond Light Source Ltd, Didcot OX11 0DE, Oxon, England
来源
关键词
Gamma detectors (scintillators; CZT; HPG; HgI etc); Hybrid detectors; Instrumentation for synchrotron radiation accelerators; READOUT CHIP;
D O I
10.1088/1748-0221/7/01/C01038
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
In this work the results on imaging and spectroscopic performances of 14 x 14 x 1 mm CdTe detectors with 55 x 55 mu m and 110 x 110 mu m pixel pitch bump-bonded to a Timepix chip are presented. The performance of the 110 x 110 mu m pixel detector was evaluated at the extreme conditions beam line I15 of the Diamond Light Source. The energy of X-rays was set between 25 and 77 keV. The beam was collimated through the edge slits to 20 mu m FWHM incident in the middle of the pixel. The detector was operated in the time-over-threshold mode, allowing direct energy measurement. Energy in the neighbouring pixels was summed for spectra reconstruction. Energy resolution at 77 keV was found to be Delta E/E = 3.9%. Comparative imaging and energy resolution studies were carried out between two pixel size detectors with a fluorescence target X-ray tube and radioactive sources. The 110 x 110 mu m pixel detector exhibited systematically better energy resolution in comparison to 55 x 55 mu m. An imaging performance of 55 x 55 mu m pixellated CdTe detector was assessed using the Modulation Transfer Function (MTF) technique and compared to the larger pixel. A considerable degradation in MTF was observed for bias voltages below -300V. Significant room for improvement of the detector performance was identified both for imaging and spectroscopy and is discussed.
引用
收藏
页数:10
相关论文
共 50 条
  • [21] Application of Timepix3 based CdTe spectral sensitive photon counting detector for PET imaging
    Turecek, Daniel
    Jakubek, Jan
    Trojanova, Eliska
    Sefc, Ludek
    Kolarova, Vera
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2018, 895 : 84 - 89
  • [22] Characterization of a pixellated CdTe detector with single-photon processing readout
    Fröjdh, C
    Graafsma, H
    Nilsson, HE
    Ponchut, C
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2006, 563 (01): : 128 - 132
  • [23] Energy sensitive X-ray phase contrast imaging with a CdTe-Timepix3 detector
    Navarrete, C.
    Procz, S.
    Fey, J.
    Roque, G.
    Avila, C.
    Schuetz, M.
    Olivo, A.
    Fiederle, M.
    2019 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2019,
  • [24] Spectroscopic performance of pixellated thallium bromide detectors
    Onodera, T
    Hitomi, K
    Shoji, T
    Hiratate, Y
    Kitaguchi, H
    2004 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-7, 2004, : 4448 - 4450
  • [25] Application of Dynamic Time-over-Threshold Method to Pixellated CdTe Detector
    Shimazoe, K.
    Nguyen, H.
    Orita, T.
    Wang, Y.
    Shi, B.
    Suzuki, T.
    Takahashi, H.
    2010 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD (NSS/MIC), 2010, : 3888 - 3890
  • [26] Spectral performance of a pixellated X-ray imaging detector with suppressed charge sharing
    Norlin, Borje
    Frojdh, Christer
    Nilsson, Hans-Erik
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2007, 576 (01): : 248 - 250
  • [27] Spectroscopic performance of pixellated thallium bromide detectors
    Onodera, T
    Hitomi, K
    Shoji, T
    Hiratate, Y
    Kitaguchi, H
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2005, 52 (05) : 1999 - 2002
  • [28] Experimental tests of a hybrid pixellated detector for gamma imaging
    Gal, O
    Mikulec, B
    Million, M
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2001, 460 (01): : 113 - 118
  • [29] An investigation of performance characteristics of a pixellated room-temperature semiconductor detector for medical imaging
    Guerra, P.
    Santos, A.
    Darambara, D. G.
    JOURNAL OF PHYSICS D-APPLIED PHYSICS, 2009, 42 (17)
  • [30] Autoradiography Imaging in Targeted Alpha Therapy with Timepix Detector
    AL Darwish, Ruqaya
    Staudacher, Alexander Hugo
    Bezak, Eva
    Brown, Michael Paul
    COMPUTATIONAL AND MATHEMATICAL METHODS IN MEDICINE, 2015, 2015