OBJECTIVE. The objective of this study was to standardize our image acquisition protocol for CT-guided biopsy procedures. MATERIALS AND METHODS. The records of consecutive patients who underwent CT-guided biopsy 3 months before (n = 598 biopsies) and 3 months after (n = 540 biopsies) standardization of our image acquisition protocol were retrospectively reviewed. CT technical parameters were individualized on the basis of the sum of the anteroposterior and transverse dimensions of the patient. Information on patient demographic characteristics, biopsy site, complications associated with the procedure, and diagnostic yield was collected. The radiation dose metrics that were evaluated included the volume CT dose index, dose-length product, and size-specific dose estimate. Image noise was quantified using the SD of the CT number measured in subcutaneous fat. Fisher exact test and one-way ANOVA were used to evaluate statistical significance. RESULTS. The mean dose-length product decreased by 72.3% (from 699.7 to 193.9 mGy x cm; p < 0.0001), and statistically significant decreases in dose-length product were observed when data were stratified according to biopsy site (i.e., lung, solid organ, lymph node, or bone; for all sites, p < 0.0001). The mean size-specific dose estimate decreased by 58.9% (from 125 to 51.4 mGy), which was statistically significant (p < 0.001). Image noise increased during the study period, but this increase was not statistically significantly different among the four biopsy sites (p = 0.46). CONCLUSION. Standardization of the image acquisition protocol used in CT-guided biopsy procedures significantly reduced patient radiation dose and decreased variability in image noise.
引用
收藏
页码:W390 / W399
页数:10
相关论文
共 50 条
  • [1] CT-guided interventional procedures without CT fluoroscopy assistance: Patient effective dose and absorbed dose considerations
    Tsalafoutas, Ioannis A.
    Tsapaki, Virginia
    Triantopoulou, Charicleia
    Gorantonaki, Akrivi
    Papailiou, John
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2007, 188 (06) : 1479 - 1484
  • [2] Radiation Dose Optimization for CT-Guided Interventional Procedures in the Abdomen and Pelvis
    Lamba, Ramit
    JOURNAL OF THE AMERICAN COLLEGE OF RADIOLOGY, 2014, 11 (03) : 279 - 284
  • [3] PANCREATIC TRANSPLANTS - CT-GUIDED BIOPSY
    BERNARDINO, M
    FERNANDEZ, M
    NEYLAN, J
    HERTZLER, G
    WHELCHEL, J
    OLSON, R
    RADIOLOGY, 1990, 177 (03) : 709 - 711
  • [4] Radiation Dose in CT-guided Interventional Procedures: Establishing a Benchmark
    Leng, Shuai
    RADIOLOGY, 2018, 289 (01) : 158 - 159
  • [5] CT-guided biopsy of an infected intervertebral disk
    Risselada, M
    Saunders, J
    Bhatti, S
    Gielen, I
    Van Ham, L
    van Bree, H
    VLAAMS DIERGENEESKUNDIG TIJDSCHRIFT, 2001, 70 (01): : 59 - 64
  • [6] CT-guided percutaneous biopsy of pancreas transplants
    Aideyan, OA
    Schmidt, AJ
    Trenkner, SW
    Hakim, NS
    Gruessner, RWG
    Walsh, JW
    RADIOLOGY, 1996, 201 (03) : 825 - 828
  • [7] CT-GUIDED BIOPSY OF SPINE LYTIC LESIONS
    FACCHINETTI, A
    GRIVETTA, MA
    BARUZZI, F
    MINONZIO, G
    MERCURI, A
    ALBERTINI, F
    BONALDI, G
    RIVISTA DI NEURORADIOLOGIA, 1995, 8 (02): : 215 - 221
  • [8] CT-guided core biopsy for retroperitoneal lesions. Experience in 136 procedures
    Lagos C, Claudio
    Gallardo, Eduardo
    Huete, Alvaro
    REVISTA MEDICA DE CHILE, 2019, 147 (10) : 1266 - 1272
  • [9] PET/CT-Guided Biopsy of Suspected Lung Lesions Requires Less Rebiopsy Than CT-Guided Biopsy Due to Inconclusive Results
    Cerci, Juliano J.
    Bogoni, Mateos
    Cerci, Rodrigo J.
    Masukawa, Margaret
    Neto, Carlos C. P.
    Krauzer, Cassiano
    Fanti, Stefano
    Sakamoto, Danielle G.
    Barreiros, Renan B.
    Nanni, Cristina
    Vitola, Joao V.
    JOURNAL OF NUCLEAR MEDICINE, 2021, 62 (08) : 1057 - 1061
  • [10] Procedure-specific CT Dose and Utilization Factors for CT-guided Interventional Procedures
    Yang, Kai
    Ganguli, Suvranu
    DeLorenzo, Matthew C.
    Zheng, Hui
    Li, Xinhua
    Liu, Bob
    RADIOLOGY, 2018, 289 (01) : 150 - 157