Posterior Glenoid Wear in Total Shoulder Arthroplasty: Eccentric Anterior Reaming Is Superior to Posterior Augment

被引:27
|
作者
Wang, Tim [1 ]
Abrams, Geoffrey D. [1 ,2 ]
Behn, Anthony W. [1 ]
Lindsey, Derek [2 ]
Giori, Nicholas [2 ]
Cheung, Emilie V. [1 ]
机构
[1] Stanford Univ, Dept Orthopaed Surg, Redwood City, CA 94063 USA
[2] Vet Affairs Palo Alto Hlth Care Syst, Ctr Tissue Regenerat Repair & Restorat, Palo Alto, CA USA
关键词
TOTAL KNEE ARTHROPLASTY; BONE DEFICIENCY; WEDGE AUGMENTATION; COMPONENTS; EROSION; OSTEOARTHRITIS; RETROVERSION; MANAGEMENT; FIXATION; DESIGN;
D O I
10.1007/s11999-015-4482-8
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background Uncorrected glenoid retroversion during total shoulder arthroplasty may lead to an increased likelihood of glenoid prosthetic loosening. Augmented glenoid components seek to correct retroversion to address posterior glenoid bone loss, but few biomechanical studies have evaluated their performance. Questions/purposes We compared the use of augmented glenoid components with eccentric reaming with standard glenoid components in a posterior glenoid wear model. The primary outcome for biomechanical stability in this model was assessed by (1) implant edge displacement in superior and inferior edge loading at intervals up to 100,000 cycles, with secondary outcomes including (2) implant edge load during superior and inferior translation at intervals up to 100,000 cycles, and (3) incidence of glenoid fracture during implant preparation and after cyclic loading. Methods A 12A degrees-posterior glenoid defect was created in 12 composite scapulae, and the specimens were divided in two equal groups. In the posterior augment group, glenoid version was corrected to 8A degrees and an 8A degrees-augmented polyethylene glenoid component was placed. In the eccentric reaming group, anterior glenoid reaming was performed to neutral version and a standard polyethylene glenoid component was placed. Specimens were cyclically loaded in the superoinferior direction to 100,000 cycles. Superior and inferior glenoid edge displacements were recorded. Results Surviving specimens in the posterior augment group showed greater displacement than the eccentric reaming group of superior (1.01 +/- 0.02 [95% CI, 0.89-1.13] versus 0.83 +/- 0.10 [95% CI, 0.72-0.94 mm]; mean difference, 0.18 mm; p = 0.025) and inferior markers (1.36 +/- 0.05 [95% CI, 1.24-1.48] versus 1.20 +/- 0.09 [95% CI, 1.09-1.32 mm]; mean difference, 0.16 mm; p = 0.038) during superior edge loading and greater displacement of the superior marker during inferior edge loading (1.44 +/- 0.06 [95% CI, 1.28-1.59] versus 1.16 +/- 0.11 [95% CI, 1.02-1.30 mm]; mean difference, 0.28 mm; p = 0.009) at 100,000 cycles. No difference was seen with the inferior marker during inferior edge loading (0.93 +/- 0.15 [95% CI, 0.56-1.29] versus 0.78 +/- 0.06 [95% CI, 0.70-0.85 mm]; mean difference, 0.15 mm; p = 0.079). No differences in implant edge load were seen during superior and inferior loading. There were no instances of glenoid vault fracture in either group during implant preparation; however, a greater number of specimens in the eccentric reaming group were able to achieve the final 100,000 time without catastrophic fracture than those in the posterior augment group. Conclusions When addressing posterior glenoid wear in surrogate scapula models, use of angle-backed augmented glenoid components results in accelerated implant loosening compared with neutral-version glenoid after eccentric reaming, as shown by increased implant edge displacement at analogous times.
引用
收藏
页码:3928 / 3936
页数:9
相关论文
共 50 条
  • [1] Posterior glenoid bone grafting in total shoulder arthroplasty for osteoarthritis with severe posterior glenoid wear
    Nicholson, Gregory P.
    Cvetanovich, Gregory L.
    Rao, Allison J.
    O'Donnell, Patrick
    JOURNAL OF SHOULDER AND ELBOW SURGERY, 2017, 26 (10) : 1844 - 1853
  • [2] Complications of shoulder arthroplasty for osteoarthritis with posterior glenoid wear
    Gallusser, N.
    Farron, A.
    ORTHOPAEDICS & TRAUMATOLOGY-SURGERY & RESEARCH, 2014, 100 (05) : 503 - 508
  • [3] Vault perforation after eccentric glenoid reaming for deformity correction in anatomic total shoulder arthroplasty
    Olszewski, Adam
    Ramme, Austin J.
    Maerz, Tristan
    Freehill, Michael T.
    Warner, Jon J. P.
    Bedi, Asheesh
    JOURNAL OF SHOULDER AND ELBOW SURGERY, 2020, 29 (07) : 1450 - 1459
  • [4] 3D fluoroscopic navigated reaming of the glenoid for total shoulder arthroplasty (TSA)
    Briem, Daniel
    Ruecker, Andreas H.
    Neumann, Joerg
    Gebauer, Matthias
    Kendoff, Daniel
    Gehrke, Thorsten
    Lehmann, Wolfgang
    Schumacher, Udo
    Rueger, Johannes. M.
    Grossterlinden, Lars G.
    COMPUTER AIDED SURGERY, 2011, 16 (02) : 93 - 99
  • [5] The effect of eccentric glenoid reaming in reverse shoulder artrhoplasty for glenohumeral osteoarthritis
    Kececi, Tolga
    Ucan, Vahdet
    Ertogrul, Rodi
    Sahin, Koray
    Bilsel, Kerem
    Kapicioglu, Mehmet
    JOURNAL OF ORTHOPAEDICS, 2024, 50 : 111 - 115
  • [6] Glenoid morphology after reaming in computer-simulated total shoulder arthroplasty
    Yongpravat, Charlie
    Lester, Jonathan D.
    Saifi, Comron
    Trubelja, Alen
    Greiwe, R. Michael
    Bigliani, Louis U.
    Levine, William N.
    Gardner, Thomas R.
    Ahmad, Christopher S.
    JOURNAL OF SHOULDER AND ELBOW SURGERY, 2013, 22 (01) : 122 - 128
  • [7] Glenoid subchondral bone density distribution in male total shoulder arthroplasty subjects with eccentric and concentric wear
    Simon, Peter
    Gupta, Anil
    Pappou, Ioannis
    Hussey, Michael M.
    Santoni, Brandon G.
    Inoue, Nozomu
    Frankle, Mark A.
    JOURNAL OF SHOULDER AND ELBOW SURGERY, 2015, 24 (03) : 416 - 424
  • [8] Clinical and Radiographic Outcomes of a Posteriorly Augmented Glenoid Component in Anatomic Total Shoulder Arthroplasty for Primary Osteoarthritis with Posterior Glenoid Bone Loss
    Ho, Jason C.
    Amini, Michael H.
    Entezari, Vahid
    Jun, Bong Jae
    Alolabi, Bashar
    Ricchetti, Eric T.
    Iannotti, Joseph P.
    JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2018, 100 (22) : 1934 - 1948
  • [9] Treatment of Glenoid Wear with the Use of Augmented Glenoid Components in Total Shoulder Arthroplasty
    Contreras, Erik S.
    Kingery, Matthew T.
    Zuckerman, Joseph D.
    Virk, Mandeep S.
    JBJS REVIEWS, 2023, 11 (10)
  • [10] Addressing glenoid bone deficiency and asymmetric posterior erosion in shoulder arthroplasty
    Hsu, Jason E.
    Ricchetti, Eric T.
    Huffman, G. Russell
    Iannotti, Joseph P.
    Glaser, David L.
    JOURNAL OF SHOULDER AND ELBOW SURGERY, 2013, 22 (09) : 1298 - 1308