摘要
Stereotactic frames are a standard tool for neurosurgical targeting, but are uncomfortable for patients and obstruct the surgical field. Microstereotactic frames are more comfortable for patients, provide better access to the surgical site, and have grown in popularity as an alternative to traditional stereotactic devices. However, clinically available microstereotactic frames require either lengthy manufacturing delays or expensive image guidance systems. We introduce a robotically-adjusted, disposable microstereotactic frame for deep brain stimulation surgery that eliminates the drawbacks of existing microstereotactic frames. Our frame can be automatically adjusted in the operating room using a preoperative plan in less than five minutes. A validation study on phantoms shows that our approach provides a target positioning error of 0.14 mm, which exceeds the required accuracy for deep brain stimulation surgery
引用
收藏
页数:11
相关论文
共 18 条
  • [1] [Anonymous], 2000, Handbook of Medical Imaging
  • [2] Comparison of accuracy and precision between frame-based and frameless stereotactic navigation for deep brain stimulation electrode implantation
    Bjartmarz, Hjalmar
    Rehncrona, Stig
    [J]. STEREOTACTIC AND FUNCTIONAL NEUROSURGERY, 2007, 85 (05) : 235 - 242
  • [3] Fitzpatrick J. M., 2012, SPIE MED IMAGING
  • [4] Fitzpatrick J. M., 2012, TECH REP
  • [5] Predicting error in rigid-body point-based registration
    Fitzpatrick, JM
    West, JB
    Maurer, CR
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 1998, 17 (05) : 694 - 702
  • [6] Accuracy of customized miniature stereotactic platforms
    Fitzpatrick, JM
    Konrad, PE
    Nickele, C
    Cetinkaya, E
    Kao, C
    [J]. STEREOTACTIC AND FUNCTIONAL NEUROSURGERY, 2005, 83 (01) : 25 - 31
  • [7] POTENTIAL HUMAN ERROR IN SETTING STEREOTAXIC COORDINATES FOR RADIOSURGERY - IMPLICATIONS FOR QUALITY ASSURANCE
    FLICKINGER, JC
    LUNSFORD, LD
    KONDZIOLKA, D
    MAITZ, A
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1993, 27 (02): : 397 - 401
  • [8] Frameless stereotaxy using bone fiducial markers for deep brain stimulation
    Holloway, KL
    Gaede, SE
    Starr, PA
    Rosenow, JM
    Ramakrishnan, V
    Henderson, JM
    [J]. JOURNAL OF NEUROSURGERY, 2005, 103 (03) : 404 - 413
  • [9] Joskowicz L, 2006, Comput Aided Surg, V11, P181, DOI 10.1080/10929080600909351
  • [10] Design and analysis of a head-mounted parallel kinematic device for skull surgery
    Kobler, Jan-Philipp
    Kotlarski, Jens
    Oeltjen, Julian
    Baron, Stephan
    Ortmaier, Tobias
    [J]. INTERNATIONAL JOURNAL OF COMPUTER ASSISTED RADIOLOGY AND SURGERY, 2012, 7 (01) : 137 - 149