The biomechanical effect of artificial and human bone density on stopping and stripping torque during screw insertion

被引:40
|
作者
Tsuji, Matthew [1 ]
Crookshank, Meghan [1 ,2 ]
Olsen, Michael [1 ,2 ]
Schemitsch, Emil H. [1 ,2 ]
Zdero, Racl [2 ,3 ]
机构
[1] Univ Toronto, Fac Med, Toronto, ON M5S 1A8, Canada
[2] Lab St Michaels Hosp, Martin Orthoped Biomech, Toronto, ON M5B 1W8, Canada
[3] Ryerson Univ, Dept Mech & Ind Engn, Toronto, ON M5B 2K3, Canada
关键词
Biomechanics; Artificial; Human; Density; Cortical; Cancellous; Stopping torque; Stripping torque; CANCELLOUS BONE; FRACTURE FIXATION; BENDING STRENGTH; PULLOUT STRENGTH; PEDICLE SCREWS; HOLDING POWER; FEMUR; SHAFT; HIP;
D O I
10.1016/j.jmbbm.2013.03.006
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Orthopedic surgeons apply torque to metal screws manually by "subjective feel" to obtain adequate fracture fixation, i.e. stopping torque, and attempt to avoid accidental overtightening that leads to screw-bone interface failure, i.e. stripping torque. Few studies have quantified stripping torque in human bone, and only one older study from 1980 reported stopping/stripping torque ratio. The present aim was to measure stopping and stripping torque of cortical and cancellous screws in artificial and human bone over a wide range of densities. Sawbone blocks were obtained having densities from 0.08 to 0.80 g/cm(3). Sixteen fresh-frozen human femurs of known standardized bone mineral density (sBMD) were also used. Using a torque screwdriver, 3.5-mm diameter cortical screws and 6.5-mm diameter cancellous screws were inserted for adequate tightening as determined subjectively by an orthopedic surgeon, i.e. stopping torque, and then further tightened until failure of the screw-bone interface, i.e. stripping torque. There were weak (R=0.25) to strong (R=0.99) linear correlations of absolute and normalized torque vs. density or sBMD. Maximum stopping torques normalized by screw thread area engaged by the host material were 15.2 N/mm (cortical screws) and 13.4 N/mm (cancellous screws) in sawbone blocks and 20.9 N/mm (cortical screws) and 6.1 N/mm (cancellous screws) in human femurs. Maximum stripping torques normalized by screw thread area engaged by the host material were 23.4 N/mm (cortical screws) and 16.8 N/mm (cancellous screws) in sawbone blocks and 29.3 N/mm (cortical screws) and 8.3 N/mm (cancellous screws) in human femurs. Combined average stopping/ stripping torque ratios were 80.8% (cortical screws) and 76.8% (cancellous screws) in sawbone blocks, as well as 66.6% (cortical screws) and 84.5% (cancellous screws) in human femurs. Surgeons should be aware of stripping torque limits for human femurs and monitor stopping torque during surgery. This is the first study of the effect of sawbone density or human bone sBMD on stopping and stripping torque. (C) 2013 Elsevier Ltd. All rights reserved.
引用
收藏
页码:146 / 156
页数:11
相关论文
共 50 条
  • [41] The effect of different screw-rod design on the anti-rotational torque: a biomechanical comparison of three conventional screwrod constructs
    Huang, Zifang
    Wang, Chongwen
    Fan, Hengwei
    Sui, Wenyuan
    Li, Xueshi
    Wang, Qifei
    Yang, Junlin
    BMC MUSCULOSKELETAL DISORDERS, 2017, 18
  • [42] Determining relationship between bone screw insertion torque and insertion speed Bestimmung des Zusammenhangs zwischen dem Drehmoment beim Eindrehen von Knochenschrauben und der Eindrehgeschwindigkeit
    Wilkie, Jack A.
    Rauter, Georg
    Moeller, Knut
    AT-AUTOMATISIERUNGSTECHNIK, 2022, 70 (11) : 976 - 991
  • [43] The effect of pilot hole size on the insertion torque and pullout strength of self-tapping cortical bone screws in osteoporotic bone
    Battula, Suneel
    Schoenfeld, Andrew J.
    Sahai, Vivek
    Vrabec, Gregory A.
    Tank, Jason
    Njus, Glen O.
    JOURNAL OF TRAUMA-INJURY INFECTION AND CRITICAL CARE, 2008, 64 (04): : 990 - 995
  • [44] Biomechanical Analysis of the human femur bone during going upstairs and sitting down
    Yousif, Albert E.
    Aziz, Mustafa Y.
    2012 FIRST NATIONAL CONFERENCE FOR ENGINEERING SCIENCES (FNCES), 2012,
  • [45] Juvenile bovine bone is an appropriate surrogate for normal and reduced density human bone in biomechanical testing: a validation study
    Fletcher, J. W. A.
    Williams, S.
    Whitehouse, M. R.
    Gill, H. S.
    Preatoni, E.
    SCIENTIFIC REPORTS, 2018, 8
  • [46] Torque dependent insertion depth of Conelog implants in artificial bone of D1 quality. An in-vitro study
    Mueller, Friedrich
    Mueller, Janine
    IMPLANTOLOGIE, 2016, 24 (03): : 327 - 333
  • [47] ASSESSMENT OF UNICORTICAL AND BICORTICAL FIXATION IN A QUASI-STATIC CADAVERIC MODEL - ROLE OF BONE-MINERAL DENSITY AND SCREW TORQUE
    RYKEN, TC
    GOEL, VK
    CLAUSEN, JD
    TRAYNELIS, VC
    SPINE, 1995, 20 (17) : 1861 - 1867
  • [48] The effect of screw taper on interference fit during load to failure at the soft tissue/bone
    Mann, CJ
    Costi, JJ
    Stanley, RA
    Dobson, PJ
    KNEE, 2005, 12 (05) : 370 - 376
  • [49] Biomechanical investigation of the effect of high hydrostatic pressure treatment on the mechanical properties of human bone
    Steinhauser, E
    Diehl, P
    Hadaller, M
    Schauwecker, J
    Busch, R
    Gradinger, R
    Mittelmeier, W
    JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART B-APPLIED BIOMATERIALS, 2006, 76B (01) : 130 - 135
  • [50] The effect of two types of resorbable augmentation materials - a cement and an adhesive - on the screw pullout pullout resistance in human trabecular bone
    Wu, Dan
    Pujari-Palmer, Michael
    Bojan, Alicja
    Palmquist, Anders
    Procter, Philip
    Ohman-Magi, Caroline
    Ferguson, Stephen J.
    Isaksson, Per
    Persson, Cecilia
    JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2020, 110 (110)