The biomechanical effect of artificial and human bone density on stopping and stripping torque during screw insertion

被引:40
|
作者
Tsuji, Matthew [1 ]
Crookshank, Meghan [1 ,2 ]
Olsen, Michael [1 ,2 ]
Schemitsch, Emil H. [1 ,2 ]
Zdero, Racl [2 ,3 ]
机构
[1] Univ Toronto, Fac Med, Toronto, ON M5S 1A8, Canada
[2] Lab St Michaels Hosp, Martin Orthoped Biomech, Toronto, ON M5B 1W8, Canada
[3] Ryerson Univ, Dept Mech & Ind Engn, Toronto, ON M5B 2K3, Canada
关键词
Biomechanics; Artificial; Human; Density; Cortical; Cancellous; Stopping torque; Stripping torque; CANCELLOUS BONE; FRACTURE FIXATION; BENDING STRENGTH; PULLOUT STRENGTH; PEDICLE SCREWS; HOLDING POWER; FEMUR; SHAFT; HIP;
D O I
10.1016/j.jmbbm.2013.03.006
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Orthopedic surgeons apply torque to metal screws manually by "subjective feel" to obtain adequate fracture fixation, i.e. stopping torque, and attempt to avoid accidental overtightening that leads to screw-bone interface failure, i.e. stripping torque. Few studies have quantified stripping torque in human bone, and only one older study from 1980 reported stopping/stripping torque ratio. The present aim was to measure stopping and stripping torque of cortical and cancellous screws in artificial and human bone over a wide range of densities. Sawbone blocks were obtained having densities from 0.08 to 0.80 g/cm(3). Sixteen fresh-frozen human femurs of known standardized bone mineral density (sBMD) were also used. Using a torque screwdriver, 3.5-mm diameter cortical screws and 6.5-mm diameter cancellous screws were inserted for adequate tightening as determined subjectively by an orthopedic surgeon, i.e. stopping torque, and then further tightened until failure of the screw-bone interface, i.e. stripping torque. There were weak (R=0.25) to strong (R=0.99) linear correlations of absolute and normalized torque vs. density or sBMD. Maximum stopping torques normalized by screw thread area engaged by the host material were 15.2 N/mm (cortical screws) and 13.4 N/mm (cancellous screws) in sawbone blocks and 20.9 N/mm (cortical screws) and 6.1 N/mm (cancellous screws) in human femurs. Maximum stripping torques normalized by screw thread area engaged by the host material were 23.4 N/mm (cortical screws) and 16.8 N/mm (cancellous screws) in sawbone blocks and 29.3 N/mm (cortical screws) and 8.3 N/mm (cancellous screws) in human femurs. Combined average stopping/ stripping torque ratios were 80.8% (cortical screws) and 76.8% (cancellous screws) in sawbone blocks, as well as 66.6% (cortical screws) and 84.5% (cancellous screws) in human femurs. Surgeons should be aware of stripping torque limits for human femurs and monitor stopping torque during surgery. This is the first study of the effect of sawbone density or human bone sBMD on stopping and stripping torque. (C) 2013 Elsevier Ltd. All rights reserved.
引用
收藏
页码:146 / 156
页数:11
相关论文
共 50 条
  • [31] The effect of different screw-rod design on the anti-rotational torque: a biomechanical comparison of three conventional screw-rod constructs
    Zifang Huang
    Chongwen Wang
    Hengwei Fan
    Wenyuan Sui
    Xueshi Li
    Qifei Wang
    Junlin Yang
    BMC Musculoskeletal Disorders, 18
  • [32] The biomechanical effect of changes in cancellous bone density on synthetic femur behaviour
    Nicayenzi, B.
    Shah, S.
    Schemitsch, E. H.
    Bougherara, H.
    Zdero, R.
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2011, 225 (H11) : 1050 - 1060
  • [33] Comparison between sheep and human cervical spines -: An anatomic, radiographic, bone mineral density, and biomechanical study
    Kandziora, F
    Pflugmacher, R
    Scholz, M
    Schnake, K
    Lucke, M
    Schröder, R
    Mittlmeier, T
    SPINE, 2001, 26 (09) : 1028 - 1037
  • [34] Effects of Velocity Profile and Plate Usage on Identified Bone Strength during Instrumented Screw Insertion
    Wilkie, Jack
    Bhave, Ashish
    Rauter, Georg
    Moeller, Knut
    IFAC PAPERSONLINE, 2023, 56 (02): : 8237 - 8241
  • [35] Complex biomechanical properties of non-augmented and augmented pedicle screws in human vertebrae with reduced bone density
    Martin Schulze
    Oliver Riesenbeck
    Thomas Vordemvenne
    Michael J. Raschke
    Julia Evers
    René Hartensuer
    Dominic Gehweiler
    BMC Musculoskeletal Disorders, 21
  • [36] Complex biomechanical properties of non-augmented and augmented pedicle screws in human vertebrae with reduced bone density
    Schulze, Martin
    Riesenbeck, Oliver
    Vordemvenne, Thomas
    Raschke, Michael J.
    Evers, Julia
    Hartensuer, Rene
    Gehweiler, Dominic
    BMC MUSCULOSKELETAL DISORDERS, 2020, 21 (01)
  • [37] The effect of pedicle screw redirection after lateral wall breach-a biomechanical study using human lumbar vertebrae
    Stauff, Michael P.
    Freedman, Brett A.
    Kim, Jin-Hwan
    Hamasaki, Takahiko
    Yoon, S. Tim
    Hutton, William C.
    SPINE JOURNAL, 2014, 14 (01) : 98 - 103
  • [38] Effect of Iliac Screw Insertion Depth on the Stability and Strength of Lumbo-Iliac Fixation Constructs An Anatomical and Biomechanical Study
    Zheng, Zhao-Min
    Yu, Bin-Sheng
    Chen, Hui
    Aladin, Darwesh M. K.
    Zhang, Kui-Bo
    Zhang, Jia-Fang
    Liu, Hui
    Luk, Keith D. K.
    Lu, William W.
    SPINE, 2009, 34 (16) : E565 - E572
  • [39] Biomechanical Considerations in Plate Osteosynthesis: The Effect of Plate-to-Bone Compression With and Without Angular Screw Stability
    Stoffel, Karl
    Lorenz, Kai-Uwe
    Kuster, Markus S.
    JOURNAL OF ORTHOPAEDIC TRAUMA, 2007, 21 (06) : 362 - 368
  • [40] Effect of Insertional Direction of Pedicle Screw on Screw Loosening: A Biomechanical Study on Synthetic Bone Vertebra under a Physiology-like Load
    Zhan, Xuqiang
    Gao, Feng
    Yang, Yangyang
    Tsai, Tsung-Yuan
    Wan, Zongmiao
    Yu, Yan
    ORTHOPAEDIC SURGERY, 2024, 16 (06) : 1461 - 1472